Empirical beam hardening correction (EBHC) for CT.

Yiannis Kyriakou1, Esther Meyer, Daniel Prell, Marc Kachelriess.   

Abstract

PURPOSE: Due to x-ray beam polychromaticity and scattered radiation, attenuation measurements tend to be underestimated. Cupping and beam hardening artifacts become apparent in the reconstructed CT images. If only one material such as water, for example, is present, these artifacts can be reduced by precorrecting the rawdata. Higher order beam hardening artifacts, as they result when a mixture of materials such as water and bone, or water and bone and iodine is present, require an iterative beam hardening correction where the image is segmented into different materials and those are forward projected to obtain new rawdata. Typically, the forward projection must correctly model the beam polychromaticity and account for all physical effects, including the energy dependence of the assumed materials in the patient, the detector response, and others. We propose a new algorithm that does not require any knowledge about spectra or attenuation coefficients and that does not need to be calibrated. The proposed method corrects beam hardening in single energy CT data.
METHODS: The only a priori knowledge entering EBHC is the segmentation of the object into different materials. Materials other than water are segmented from the original image, e.g., by using simple thresholding. Then, a (monochromatic) forward projection of these other materials is performed. The measured rawdata and the forward projected material-specific rawdata are monomially combined (e.g., multiplied or squared) and reconstructed to yield a set of correction volumes. These are then linearly combined and added to the original volume. The combination weights are determined to maximize the flatness of the new and corrected volume. EBHC is evaluated using data acquired with a modern cone-beam dual-source spiral CT scanner (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany), with a modern dual-source micro-CT scanner (Tomo-Scope Synergy Twin, CT Imaging GmbH, Erlangen, Germany), and with a modern C-arm CT scanner (Axiom Artis dTA, Siemens Healthcare, Forchheim, Germany). A large variety of phantom, small animal, and patient data were used to demonstrate the data and system independence of EBHC.
RESULTS: Although no physics apart from the initial segmentation procedure enter the correction process, beam hardening artifacts were significantly reduced by EBHC. The image quality for clinical CT, micro-CT, and C-arm CT was highly improved. Only in the case of C-arm CT, where high scatter levels and calibration errors occur, the relative improvement was smaller.
CONCLUSIONS: The empirical beam hardening correction is an interesting alternative to conventional iterative higher order beam hardening correction algorithms. It does not tend to over- or undercorrect the data. Apart from the segmentation step, EBHC does not require assumptions on the spectra or on the type of material involved. Potentially, it can therefore be applied to any CT image.

Entities:  

Mesh:

Year:  2010        PMID: 21089751     DOI: 10.1118/1.3477088

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality.

Authors:  Lifeng Yu; Jodie A Christner; Shuai Leng; Jia Wang; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Simplified Statistical Image Reconstruction for X-ray CT With Beam-Hardening Artifact Compensation.

Authors:  Monica Abella; Cristobal Martinez; Manuel Desco; Juan Jose Vaquero; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2019-06-10       Impact factor: 10.048

3.  Reduction of Beam Hardening Artifacts in Cone-Beam CT Imaging via SMART-RECON Algorithm.

Authors:  Yinsheng Li; John Garrett; Guang-Hong Chen
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-22

4.  Known-component metal artifact reduction (KC-MAR) for cone-beam CT.

Authors:  A Uneri; X Zhang; T Yi; J W Stayman; P A Helm; G M Osgood; N Theodore; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2019-08-21       Impact factor: 3.609

5.  Convolutional Neural Network Based Metal Artifact Reduction in X-Ray Computed Tomography.

Authors:  Yanbo Zhang; Hengyong Yu
Journal:  IEEE Trans Med Imaging       Date:  2018-06       Impact factor: 10.048

6.  Calibration Free Beam Hardening Correction for Cardiac CT Perfusion Imaging.

Authors:  Jacob Levi; Rachid Fahmi; Brendan L Eck; Anas Fares; Hao Wu; Mani Vembar; Amar Dhanantwari; Hiram G Bezerra; David L Wilson
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-21

Review 7.  Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Joel G Fletcher
Journal:  Radiology       Date:  2015-09       Impact factor: 11.105

Review 8.  Modelling the physics in the iterative reconstruction for transmission computed tomography.

Authors:  Johan Nuyts; Bruno De Man; Jeffrey A Fessler; Wojciech Zbijewski; Freek J Beekman
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

9.  Non-contact respiration monitoring for in-vivo murine micro computed tomography: characterization and imaging applications.

Authors:  Laurel M Burk; Yueh Z Lee; J Matthew Wait; Jianping Lu; Otto Z Zhou
Journal:  Phys Med Biol       Date:  2012-09-05       Impact factor: 3.609

10.  Low dose dynamic myocardial CT perfusion using advanced iterative reconstruction.

Authors:  Brendan L Eck; Rachid Fahmi; Christopher Fuqua; Mani Vembar; Amar Dhanantwari; Hiram G Bezerra; David L Wilson
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.