Literature DB >> 27936043

Reproducibility for Heart Rate Variability Analysis during 6-Min Walk Test in Patients with Heart Failure and Agreement between Devices.

Lays Magalhães Braga1, Gustavo Faibischew Prado2, Iracema Ioco Kikuchi Umeda3, Tatiana Satie Kawauchi1, Adriana Marques Fróes Taboada3, Raymundo Soares Azevedo4, Horacio Gomes Pereira Filho5, César José Grupi5, Hayala Cristina Cavenague Souza3, Dalmo Antônio Ribeiro Moreira3, Naomi Kondo Nakagawa1.   

Abstract

Heart rate variability (HRV) analysis is a useful method to assess abnormal functioning in the autonomic nervous system and to predict cardiac events in patients with heart failure (HF). HRV measurements with heart rate monitors have been validated with an electrocardiograph in healthy subjects but not in patients with HF. We explored the reproducibility of HRV in two consecutive six-minute walk tests (6MW), 60-minute apart, using a heart rate monitor (PolarS810i) and a portable electrocardiograph (called Holter) in 50 HF patients (mean age 59 years, NYHA II, left ventricular ejection fraction ~35%). The reproducibility for each device was analysed using a paired t-test or the Wilcoxon signed-rank test. Additionally, we assessed the agreement between the two devices based on the HRV indices at rest, during the 6MW and during recovery using concordance correlation coefficients (CCC), 95% confidence intervals and Bland-Altman plots. The test-retest for the HRV analyses was reproducible using Holter and PolarS810i at rest but not during recovery. In the second 6MW, patients showed significant increases in rMSSD and walking distance. The PolarS810i measurements had remarkably high concordance correlation [0.86<CCC<0.99] based on Holter at rest, during 6MW and recovery. At higher rates, a small effect in increasing differences between Holter and Polar in R-R intervals was observed. In conclusion, our study showed good reproducibility of HRV at rest in two consecutive 6MW using Holter and PolarS810i. Additionally, PolarS810i produced good agreements in short-term HRV indices based on Holter simultaneous recordings at rest, during the 6MW and recovery in HF patients.

Entities:  

Mesh:

Year:  2016        PMID: 27936043      PMCID: PMC5147870          DOI: 10.1371/journal.pone.0167407

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

In recent decades, studies have demonstrated that beat-to-beat interval variation analysis, called heart rate variability (HRV), is a useful non-invasive method to assess the autonomic modulation of the sinus node. HRV describes the dispersion of R-R intervals in determined temporal series that reflect the balance between the dynamic interferences of the sympathetic and parasympathetic nervous systems with positive and negative chronotropic effects, respectively. HRV analysis derived from an 24-hour electrocardiogram (ECG) has shown stability in healthy subjects [1], in patients with heart failure (HF) New York Heart Association (NYHA) functional class II-III taking their usual medications [2], or in patients with congestive HF secondary to coronary artery disease [3]. Studies on patients with HF have shown associations between reduced HRV and sudden death [4], one-year mortality [5] and overall mortality [6] using a 24-hour ECG. Using a short-term ECG, studies have shown associations between HRV reduction and an increased hazard ratio of new HF in hypertensive subjects [7] and sudden death in patients with HF [8]. Increased HRV was shown to have a protective role in the prevention of cardiac events [9]. Heart rate monitors are inexpensive and make it moderately simple to record, edit and analyse R-R intervals and HRV indices. In healthy subjects, the reproducibility of short-term HRV analysis from heart rate monitors based on ECGs has been extensively investigated, and its ability has been shown to be comparable to portable ECGs [10-17]. However, it is of interest to evaluate whether there is agreement between short-term HRV analysis from heart rate monitors and ECGs in patients with HF. To our knowledge, the reproducibility of short-term HRV analysis using portable ECGs and heart rate monitors and the agreement between heart rate monitors and ECGs, based on portable ECGs have not been investigated in patients with HF. We hypothesized that the two devices may have a similar ability to reproduce HRV indices in patients with moderate HF. Good reproducibility and agreement between devices are essential for the maximal utility of instruments in clinical follow-up and research. Due to the limited information available, we investigated the reproducibility of short-term HRV using simultaneously a portable ECG (Holter Cardiolight, Cardios Ind., São Paulo, Brazil) and a heart rate monitor (PolarS810i, Polar Electro-OY, Kempele, Finland) at rest and during recovery from two consecutive 6MW, 60-min apart. We also investigated the agreement of short-term HRV indices from Polar S810i based on Holter at rest, during the six-min walk test (6MW) and during recovery in stable patients with moderate HF.

