Martin Reinhardt1,2, Paolo Piaggi1, Barbara DeMers1, Cathy Trinidad1, Jonathan Krakoff1. 1. Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch, National Institute of Diabetes and Digestive and Kidney Diseases, National Institutes of Health, Phoenix, Arizona, USA. 2. Department of Diagnostic and Interventional Radiology, University of Leipzig, Germany.
Abstract
OBJECTIVE: Compare total percentage body fat (pfat) measurements between two densitometers and visceral adipose tissue (VAT) estimates between iDXA and magnetic resonance imaging (MRI) from the same defined abdominal region. METHODS: Participants [N = 93 (50 men, 43 women), BMI: 19.1-57.6 kg/m2 ] underwent dual-energy X-ray absorptiometry (DXA) scans on two DXA systems (GE Healthcare Lunar iDXA and Lunar Prodigy), and a subgroup underwent abdominal MRI imaging for quantification of VAT. RESULTS: Pfat correlated strongly between both machines (r2 = 0.98, P < 1.0E-14). Bland-Altman plots showed a bias with higher measured pfat on iDXA versus Prodigy in leaner subjects and the opposite in more overweight subjects. The R2 for regression of MRI on iDXA VAT values was 0.948. Bland-Altman bias was +104.1 cm3 with 95% limits of agreement of -681.9 to 890.0 cm3 . For both DXA methods, and iDXA versus MRI determined VAT, comparison using rank regression demonstrated no order bias. CONCLUSIONS: The total pfat measured by both machines was strongly and linearly associated, allowing for conversion (equations are provided) of iDXA for assessment of longitudinal body fat changes. Despite a bias of abdominal VAT measures of iDXA versus MRI, the high rank correlation makes iDXA a good alternative to the more complicated and time-consuming MRI for use in larger cross-sectional and longitudinal studies. Published 2016. This article is a U.S. Government work and is in the public domain in the USA. Obesity published by Wiley Periodicals, Inc. on behalf of The Obesity Society (TOS).
OBJECTIVE: Compare total percentage body fat (pfat) measurements between two densitometers and visceral adipose tissue (VAT) estimates between iDXA and magnetic resonance imaging (MRI) from the same defined abdominal region. METHODS:Participants [N = 93 (50 men, 43 women), BMI: 19.1-57.6 kg/m2 ] underwent dual-energy X-ray absorptiometry (DXA) scans on two DXA systems (GE Healthcare Lunar iDXA and Lunar Prodigy), and a subgroup underwent abdominal MRI imaging for quantification of VAT. RESULTS: Pfat correlated strongly between both machines (r2 = 0.98, P < 1.0E-14). Bland-Altman plots showed a bias with higher measured pfat on iDXA versus Prodigy in leaner subjects and the opposite in more overweight subjects. The R2 for regression of MRI on iDXA VAT values was 0.948. Bland-Altman bias was +104.1 cm3 with 95% limits of agreement of -681.9 to 890.0 cm3 . For both DXA methods, and iDXA versus MRI determined VAT, comparison using rank regression demonstrated no order bias. CONCLUSIONS: The total pfat measured by both machines was strongly and linearly associated, allowing for conversion (equations are provided) of iDXA for assessment of longitudinal body fat changes. Despite a bias of abdominal VAT measures of iDXA versus MRI, the high rank correlation makes iDXA a good alternative to the more complicated and time-consuming MRI for use in larger cross-sectional and longitudinal studies. Published 2016. This article is a U.S. Government work and is in the public domain in the USA. Obesity published by Wiley Periodicals, Inc. on behalf of The Obesity Society (TOS).
