Accelerating the washout of inhalational anesthetics from the Dräger Primus anesthetic workstation: effect of exchangeable internal components.

Mark W Crawford1, Heike Prinzhausen, Guy C Petroz.   

Abstract

BACKGROUND: To establish guidelines for the preparation of the Primus anesthetic workstation (Dräger, Lübeck, Germany) for malignant hyperthermia-susceptible patients, the authors evaluated the effect of replacing the workstation's exchangeable internal components on the washout of isoflurane.
METHODS: Primus workstations were exposed to isoflurane, and contaminated internal components were replaced as follows: group 1, no replacement; group 2, new ventilator diaphragm; group 3, autoclaved ventilator diaphragm; group 4, autoclaved integrated breathing system; group 5, flushed integrated breathing system; group 6, autoclaved ventilator diaphragm and integrated breathing system. The fresh gas flow was set at 10 l/min, and subsequently reduced to 3 l/min when a concentration of 5 ppm was achieved. Isoflurane concentration was measured in the inspiratory limb of the circle circuit every minute.
RESULTS: Washout times for isoflurane decreased in the following order: group 1 (67 +/- 6.5 min) > groups 2 and 3 (50 +/- 4.1 and 50 +/- 5.7 min, respectively) > group 5 (43 +/- 9.5 min) > group 4 (12 +/- 1.5 min) > group 6 (3.2 +/- 0.4 min). Isoflurane concentration increased approximately threefold when the fresh gas flow was reduced to 3 l/min.
CONCLUSION: Washout of isoflurane increased 20-fold with the use of an autoclaved ventilator diaphragm and integrated breathing system. To prepare the Primus for malignant hyperthermia-susceptible patients, the authors recommend replacing the ventilator diaphragm and integrated breathing system with autoclaved components, flushing the workstation for 5 min at a fresh gas flow of 10 l/min, and maintaining this flow for the duration of anesthesia.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17264723     DOI: 10.1097/00000542-200702000-00017

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  2 in total

Review 1.  Management of malignant hyperthermia: diagnosis and treatment.

Authors:  Daniel Schneiderbanger; Stephan Johannsen; Norbert Roewer; Frank Schuster
Journal:  Ther Clin Risk Manag       Date:  2014-05-14       Impact factor: 2.423

2.  The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration.

Authors:  Hye Won Shin; Hae Na Yu; Go Eun Bae; Hyub Huh; Ji Yong Park; Ji Young Kim
Journal:  BMC Anesthesiol       Date:  2017-01-19       Impact factor: 2.217

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.