Methods

Participants

This study followed the ethical guidelines of the Declaration of Helsinki, and it was approved by the Ethical Committees of University of Sao Paulo School of Medicine and Dante Pazzanese Heart Institute of Sao Paulo State (CEP 232/12). The patients entered into the study after providing their writing informed consent. Consecutive patients with HF, NYHA functional class II with a left ventricular ejection fraction (LVEF) < 40% were recruited from a list of a cardiac rehabilitation centre in a six-month period in 2014. Exclusion criteria were uncontrolled arrhythmia, chronic atrial fibrillation, acute HF, decompensated HF, ectopic beats in ≥ 5% of measurements, the use of a paced rhythm, current smoking or a smoking time cessation < 24 months and any orthopaedic or cognitive problems that could affect 6MW performance. The subjects’ weight and height were used to calculate their body mass index. Hypertension, diabetes and smoking history were recorded. Hypertension was defined according to the American Association of Cardiology [18]. Diabetes was defined in accordance with the American Association of Diabetes [19]. Medications routinely used by the patients were recorded. Subjects self-reported their regular physical activity during the week prior to the study using the International Physical Activity Questionnaire validated for the Brazilian population [20]. Their answers allowed them to be classified as “active” (30 min of regular physical activity ≥ 3 times per week) or “sedentary”. All subjects reported to the laboratory between 8 and 11 A.M. on one single occasion to perform two consecutive 6MW, 60-min apart. Holter and Polar S810i were used simultaneously. Subjects were instructed to eat a light breakfast at least 2 hours before testing, to take their usual medications, to abstain from consuming tea, alcohol and beverages containing caffeine, and to refrain from heavy physical exertion in the 24 hours preceding the assessments. Subjects underwent a physical examination in a controlled environment (24°C room temperature and 55–60% relative humidity) and stayed at supine rest for 25 min before and after the two 6MW. The two 6MW were performed in accordance with the ATS guidelines [21]. For the reproducibility of the test-retest 6MW, R-R intervals were recorded and the HRV data were analysed at rest and during recovery from the two 6MW using both devices, Holter and PolarS810i. For the HRV analysis of agreement between PolarS810i and holter, based on Holter, we used data from the second 6MW (Fig 1).
Fig 1

The study design shows the two consecutive 6MW to analyse the reproducibility of the two devices, Holter and PolarS810i at rest and during recovery and the reliability of PolarS810i based on Holter at rest, 6MW and during recovery from the second 6MW.