Dual-energy X-ray absorptiometry (DXA) provides accurate measurement of body
composition, which is essential for the quantification of adipose and lean tissue,
and their distribution during and after interventions in energy balance studies. DXA
systems have replaced underwater weighing as a method to determine body composition
in nearly all clinical- and research facilities. Technical innovations led to a
transition from the early pencil-beam to modern fan-beam densitometers, which
utilize multiple detectors in favor of faster scan times and near-radiographic image
quality (1). Longitudinal studies with
repeated body composition measurements require comparable assessments of the
percentage body fat (pfat), fat mass, and fat free mass (FM, FFM). When baseline-
and follow-up body composition measurements are not performed with the same DXA
machine, e.g. due to system upgrades, cross-calibration studies are necessary to
compare results from different DXA systems. Recent cross-calibration studies
comparing precision of body composition measurements between different fan-beam
systems report excellent (2) or good agreement
(3) and propose the use of translational
equations to ensure comparability of body composition measurements (3).The recently introduced fan-beam iDXA (GE Healthcare Lunar) densitometer
allows, for the first time, fully automated dedicated measurement of visceral
adipose tissue (VAT), separately form subcutaneous adipose tissue (SAT). Adipose
tissue distribution (in particular VAT vs. SAT) rather than total adiposity has been
associated with adverse health events (4–7). Previous reports on
the precision of VAT measurements done by iDXA in comparison to the time- and
resource intense gold standard methods computed tomography (CT), and magnetic
resonance imaging (MRI) report a strong correlation of VAT measured with the
respective imaging modalities throughout a wide range of body sizes and different
ethnicities (8–10). However, to our knowledge previous reports compared a
fixed region of interest from CT or MRI scans, defined as 150 mm of the abdomen,
beginning at the top of S1 and moving toward the head (8–10), while the
CoreScan® software (iDXA) individually defines a region-of-interest whose
caudal limit is the top of the iliac crest and in which the height is set to
20% of the distance from the top of the iliac crest to the base of the
mandible to account for inter-individual differences in body size.The aim of this study was to precisely compare total percentage body fat
measurements between two fan beam densitometers (GE Lunar Prodigy and GE Lunar
iDXA), and VAT estimates between iDXA and MRI from the same defined abdominal
region.
Methods
Subjects
We recruited 100 subjects without reported health problems (50 women and
50 men, aged 18.6 to 64.7 years) from the greater Phoenix area with a broad
range of BMI (19.1 – 57.6 kg/m2), who underwent DXA scans on
both machines (Lunar Prodigy and iDXA), in random order. The only exclusion
criteria were age <18 years, pregnancy or breastfeeding or weight
>159 kg (weight limit for Lunar Prodigy). By design, we recruited
approximately equal numbers of women and men in the following BMI groups
< 25 kg/m2 (normal weight; 16 women, 16 men), ≥ 25
kg/m2 <30 kg/m2 (overweight; 16 women, 18
men), and ≥30kg/m2 (obese; 18 women, 16 men). Seven
participants (all with BMI >30) were excluded from the analyses. These 7
scans were performed as half scans of the right side of the body, previously
validated for calculation of total body fat on the same scanner (14). A limitation of available DXA systems
is the size of the scanning area (~190 × 60 cm), meaning the
body dimensions can exceed the scanning area. These 7 participants had been
improperly positioned on the table during one of the scans. Thus, in the results
section we report results of 93 participants (Table 1), whose demographic and anthropometric characteristics did
not differ significantly from the initially recruited 100 subjects.
Table 1
Demographic and anthropometric characteristics of the study
population.
