Tatsuya Kawai1, Yoshiyuki Ozawa1, Masaki Ogawa1, Kazuya Ohashi2, Kazushi Suzuki1, Yuta Shibamoto1.
Abstract
BACKGROUND: The effects of the reduction of low-energy X-ray spectrum on lung perfusion images created by dual-energy CT have not been well evaluated. The aim of this study is to investigate the reliability of lung perfusion blood volume (PBV) images created by dual-energy CT (DECT) equipped with or without a tin filter, focusing on its accuracy adjacent to high-attenuation areas. MATERIAL/Entities:
Keywords: Perfusion Imaging; Radiography, Dual-Energy Scanned Projection; Tomography Scanners, X-Ray Computed
Year: 2016 PMID: 27994698 PMCID: PMC5154711 DOI: 10.12659/PJR.899502
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Figure 1Schema for patient selection.
Figure 2Scoring criteria of pulmonary enhancement defect (PED) (A) and tracheal pseudo-enhancement (TPE) (B). For PED, score 0 was defined as no or very slight defect, score 3 was broad defect which might be judged as pulmonary embolism. TPE was classified into 4 grades dependent on its area in the cross-sectional trachea.
Summary of the results.
| Group A (n=18) | Group B (n=20) | ||
|---|---|---|---|
| Age (years, mean) | 62 | 66 | 0.42 |
| Gender (F/M) | 12/6 | 17/3 | 0.18 |
| Injection side (R/L) | 15/3 | 19/1 | 0.24 |
| Attenuation of BCV on WAI | 1165±357 | 1424±401 | 0.09 |
| PED score (mean) | 0.8 | 1.9 | <0.0001 |
| TPE score (mean) | 1.1 | 2.2 | <0.0001 |
| CTDI volume (mGy, mean ±SD) | 4.90±1.14 | 12.98±3.15 | <0.0001 |
WAI – weighted average image (a mixed two-kilovolt image equivalent to 120-kVp image).
Student’s t-test;
Chi-square test;
Mann-Whitney u test.
Figure 3Correlations between the artifact score and attenuation of the brachiocephalic vein ipsilateral to the injection site of the contrast medium. (A) Group A and (B) Group B. A positive correlation was observed for PED in Group A (P=0.038).