| Literature DB >> 28245864 |
Sean M Hamlet1,2, Christopher M Haggerty2,3,4, Jonathan D Suever2,3,4, Gregory J Wehner2,5, Kristin N Andres2, David K Powell5, Richard J Charnigo6, Brandon K Fornwalt7,8,9,10,11,12.
Abstract
BACKGROUND: Left ventricular (LV) torsion is an important indicator of cardiac function that is limited by high inter-test variability (50% of the mean value). We hypothesized that this high inter-test variability is partly due to inconsistent breath-hold positions during serial image acquisitions, which could be significantly improved by using a respiratory navigator for cardiovascular magnetic resonance (CMR) based quantification of LV torsion.Entities:
Keywords: Breath-holds; Cardiovascular magnetic resonance; DENSE; Left ventricular torsion; Respiratory navigator gating
Mesh:
Year: 2017 PMID: 28245864 PMCID: PMC5331707 DOI: 10.1186/s12968-017-0338-6
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Fig. 1Computation of LV torsion from basal and apical images. The curved arrows represent the relative twist along the longitudinal axis of the left ventricle. LV twist (ϕ) was measured as the difference in rotation between the apex (ϕ ) and base (ϕ ) (twist direction shown as viewed from foot to head). Torsion (τ) was computed as LV twist divided by the distance (d) between basal and apical image locations
Fig. 2Real-time images of the diaphragm as it translates during a respiratory cycle. During respiration, diaphragm motion causes the heart to translate a substantial distance through the fixed imaging plane
Fig. 3Respiratory navigator gating. (Left) The diaphragm position is at the high-contrast interface between the lung (dark) and the liver (bright). (Right) Image of a measured diaphragm position over time for separate breath-holds. For this subject, there was an 11 mm difference in end-expiratory position between breath-hold 1 and 10
Fig. 4Measured end-expiratory diaphragm positions were used to define subject-specific maximum, middle, and minimum end-expiratory positions. The maximum diaphragm position was defined as being closer to the end-expiratory position while the minimum diaphragm position was defined as being closer to the end-inspiratory position
Fig. 5The nine possible torsion permutations were constructed from three basal and three apical images. One basal and apical image was acquired for each subject-specific end-expiratory position (maximum, middle, and minimum). Image acquisitions were repeated at the middle position to assess inter-test variability (far right)
Fig. 6Inconsistent end-expiratory positions across ten consecutive breath-holds in patients and healthy controls. There were no significant differences in either the range (a) or standard deviation (b) of end-expiratory position between the healthy and patient groups. Solid red lines denote the mean for each group
Fig. 7DENSE images from a representative subject show the relative twist differences between the basal and apical images at end-systole. Twist in the basal region is predominantly in the clockwise direction, while the apex is predominantly counter-clockwise
Mean (± standard deviation) of torsion across the volunteers within each experiment
| Method (experiment) | Torsion (°/cm) |
|
|---|---|---|
| Experiment 1a | ||
| Enforced inconsistent positions | 3.4 ± 0.4 | 0.85 |
| Consistent positions with navigator | 3.4 ± 0.2 | |
| Experiment 2 | ||
| Breath-holds | 3.6 ± 0.3 | 0.32 |
| Consistent positions with navigator | 3.5 ± 0.2 |
aReported values are from combined group of healthy and patient volunteers
Fig. 8Variability of torsion due to enforced inconsistent end-expiratory positions versus the subject-specific range of end-expiratory position. There was a moderate positive correlation between RMSE of LV torsion due to inconsistent expiratory positions and the range of end-expiratory position (r = 0.50, p = 0.049). The dashed gray line illustrates the linear best fit
Fig. 9Variability of torsion due to naturally inconsistent end-expiratory positions versus the standard deviation of end-expiratory position. There was a moderate positive correlation between the standard deviation of LV torsion and the standard deviation of end-expiratory position (r = 0.34, p = 0.03). The dashed gray line illustrates the linear best fit
Sample sizes required to detect a 10% relative change in LV torsion calculated using data in this and previous studies
| Method | Sample size (n) |
|---|---|
| Experiment 1 | |
| Enforced inconsistent positions | 66 |
| Consistent positions with navigator | 16 |
| Experiment 2 | |
| Breath-holds | 26 |
| Consistent positions with navigator | 15 |
| Previous Studies | |
| CMR Tagging [ | 107 |
| CMR Feature Tracking [ | 81 |
| 3D Speckle Tracking [ | 80 |