Heart rate variability using Holter and Polar S810i

Subjects underwent a simultaneous 60-min R-R recording using Holter and PolarS810i for each 6MW. In brief, subjects were prepared for the attachment of a 3-lead ECG (called Holter, CardioLight model, Cardios Inc., São Paulo, Brazil) that was in a standard configuration and placed on the sternum manubrium, at the xiphoid appendix and at the 5th intercostal space in the middle clavicular line. Then, a strap of the PolarS810i was fixed around the chest of the subject in accordance with the manufacturer’s instructions. Subjects were instructed to rest in supine with their head supported by a pillow and not to talk or move excessively any part of their body. R-R recordings were done when the subjects were resting in supine at baseline, walking for 6 min and recovering from the 6MW. In each phase of the 6MW, only a 5 min time series with a minimum of 256 R-R intervals was analysed for HRV indices, as recommended [1]. At rest, the 5-min right before beginning the 6MW was taken to limit possible influence of equipment installation, talking and deambulation (the washout effect). At 6-minute period of the 6MW, the first minute was excluded to get more stable and comparable signals. At recovery, the first minutes from the total 25-minute were taken as the time-series for comparisons. All assessors were blinded to subjects’ identification and conditions. ECG data acquisition was performed using a Holter device at a sampling frequency of 800 Hz with temporal resolution of 1 ms for each R-R interval. Data were stored in a computer and further analysed by two physicians using CardioSmart S550 Software (Cardios Inc., São Paulo, Brazil). Each ECG recording was inspected for ectopic beats and artefacts, which were replaced with the mean value of the previous and subsequent intervals with the accepted R-R interval in accordance with the recommendations of the Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology [1]. The R-R intervals of the Polar S810i registered throughout the study were downloaded and edited on a laptop computer with the aid of software (Polar Pro-Trainer 5 Software, Kempele, Finland). A sampling frequency of 1,000 Hz with temporal resolution of 1-ms for each R-R interval was chosen. The Polar software has an automatic R-R interval filtering that removes detected errors and aberrant beats and interpolates intervals with the aid of an algorithm described above. After this, all registrations were visually inspected for removal of artifacts or aberrant beats within the limit of 5% of the R-R intervals. Data were analysed by a physiotherapist and a physician using Kubios HRV Analysis Software 2.1 (Biomedical Signal Analysis Group, Kupio, Finland). For the analysis of agreement between the Polar and Holter for the data acquisition, R-R intervals registrations were tested. For the time domain, HRV analysis included the SDNN (standard deviation of all R-R intervals), the root mean square successive difference between R-R intervals (rMSSD), and the percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms (pNN50). For the frequency-domain, spectral analysis was calculated using fast Fourier transform algorithms to analyse LF in normalized units (LF nu), the high frequency in normalized units (HF nu) and the LF/HF ratio.

Statistical analysis

Statistical analyses were conducted using SPSS statistical software version 19.0 (SPSS Inc., IL, USA) and R 3.2.1 software. Data are summarized as means, standard deviations and proportional frequencies, when appropriate. The reproducibility of the HRV indices (R-R intervals, SDNN, rMSSD, pNN50, LF nu, HF nu and LF/HF ratio) and the walking distance of the test-retest 6MW using Holter and Polar S810i were compared using a paired t-test or a Wilcoxon signed-rank test. A p-value < 0.05 was considered statistically significant. The agreement of the HRV analysis with between PolarS810i and Holter, based on Holter was tested in the second 6MW in HF patients (at rest, during the 6MW and during recovery) and was estimated using the Concordance Correlation Coefficient (CCC) with 95% confidence intervals (95% CI) [22], and Bland-Altman plots [23]. Linear regression models were developed to explore the possibility of biases of proportionality adopting the average value of Holter and Polar measures as independent variables and the differences between both measures as dependent variables.

Results

Fifty-four patients with mild/moderate systolic dysfunction (17% ≤ LVEF ≤ 55%) were admitted to the study and underwent two consecutive 6MW, 60-minute apart. Holter and Polar S810i were used simultaneously. The recordings of four patients were excluded due to aberrant beats/artefacts ≥ 5% after editing. Demographic and clinical data (Table 1) showed that the majority of patients were older, male, hypertensive and sedentary and had ischaemic heart disease.
Table 1

Demographic and clinical data of 50 patients with HF NYHA class II are presented as mean values (SD) or absolute numbers and proportion of patients when appropriate.

HF subjects n = 50
Age, years59 (6.2)
BMI, kg/m224.6 (5.2)
Male, n (%)39 (78)
LVEF, %35.4 (7.8)
Etiology, n (%)
Ischemic38 (76)
Non-ischemic12 (24)
Other morbidities, n (%)
Hypertension50 (100)
Diabetes21 (42)
Myocardial infarction38 (76)
Medications, n (%)
Antiarrhythmics9 (18)
Anticoagulants8 (16)
Angiotensin receptor blockers12 (24)
ACE inibitors31 (62)
Antiplatelet agents45 (90)
Beta-blockers49 (98)
Digitalics13 (26)
Diuretics49 (98)
Antidepressants27 (54)
Arrhythmias > 10 beats/hour, n (%)
Atrial premature beats9 (18)
Ventricular premature beats19 (38)
Smoking
Time of cessation, years11 (6)
Pack-years22.8 (14.7)
Physical activity, n patients (%)
Sedentary38 (76)
Active12 (24)