Total study cohort with both DXA
scans
Subgroup with both DXA scans,
andabdominal MRI scan
Total(N=93)
Women(N=43)
Men(N=50)
Total (N=40)
Women(N=20)
Men(N=20)
Race
B=1, H=2,NA=4
B=1, H=1,NA=3
B=7, H=8,NA=16,W=19
A=1,
B=5,H=5,NA=15,W=14
A=1,
B=2,H=1,NA=10,W=6
B=3, H=4,NA=5,W=8
Age (years)
38,6 ± 11.6
37.5 ± 11.4
39.7 ± 11.6
36.6 ± 10.7
37.4 ± 11.5
35.8 ± 9.9
Height (cm)
168,6 ± 9.6
161.2 ± 6.5***
175.3 ± 6.3***
168.3 ± 9.1
161.9 ± 7.3
174.7 ± 5.3***
Body weight(kg)
79,1 ± 17.7
73.4 ± 18.4**
86.4 ± 17.4*
82.1 ± 20.6
76.9 ± 21.7
87.3 ± 18.6
BMI (kg/m2)
27,8 ± 5.7
28.2 ± 6.5
28.2 ± 5.9
29.1 ± 7.5
29.5 ± 8.8
28.6 ± 6.1
pfat Lunar DXA(%)
34,2 ± 11.0
41.5 ± 8.5***
28.2 ± 9.2***
35.1 ± 12.0
42.9 ± 8.8
27.3 ± 9.5***
pfat iDXA (%)
34,1 ± 10.0
40.4 ± 8.3***
29.3 ± 8.8***
35.1 ± 11.3
41.7 ± 9.3
28.5 ± 9.1***
FM Lunar DXA(kg)
27,8 ± 12.5
31.6 ± 13.9**
25.6 ± 12.4*
29.8 ± 15.2
34.3 ± 16.0
25.2 ± 13.2
FM iDXA (kg)
27,7 ± 12.5
30.8 ± 13.8*
26.5 ± 12.7*
30.0 ± 15.9
33.6 ± 17.3
26.3 ± 13.7
FFM LunarDXA (kg)
51,3 ± 11.4
41.8 ± 6.3***
60.8 ± 8.0***
52.4 ± 12.4
42.6 ± 7.1
62.1 ± 8.0***
FFM iDXA (kg)
51,4 ± 10.8
42.5 ± 6.1***
59.9 ± 7.5***
52.2 ± 11.1
43.3 ± 6.0
61.1 ± 7.1***
VAT iDXA(cm3)
.
.
.
1210 ± 1245
997 ± 948
1423 ± 1480
VAT MRI (cm3)
.
.
.
1106 ± 930
893 ± 584
1318 ± 1157
Shown is the total study population N=93, and N=40 with both DXA
scans, and abdominal MRI scan. Values are presented as mean ±
SD.
Differences between men and women were assessed by two sample
Student’s t-test (* = P<0.05; ** = <0.01; *** =
P<0.0001.
Abbreviations: A = Asian; B = Black; H = Hispanic; NA = Native
American; W = White; pfat = percentage body fat; BMI = Body mass index; FM =
Fat mass; FFM = Fat free mass
From the initial 100 individuals, forty (20 women and 20 men) underwent
additional abdominal magnetic resonance imaging (< 25 kg/m2
(6 women, 6 men), ≥ 25 kg/m2–<30
kg/m2 (7 women, 7 men), and ≥30kg/m2 (7 women,
7 men)). Female participants underwent a urine pregnancy test within 1 hour of
the DXA scans. Volunteers were fully informed of the nature and purpose of the
study they participated in, and written informed consent was obtained before
admission. DXA measurements were done at the clinical research unit of the
Obesity and Diabetes Clinical Research Section of the National Institute of
Diabetes and Digestive and Kidney Diseases (NIDDK) in Phoenix, AZ. MRIs were
performed at St. Joseph’s Hospital and Medical Center at the Kelly
Center for Imaging Innovation, Phoenix, AZ. The protocol was approved by the
Institutional Review Board of the NIDDK and by the St. Joseph’s Hospital
and Medical Center at the Kelly Center for Imaging Innovation’s
Institutional Review Board.
Dual-energy X-ray absorptiometry
All total body DXA scans were performed in the morning after an
overnight fast (no caloric intake after 8pm the night before the scans).
