Literature DB >> 16434520

The premature failure of the Charnley Elite-Plus stem: a confirmation of RSA predictions.

J Hauptfleisch1, S Glyn-Jones, D J Beard, H S Gill, D W Murray.   

Abstract

We performed a clinical and radiological study to determine the rate of failure of the Charnley Elite-Plus femoral component. Our aim was to confirm or refute the predictions of a previous roentgen stereophotogrammetric analysis study in which 20% of the Charnley Elite-Plus stems had shown rapid posterior head migration. It was predicted that this device would have a high early rate of failure.We examined 118 patients at a mean of nine years after hip replacement, including the 19 patients from the original roentgen stereophotogrammetric study. The number of revision procedures was recorded and clinical and radiological examinations were performed. The rate of survival of the femoral stems at ten years was 83% when revision alone was considered to be a failure. It decreased to 59% when a radiologically loose stem was also considered to be a failure. All the patients previously shown in the roentgen stereophotogrammetric study to have high posterior head migration went on to failure. There was a highly significant difference (p = 0.002) in posterior head migration measured at two years after operation between failed and non-failed femoral stems, but there was no significant difference in subsidence between these two groups. Our study has shown that the Charnley Elite-Plus femoral component has an unacceptably high rate of failure. It confirms that early evaluation of new components is important and that roentgen stereophotogrammetric is a good tool for this. Our findings have also shown that rapid posterior head migration is predictive of premature loosening and a better predictor than subsidence.

Entities:  

Mesh:

Year:  2006        PMID: 16434520     DOI: 10.1302/0301-620X.88B2.17055

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  26 in total

1.  Early Subsidence Predicts Failure of a Cemented Femoral Stem With Minor Design Changes.

Authors:  Per-Erik Johanson; Martin Antonsson; Bita Shareghi; Johan Kärrholm
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

2.  The Elite-Plus stem migrates more than the flanged Charnley stem.

Authors:  Thord von Schewelov; Lennart Sanzén; Jack Besjakov; Ake Carlsson
Journal:  Acta Orthop       Date:  2010-06       Impact factor: 3.717

3.  Migration pattern of cemented Exeter short stem in Dorr type A femurs. A prospective radiostereometry study with 2-year follow-up.

Authors:  Tobias Dahl Vind; Peter Bo Jørgensen; Dovydas Vainorius; Stig Storgaard Jakobsen; Kjeld Søballe; Maiken Stilling
Journal:  Arch Orthop Trauma Surg       Date:  2022-02-03       Impact factor: 3.067

Review 4.  Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties.

Authors:  Bart G Pijls; Edward R Valstar; Klaas-Auke Nouta; Josepha Wm Plevier; Marta Fiocco; Saskia Middeldorp; Rob Ghh Nelissen
Journal:  Acta Orthop       Date:  2012-11-09       Impact factor: 3.717

5.  Excessive distal migration of fiber-mesh coated femoral stems.

Authors:  Thomas Baad-Hansen; Søren Kold; Niels Olsen; Finn Christensen; Kjeld Søballe
Journal:  Acta Orthop       Date:  2011-04-19       Impact factor: 3.717

6.  Effect of local TGF-beta1 and IGF-1 release on implant fixation: comparison with hydroxyapatite coating: a paired study in dogs.

Authors:  Anders Lamberg; Joan E Bechtold; Jørgen Baas; Kjeld Søballe; Brian Elmengaard
Journal:  Acta Orthop       Date:  2009-08       Impact factor: 3.717

Review 7.  Temporal subsidence patterns of cemented polished taper-slip stems: a systematic review.

Authors:  Kwaku Baryeh; Jeewaka Mendis; David H Sochart
Journal:  EFORT Open Rev       Date:  2021-05-04

8.  5-year clinical and radiostereometric analysis (RSA) follow-up of 39 CUT femoral neck total hip prostheses in young osteoarthritis patients.

Authors:  Marc J Nieuwenhuijse; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2012-08-10       Impact factor: 3.717

9.  Does a new knee design perform as well as the design it replaces?

Authors:  M Molt; P Ljung; S Toksvig-Larsen
Journal:  Bone Joint Res       Date:  2012-12-01       Impact factor: 5.853

Review 10.  Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies.

Authors:  Bart G Pijls; Marc J Nieuwenhuijse; Marta Fiocco; Josepha Wm Plevier; Saskia Middeldorp; Rob Ghh Nelissen; Edward R Valstar
Journal:  Acta Orthop       Date:  2012-11-05       Impact factor: 3.717

View more