A monitoring, feedback, and triggering system for reproducible breath-hold MR imaging.

Y L Liu1, S J Riederer, P J Rossman, R C Grimm, J P Debbins, R L Ehman.   

Abstract

A technique is described that provides improved reproducibility of breath-holding for MR image acquisition by monitoring the superior-inferior (S/I) position of the diaphragm. The method incorporates detection of the level of inspiration using an MR signal, rapid display to the patient of diaphragm position to enable breath-hold adjustment, and triggering of image data acquisition once appropriate position is attained. The response time of the system is short, approximately 10 ms. Studies in six volunteers using this method demonstrate a considerable decrease in the S/I range of diaphragm position over 10 consecutive periods of suspended respiration. The mean range is 1.3 mm with the system, while it is 8.3 mm without using it. It is expected that this method will be of assistance in many abdominal and cardiothoracic studies that use breath-hold techniques.

Entities:  

Mesh:

Year:  1993        PMID: 8255201     DOI: 10.1002/mrm.1910300416

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  22 in total

1.  Prospective navigator gating with a dual acceptance window technique to reduce respiratory motion artifacts in 3D MR coronary angiography.

Authors:  Yiping P Du
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

Review 2.  Magnetic resonance coronary angiography.

Authors:  Phillip C Yang; Michael V McConnell; Dwight G Nishimura; Bob S Hu
Journal:  Curr Cardiol Rep       Date:  2003-01       Impact factor: 2.931

3.  On spatially selective RF excitation and its analogy with spiral MR image acquisition.

Authors:  P Börnert; B Aldefeld
Journal:  MAGMA       Date:  1998-12       Impact factor: 2.310

4.  3D motion adapted gating (3D MAG): a new navigator technique for accelerated acquisition of free breathing navigator gated 3D coronary MR-angiography.

Authors:  M Hackenbroch; K Nehrke; J Gieseke; C Meyer; K Tiemann; H Litt; O Dewald; C P Naehle; H Schild; T Sommer
Journal:  Eur Radiol       Date:  2005-04-07       Impact factor: 5.315

5.  Dual-contrast single breath-hold 3D abdominal MR imaging.

Authors:  Richard Winkelmann; Peter Börnert; Jan De Becker; Romhild Hoogeveen; Peter Mazurkewitz; Olaf Dössel
Journal:  MAGMA       Date:  2006-11-24       Impact factor: 2.310

6.  Coronary vessel-wall and lumen imaging using radial k-space acquisition with MRI at 3 Tesla.

Authors:  Andrew N Priest; P Martin Bansmann; Kai Müllerleile; Gerhard Adam
Journal:  Eur Radiol       Date:  2006-10-05       Impact factor: 5.315

7.  Facilitated acquisition of whole-heart coronary magnetic resonance angiography with visual feedback of respiration status.

Authors:  Tomohisa Okada; Shigehide Kuhara; Shotaro Kanao; Ayako Ninomiya; Saori Sato; Toshikazu Kamae; Kimio Gotoh; Kaori Togashi
Journal:  Int J Cardiovasc Imaging       Date:  2008-12-16       Impact factor: 2.357

8.  Two-dimensional coronary MRA: limitations and artifacts.

Authors:  A J Duerinckx; D P Atkinson; J Mintorovitch; O P Simonetti; M K Vrman
Journal:  Eur Radiol       Date:  1996       Impact factor: 5.315

9.  Correction of breathing-induced errors in magnetic resonance thermometry of hyperthermia using multiecho field fitting techniques.

Authors:  Cory R Wyatt; Brian J Soher; James R MacFall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

10.  Feasibility and diagnostic accuracy of whole heart coronary MR angiography using free-breathing 3D balanced turbo-field-echo with SENSE and the half-fourier acquisition technique.

Authors:  Young Jin Kim; Jae-Seung Seo; Byoung Wook Choi; Kyu Ok Choe; Yangsoo Jang; Young-Guk Ko
Journal:  Korean J Radiol       Date:  2006 Oct-Dec       Impact factor: 3.500

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.