Consecutive whole body composition measurements were performed in random order
on two fan beam DXA machines – one being the Lunar Prodigy (enCORE 2003
software version 7.53.002, GE Lunar, Madison, WI, USA) and the second being the
iDXA (enCORE version 15, GE Lunar Healthcare). Subjects were scanned using
standard imaging and positioning protocols. With both DXA systems, the operator
remained in the room with the subject during the scan. For measuring VAT, a
region of interest was automatically defined by the iDXA software algorithm,
with the caudal limit at the top of the iliac crest and the height set to
20% of the distance from the top of the iliac crest to the base of the
mandible. FM data (gram) from iDXA were transformed into MRI adipose tissue
volume using a constant correction factor (0.94 g/cm3), assuming a
constant density for adipose tissue (11).
Magnetic resonance imaging
Abdominal imaging was done with a 3.0T Philips Ingenia scanner (Philips,
Amsterdam, Netherlands). Volunteers were examined in a supine position, using 2
dS anterior coils and a dS posterior coil. Images were obtained in transverse
slice orientation (two-point mDixon sequence; 8 stacks containing 8–12
slices each; thickness 10 mm; and 0.1 mm interslice gap; TR shortest (3,7ms); TE
1.4 ms; flip angle 10°). VAT volume as measured by MRI was determined
according to an individually sized regions of interest (ROI) based on the
respective VAT ROI estimates from the automated iDXA software algorithm (see
Subjects and Methods/ Dual-energy X-ray absorptiometry) and
were quantified using OsiriX image processing software (version 6.5.2). All MRI
– VAT volumes were quantified twice in random order by a single
radiologist blinded to participant data (CV 2.8% ± 3.1
(SD)).
Sample size calculation
Based on previous studies (3,
12) comparing the Lunar Prodigy and
iDXA, using total FM of 23027 ± 10167 grams for Prodigy and 24351
± 9543 grams for iDXA with a correlation coefficient of 0.97 between the
measurements at a 2-sided alpha of 0.05, we calculated that 50 participants
would give us a power of 0.94 to detect a difference in pfat. An additional 50
subjects were recruited to validate the prediction equation calculated in the
first 50 subjects (see below).Previous studies of VAT comparisons using CT scans and iDXA demonstrated
an r-squared for the univariate model of DXA visceral fat regressed on CT
visceral fat equal to 0.96. For our power calculation, we used a more
conservative r-squared of 0.80 for the univariate model of MRI visceral fat with
iDXA visceral fat. Thus, 40 participants would give us power of >0.9 to
detect a significant association between these two measurements.
Statistical Analysis
To compare body composition measurements between the GE Lunar Prodigy
DXA and GE Lunar iDXA systems on the same individuals, linear regression models
and paired Student’s t-test were used for each of the main variables of
interest (pfat, FM, and FFM). An equation was used to convert the pfat values,
measured by iDXA on the first n=50 (n=47, please see Methods/Subjects section
for clarification) individuals, to values measured by Lunar Prodigy in the same
individuals. The results of this prediction were then validated against the
measurements of the second n=50 (n=46) individuals. To compare the measured pfat
values of the second 46 subjects (iDXA) to the predicted Lunar Prodigy values
obtained by using the regression equation calculated from the first 47
subjects’ values, we calculated the difference between measured minus
predicted pfat values for each subject and then tested these differences against
0, using one-sample Student’s t-tests.We used linear regression models to assess the association between pfat
values acquired with the Lunar Prodigy versus iDXA, as well as between visceral
fat measured by iDXA and MRI. For both comparisons, we assessed the agreement
between different imaging techniques and evaluated any systematic (intercept) or
proportional (slope) bias by two methods using: Bland-Altman diagrams and the
95% limits of agreement procedure (13), and the non-parametric Passing-Bablok regression (14) which is more robust to the presence of
outliers. The Cusum test was used to evaluate the linearity of the relationship.
Lastly, the coefficient of variation (CV), the intra-class correlation (ICC) and
the Spearman's rank-order correlation (rho) between the measures of body
fat on each machine were also calculated.
Results
Comparison analysis Lunar Prodigy DXA vs. iDXA
Demographic characteristics of the study population are presented in
Table 1.In the first 47 subjects, over a wide range of adiposity we found a
strong correlation between the pfat values measured on both machines
(r2=0.98, P<1.0E-14; Fig.
