OBJECTIVE: To evaluate the accuracy and precision of HemoCue 201 (HemoCue) and Masimo Pronto 7 (Masimo) devices for measuring hemoglobin (Hb) in epidemiological studies, having venous blood samples as a gold standard.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28170445 PMCID: PMC5295714 DOI: 10.1371/journal.pone.0170990
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Descriptive statistics for repeated Masimo Hb measurements.
| Measurement | Mean | Std. Dev. | 95% CI |
|---|---|---|---|
| 1st | 12.65 | 1.21 | 12.5, 12.8 |
| 2nd | 12.65 | 1.15 | 12.5, 12.8 |
| 3rd | 12.64 | 1.03 | 12.5, 12.8 |
Mean and confidence intervals of Hb by measuring procedure.
| Device | Hb g/dL | 95% CI |
|---|---|---|
| HemoCue 201 | 12.7 | 12.5, 12.9 |
| Masimo Pronto 7 | 12.6 | 12.4, 12.8 |
| Gold Standard | 14.2 | 14.0, 14.5 |
*Significant t-test results in comparison of devices against the Gold Standard (p<0.001).
Linear Regression models for predicting Hb (Gold Standard) using HemoCue 201 and Masimo Pronto 7 as predictors.
| Regression model | β | Test for β = 1 F (1, 136) | P | R-squared |
|---|---|---|---|---|
| HemoCue 201 vs. Gold Stadard | 0.887 | 168.28 | 0.000 | 0.987 |
| Masimo Pronto 7 vs. Gold Stadard | 0.876 | 192.63 | 0.000 | 0.986 |
Concordance Correlation Coefficient and Bland-Altman Limits of Agreement for HemoCue 201 and Masimo Pronto 7 as compared to the Gold Standard.
| Device | Concordance Correlation | Pr(r = 0) | Precision | Accuracy |
|---|---|---|---|---|
| HemoCue 201 | 0.183 | 0.000 | 0.313 | 0.586 |
| Masimo Pronto 7 | 0.166 | 0.000 | 0.299 | 0.556 |
| Device vs. Gold Standard | Difference | Std. Dev. | 95% Limits Of Agreement | |
| (Bland & Altman, 1986) | ||||
| HemoCue 201 | -1.51 | 1.58 | -4.60 | 1.58 |
| Masimo Pronto 7 | -1.67 | 1.62 | -4.84 | 1.51 |
Variance components for the Masimo Pronto 7 Hb measurements.
| Hierarchical level | Variance component | Percentage |
|---|---|---|
| Masimo devices | 0.193 | 16.28% |
| Subjects | 0.322 | 25.81% |
| Within subject | 0.724 | 57.91% |
| Total | 1.239 | 100% |