Richard A Parker1, Christopher J Weir1, Noah Rubio2, Roberto Rabinovich2, Hilary Pinnock3, Janet Hanley1,4, Lucy McCloughan5, Ellen M Drost2, Leandro C Mantoani2, William MacNee2, Brian McKinstry1,5.
Abstract
INTRODUCTION: The Bland-Altman limits of agreement method is widely used to assess how well the measurements produced by two raters, devices or systems agree with each other. However, mixed effects versions of the method which take into account multiple sources of variability are less well described in the literature. We address the practical challenges of applying mixed effects limits of agreement to the comparison of several devices to measure respiratory rate in patients with chronic obstructive pulmonary disease (COPD).Entities:
Mesh:
Year: 2016 PMID: 27973556 PMCID: PMC5156413 DOI: 10.1371/journal.pone.0168321
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
95% limits of agreement results for the respiratory rate measurements for each of the devices compared to the gold standard measure.
| Device | No. of participants(total valid differences/ total after removing outliers) | Mean bias (Fixed effects 95% LoA) | Mean bias (Mixed effects 95% LoA) | Mean bias (REVISED mixed effects 95% LoA |
|---|---|---|---|---|
| Camera (rate per second) | 21 (192/188) | -3.32 (-13.35 to 6.72) | -3.21 (-12.71 to 6.30) | -3.15 (-11.54 to 5.24) |
| Camera (rate per minute) | 21 (192/188) | -4.43 (-15.54 to 6.69) | -4.35 (-14.98 to 6.28) | -4.43 (-13.13 to 4.26) |
| PPG (raw) | 21 (378/377) | 3.53 (-10.30 to 17.35) | 3.53 (-10.30 to 17.35) | 3.46 (-10.06 to 16.98) |
| PPG (median filtered) | 21 (378/376) | 3.01 (-11.16 to 17.17) | 3.01 (-11.17 to 17.19) | 3.02 (-10.53 to 16.57) |
| Impedance | 20 (304/304) | -1.17 (-20.07 to 17.73) | -1.18 (-20.07 to 17.72) | -1.18 (-20.07 to 17.72) |
| Accelerometer | 20 (284/282) | -2.18 (-8.74 to 4.38) | -2.18 (-8.63 to 4.27) | -2.14 (-7.91 to 3.63) |
| Chest-band | 21 (385/384) | -1.61 (-9.99 to 6.78) | -1.60 (-9.99 to 6.80) | -1.67 (-9.64 to 6.30) |
Mean bias = Average difference; LoA = Limits of Agreement.
*Outliers removed and emphasis is placed on model fitting rather than consistency in methods across devices.
Fig 1Bland-Altman Plots showing the paired differences against the average for two devices (Camera and PPG).
Mean bias and limits of agreement are shown by the dashed lines, while confidence intervals are shown by the dotted lines. (A) Camera: rate per second. (B) Camera: rate per minute. (C) PPG: raw. (D) PPG: median filtered.
Fig 2Bland-Altman Plots showing the paired differences against the average for three devices (Impedance, Accelerometer, Chest-band).
Mean bias and limits of agreement are shown by the dashed lines, while confidence intervals are shown by the dotted lines. (A) Impedance. (B) Accelerometer. (C) Chest-band.
Comparison of the variabilities of differences across devices.
| Device | Original Mixed Effects Model | Revised Mixed Effects Model | ||||||
|---|---|---|---|---|---|---|---|---|
| Within- participant SD | Combined SD | Maximum negative difference | Maximum positive difference | Within- participant SD | Combined SD | Maximum negative difference | Maximum positive difference | |
| Camera (rate per second) | 4.25(3.87 to 4.79) | 4.85(4.44 to 5.76) | 35.00 | 14.71 | 3.92 (3.58 to 4.44) | 4.28 (3.93 to 4.94) | 35.00 | 9.65 |
| Camera (rate per minute) | 5.18 (4.74 to 5.86) | 5.42 (4.97 to 6.15) | 35.00 | 10.63 | 4.31 (3.95 to 4.89) | 4.44 (4.07 to 4.97) | 35.00 | 10.63 |
| PPG (raw) | 6.02 (5.64 to 6.53) | 7.05 (6.61 to 8.60) | 18.66 | 30.19 | 5.85(5.46 to 6.37) | 6.90 (6.45 to 8.26) | 18.66 | 23.44 |
| PPG (median filtered) | 6.20 (5.81 to 6.77) | 7.23 (6.76 to 8.63) | 25.00 | 32.39 | 5.84 (5.46 to 6.36) | 6.91 (6.44 to 8.29) | 22.00 | 23.65 |
| Impedance | 8.94 (8.29 to 9.84) | 9.64 (8.97 to 10.91) | 32.00 | 34.00 | 8.94(8.29 to 9.84) | 9.64(8.97 to 10.91) | 32.00 | 34.00 |
| Accelerometer | 3.23 (3.00 to 3.59) | 3.29 (3.05 to 3.62) | 24.84 | 12.99 | 2.85(2.64 to 3.15) | 2.94(2.73 to 3.25) | 14.82 | 12.99 |
| Chest-band | 4.17(3.91 to 4.53) | 4.28 (4.02 to 4.65) | 24.65 | 27.80 | 3.98(3.73 to 4.33) | 4.07(3.81 to 4.42) | 24.65 | 20.76 |
*Outliers removed and emphasis is given to model fitting rather than consistency in methods across devices.
95% confidence intervals around each of the limits of agreement for the original mixed effects method (results after removing outliers are shown in italics for comparison).
| Device | Mean bias | 95% repeated differences LoA | 95% bootstrap CI of lower limit | 95% bootstrap CI of upper limit |
|---|---|---|---|---|
| Camera (rate per second) | -3.21 | -12.71 to 6.30 | -14.84 to -11.42 | 5.00 to 8.39 |
| Camera (rate per minute) | -4.35 | -14.98 to 6.28 | -16.70 to -13.65 | 4.96 to 8.00 |
| PPG (raw) | 3.53 | -10.30 to 17.35 | -14.23 to -8.58 | 15.58 to 21.16 |
| PPG (median filtered) | 3.01 | -11.17 to 17.19 | -15.10 to -9.39 | 15.35 to 20.79 |
| Impedance | -1.18 | -20.07 to 17.72 | -23.84 to -17.81 | 15.49 to 21.65 |
| Accelerometer | -2.18 | -8.63 to 4.27 | -9.45 to -7.96 | 3.62 to 5.21 |
| Chest-band` | -1.60 | -9.99 to 6.80 | -11.04 to -9.19 | 6.05 to 7.86 |
95% confidence intervals were calculated using a parametric bootstrap-t method based on 1999 resamples.
*No outliers