1A). We found an equally strong correlation when we used the
regression equation obtained from the first 47 subjects (Lunar Prodigy DXA pfat
= −3.417 + 1.099 × iDXA pfat) to calculate the predicted Lunar
Prodigy DXA pfat values in the second cohort and compare them to measured values
(r2=0.98, P<1.0E-14; Fig.1B). Furthermore, there was an equally strong correlation
between the measured values of the second 46 subjects (r2=0.98,
P<1.0E-14; Fig. 1C). The mean
difference between the predicted Lunar Prodigy- versus measured (iDXA) pfat
values was not significantly different from 0 (mean pfat difference =
0.27%, 95% confidence interval: −0.15 to 0.70, P=0.21).
The mean absolute error (MAE) of the pfat prediction was 1.7%, ranging
from 0.1% to 9.8%. The ICC for all 93 pfat measurements
performed on both machines was 0.986 (95% CI: 0.979 to 0.991), the CV
was 3.6%. In the entire group (n=93), pfat, FM, and FFM measurements
were not significantly different between the two systems (paired t-test: P=0.59;
P=0.58; P=0.58, respectively). The Bland-Altman plots (Fig 2A–B) demonstrated a bias with higher measured
pfat on iDXA versus Lunar Prodigy DXA in leaner subjects and the opposite in
more overweight subjects. As mentioned above, 7 participants with improperly
positioned half-right body scans were excluded from all analyses. To assess the
impact of excluding those half scans, we performed a sensitivity analysis with
and without these 7 scans included, and then without all half-scans. Compared to
the correlation between the two scans calculated in the whole cohort (N=100,
r2=0.95, P<1.0E-14), excluding these scans did produce a
marginal improvement in the correlation between the two DXA systems (N=93,
r2=0.98, P<1.0E-14; N=77, r2=0.98,
P<1.0E-14). For the half scans only (n=23), the ICC was 0.89
(95% CI: 0.76 to 0.95), and the CV was 6.5%. After removal of
all 23 half-right scans, the Bland-Altman bias was more pronounced (Fig. 2A, and B). Linearity of relationships
between methods was confirmed by the Cusum test for all N=100, N=93, and N=77
(P=0.70, P=0.64, P=0.71, Fig.3 A–B,
respectively). Preservation of individual ranks was also confirmed by the
Spearman's correlation (n=100: rho=0.98; n=77: rho=0.99, both
P<0.0001). Non-parametric Passing-Bablok regression confirmed the bias
seen in the Bland-Altman analyses with a significant proportional difference
(slope=3.0, 95% CI: 2.24 to 3.85, N=93) and systematic difference
(intercept=0.90, 95% CI: 0.88 to 0.93, N=93) between the two
methods.
Figure 1
Correlation between Lunar Prodigy and iDXA total body percentage body fat
measurements
Figures 1A shows the correlation between pfat measured with the Lunar
Prodigy and iDXA in the first 47 men and women (r2=0.98, P<1.0E-14).
Figures 1B shows the predicted n=46 Lunar Prodigy pfat values based on iDXA
measurements versus the actual measures for the second cohort (r2=0.98,
P<1.0E-14, Lunar Prodigy = −3.417 + 1.099 × iDXA), and
Figure 1C shows the correlation of second cohort (n=46) subjects’
percentage pfat measured by lunar DXA and iDXA (r2=0.98, P<1.0E-14;
Lunar Prodigy = −2.506 + 1.080 × iDXA).
Figure 2
Bland-Altman plots
Bland-Altman plots showing the limits of agreement (mean and 95%
confidence interval) between paired values for pfat measured with the Lunar
Prodigy and Lunar iDXA for all n=93 subjects (A), and after exclusion of all 23
half-right scans (B).
