Adverse events with medical devices in anesthesia and intensive care unit patients recorded in the French safety database in 2005-2006.

Laurent Beydon1, Pierre Yves Ledenmat, Christophe Soltner, Frédéric Lebreton, Vincent Hardin, Dan Benhamou, François Clergue, Gérard Laguenie.   

Abstract

BACKGROUND: French regulations require that adverse events involving medical devices be reported to the national healthcare safety agency. The authors evaluated reports made in 2005-2006 for patients in anesthesiology and critical care.
METHODS: For each type of device, the authors recorded the severity and cause of the event and the manufacturer's response where relevant. The authors compared the results with those obtained previously from the reports (n = 1,004) sent in 1998 to the same database.
RESULTS: The authors identified 4,188 events, of which 91% were minor, 7% severe, and 2% fatal. The cause was available for 1,935 events (46%). Faulty manufacturing was the main cause of minor events. Inappropriate use was the cause in a significantly larger proportion of severe events than minor events (P < 0.001) and was usually considered preventable via improved knowledge or device verification before use. Compared to with that in 1998, the annual number of reported events doubled and the rate of severe events decreased slightly (12-10%, P = 0.03). The rate of events related to manufacturing problems remained stable (59-60%, P = nonsignificant), and the rate of events caused by human errors was 32-42% (P = 0.01). There were no changes in the mortality rate (2% in both studies).
CONCLUSIONS: The number of adverse events related to medical devices indicates a need for greater attention to these complex pieces of equipment that can suffer from faulty design and manufacturing and from inappropriate use. Improvements in clinician knowledge of medical devices, and to a lesser extent improvement in manufacturing practices, should improve safety.

Entities:  

Mesh:

Year:  2010        PMID: 20098139     DOI: 10.1097/ALN.0b013e3181ca2e55

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  4 in total

1.  Analysis of event logs from syringe pumps: a retrospective pilot study to assess possible effects of syringe pumps on safety in a university hospital critical care unit in Germany.

Authors:  Marc Kastrup; Felix Balzer; Thomas Volk; Claudia Spies
Journal:  Drug Saf       Date:  2012-07-01       Impact factor: 5.606

2.  Effects and Satisfaction of Medical Device Safety Information Reporting System Using Electronic Medical Record.

Authors:  Hye Jung Jang; Young Deuk Choi; Nam Hyun Kim
Journal:  Healthc Inform Res       Date:  2017-04-30

3.  The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration.

Authors:  Hye Won Shin; Hae Na Yu; Go Eun Bae; Hyub Huh; Ji Yong Park; Ji Young Kim
Journal:  BMC Anesthesiol       Date:  2017-01-19       Impact factor: 2.217

4.  Thinking outside the (acrylic) box: a framework for the local use of custom-made medical devices.

Authors:  M C Turner; L V Duggan; B A Glezerson; S D Marshall
Journal:  Anaesthesia       Date:  2020-06-15       Impact factor: 12.893

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.