Figure 3
Residual diagrams
Residual diagrams showing pfat values measured with the Lunar Prodigy
and Lunar iDXA for all n=93 subjects (A), and after exclusion of all 23
half-right scans (B). Values are presented for men (squares) and women (circles)
around the mean (solid line).
VAT measured with MRI vs. iDXA
Demographic characteristics of the 20 men and 20 women who also
underwent a MRI scan are presented in table
1.The coefficient of determination (r2) for regression of MRI
on iDXA values was 0.958 for women, 0.959 for men, and 0.948 combined (Fig.4A). Bland-Altman bias was +103.6
cm3 for women and +104.6 cm3 for men. The 95%
limits of agreement were −669.9 to +877.0 cm3 for women and
−713.7 to +923.0 cm3 for men. Combined Bland-Altman bias was
+104.1 cm3 with 95% limits of agreement of −681.9 to
890.0 cm3 (Fig.4B). Linearity of
relationships between MRI and iDXA measurements of VAT was confirmed by the
Cusum test (P=0.97). A significant proportional bias (slope=1.40, 95%
CI: 1.28 to 1.57, n=40) between the two methods was confirmed by the
Passing-Bablok non-parametric test (Fig.4C). A systematic difference over the total range of values
(intercept) was estimated as 287 cm3 (95% CI: −442 to
−159). The ICC of the VAT measurements performed by both methods was
0.93 (95% CI: 0.87 to 0.96), the CV was 25.0%.
Figure 4
Comparison of VAT mesaurement between iDXA and MRI methods
Figure 4A shows the correlation between VAT measured with iDXA and MRI
(r2=0.95, P<1.0E-14, N=40, VAT (iDXA) = −232.5 + 1.304 ×
VAT (MRI)). The Bland-Altman plot (B) shows the limits of agreement (mean and
95% confidence interval) between paired values for VAT measured with
iDXA and MRI. The residual diagram (C) shows VAT volume values measured with the
Lunar iDXA for all n=40 subjects. Values are presented for men (squares) and
women (circles) around the mean (solid line).
Discussion
This study compared body composition from two fan-beam DXA systems and VAT
volume of iDXA versus MRI. While the overall correlation between percentage body fat
values measured on the two different machines was very high in the 93 subjects over
a wide BMI range, we did find a trend towards an overestimation of pfat measured
with iDXA compared to Lunar Prodigy in leaner individuals and a slight
underestimation of pfat in those at the higher ranges of adiposity. We also found
that VAT measured by iDXA showed a strong overall correlation with MRI but was
slightly underestimated for lower VAT-volumes and overestimated for higher volumes,
compared to MRI. For both methods comparison using rank regression did not
demonstrate a bias in terms of order, indicated by the given linearity of the
respective relationships.Given the reported differences in pfat values measured by the two systems
and in agreement with previous works (3, 12, 17)
we recommend the employment of conversion equations derived from cross-calibration
studies like ours. This procedure becomes crucial when upgrading DXA units during
ongoing longitudinal data collection, e.g. in order to observe and quantify body
composition changes during long-term weight gain or weight loss trials. However,
albeit statistically significant, the small discrepancy between the two methods
(pFAT error <1%) is not clinically relevant. The main component of
the metabolic syndrome - visceral adipose tissue – seems to be strongly
associated with cardio-metabolic risk factors, due to its metabolic activity (15). Furthermore, VAT has been shown to be
predictive of all-cause mortality (16).
Hence, precise estimation of VAT volume in both, cross sectional and longitudinal
settings might be one of the important parameters in monitoring obesity.With a population based overall rank correlation of r2=0.948,
iDXA´s fully automated VAT estimation and MRI derived values are in strong
agreement and consistent with previous studies (8, 10). However, we found an
underestimation in leaner subjects and overestimation of VAT in subjects with
greater adiposity in measuring VAT using iDXA versus MRI. This differs from prior
studies, that found no significant BMI dependent variability (8), or the opposite directionality in the bias for measuring VAT
(10). However, these studies compared
body size-adjusted iDXA-VAT estimates with an arbitrarily standardized CT or MRI
region of interest that was defined as 150 mm of the abdomen, starting at the top of
S1 and moving toward the head. As noted in our methods, we were able to compare the
same ROI designated by the iDXA software to the MRI. Thus, the strength of our study
is that we compared that same estimated abdominal VAT from the same region of
interest for both the iDXA and MRI. Given the demonstrated, clinically significant,
systematic difference over the total range of VAT values between iDXA and the gold
standard method MRI, we propose that iDXA-VAT measurements might be useful for
relative VAT assessment in longitudinal studies, rather than for accurate absolute
measurements.This study did not include subjects under the age of 18 and over the age of
65, which limits the interpretation of our findings for these age groups. The
conversion equations provided are tailored to our specific (though ethnically
diverse and representative of a wide BMI range) study population, and might however
not be perfectly suitable for any other study population. In addition, the half-body
DXA scans used for some of the more obese subjects resulted in a greater difference
between the measurements. The half scan has been validated on the same machine but
not between machines. Given the automated algorithm for calculation of pfat, a
slight difference in alignment on the table might account for these errors. We
performed several sensitivity analyses to assess the impact of these scans, and
still found an excellent agreement between the measurements.In conclusion, the total percentage body fat measured by iDXA and Lunar
Prodigy are strongly, and linearly associated and allow for conversion of iDXA for
assessment of longitudinal body fat changes. Despite the under- and overestimation
of pfat in subjects with very low and very high BMIs, the high rank correlation of
abdominal visceral adipose tissue measures of iDXA with MRI which is considered the
gold standard for this measurement makes iDXA a reliable, cost- and time effective
alternative to MRI for use in cross-sectional and longitudinal studies.
Authors: Sanjiv Kaul; Megan P Rothney; Dawn M Peters; Wynn K Wacker; Cynthia E Davis; Michael D Shapiro; David L Ergun Journal: Obesity (Silver Spring) Date: 2012-01-26 Impact factor: 5.002
Authors: Janet-Mae P Soriano; Elli Ioannidou; Jack Wang; John C Thornton; Mary N Horlick; Dympna Gallagher; Steven B Heymsfield; Richard N Pierson Journal: J Clin Densitom Date: 2004 Impact factor: 2.963
Authors: M Negro; S Perna; D Spadaccini; L Castelli; L Calanni; M Barbero; C Cescon; M Rondanelli; G D'Antona Journal: J Nutr Health Aging Date: 2019 Impact factor: 4.075
Authors: Susanne B Votruba; Pamela A Shaw; Eric J Oh; Colleen A Venti; Susan Bonfiglio; Jonathan Krakoff; Diane M O'Brien Journal: Am J Clin Nutr Date: 2019-12-01 Impact factor: 7.045
Authors: Jessica J Johnson; Pamela A Shaw; Matthew J Wooller; Colleen A Venti; Jonathan Krakoff; Susanne B Votruba; Diane M O'Brien Journal: J Nutr Date: 2022-09-06 Impact factor: 4.687
Authors: Cassie M Mitchell; Paolo Piaggi; Diane M O'Brien; Jonathan Krakoff; Susanne B Votruba Journal: Obesity (Silver Spring) Date: 2021-05-03 Impact factor: 9.298
Authors: Alessio Basolo; Mujtaba H Shah; Varsha Parthasarathy; Shannon Parrington; Mary Walter; Susanne B Votruba; Jonathan Krakoff; Paolo Piaggi; Douglas C Chang Journal: Obesity (Silver Spring) Date: 2020-04-30 Impact factor: 9.298
Authors: Jessica J Johnson; Pamela A Shaw; Eric J Oh; Matthew J Wooller; Sean Merriman; Hee Young Yun; Thomas Larsen; Jonathan Krakoff; Susanne B Votruba; Diane M O'Brien Journal: Am J Clin Nutr Date: 2021-05-08 Impact factor: 7.045