Abbreviations: HF, heart failure, NYHA, New York Heart Association, BMI, body mass index, LVEF, left ventricular ejection fraction, ACE, angiotensin-converting enzyme

Abbreviations: HF, heart failure, NYHA, New York Heart Association, BMI, body mass index, LVEF, left ventricular ejection fraction, ACE, angiotensin-converting enzyme The walking distance was higher in the second 6MW (489.2 ± 64.7 m) than the first 6MW (474.1 ± 65.8 m, p < 0.001). The reproducibility of R-R intervals was first analysed. Similar mean values were shown in the test-retest 6MW at rest and during recovery when using Holter (p = 0.938 and p = 0.714, respectively) and PolarS810i (p = 0.919 and p = 0.745, respectively). The HRV indices using both devices also showed similar mean values at rest and during recovery (Table 2). The rMSSD data showed a slightly significant difference between the two 6MW in the recovery phase using both devices.
Table 2

Reproducibility of HRV analysis by Holter and by PolarS810i at rest and during recovery from the 6MW.

Data are presented as mean (min-max) values.

RestRecovery
Test 1Test 21x2 p-valueTest 1Test 21x2 p-value
Holter
HR,bpm63 (50–96)62 (48–95)67 (50–110)66 (50–109)
SDNN, ms29.9 (13–56)29.3 (9–57).59129.7 (11–51)31.0 (14–55).306
rMSSD, ms16.8 (5–35)17.2 (8–30).35215.9 (7–32)17.3 (8–34).020
pNN50, %1.3 (0–7)1.4 (0–8).3941.1 (0–7)1.1 (0–5).766
LF nu73.4 (54–100)72.2 (25–93).69672.0 (46–93)70.5 (16–100).904
HF nu28.9 (9–49)27.3 (8–49).54028.5 (9–58)28.1 (5–57).915
LF/HF3.4 (1–11)3.6 (1–12).6973.3 (1–9)3.6 (1–19).569
PolarS810i
HR,bpm62 (50–94)62 (47–95)66 (49–109)66(49–110)
SDNN, ms28.4 (12–53)27.4 (9–55).50528.0 (10–49)29.4 (54–12).295
rMSSD, ms15.2 (4–33)15.8 (6–28).33814.3 (6–31)15.7 (6–31).025
pNN50, %1.0 (0–6)1.0 (0–7).8710.7 (0–6)0.8 (0–4).715
LF nu72.2 (51–98)73.1 (22–94).33771.7 (42–94)72.0 (17–95).412
HF nu28.7 (7–48)27.2 (8–50).42928.0 (9–51)27.8 (5–53).866
LF/HF3.5 (1–14)3.6 (1–12).4183.3 (1–10)3.8 (1–20).601

Abbreviations: HR, heart rate, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio, 6MW, six-min walk test

Reproducibility of HRV analysis by Holter and by PolarS810i at rest and during recovery from the 6MW.

Data are presented as mean (min-max) values. Abbreviations: HR, heart rate, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio, 6MW, six-min walk test The agreement between crude data of R-R intervals (Fig 2) was high (CCC = 0.999, p<0.001) with lower dispersion beyond the limits (± 1.96 SD) from the Bland-Altman diagram.
Fig 2

Reliability between Holter and PolarS810i for R-R intervals at rest and movement using Bland-Altman plots.

There was a small but significant effect (β = -0.010) of the independent variable (the average R-R intervals between Holter and PolarS810i) on the differences between measures obtained with both, Holter and Polar (Table 3). The other regression models proved to be not statistically significant, removing the possibility of any biases of proportionality. Bland-Altman plots showed good agreement between the Holter and PolarS810i in time and frequency domains that were assessed at rest (Fig 3). PolarS810i showed good predictive performance to assess Holter HRV values in different ranges of variation of variables. During the 6MW and recovery, the CCC between values from Holter and Polar S810i showed good agreements for most HRV indices (Table 4).
Table 3

Linear regression models for proportionality bias in absolute differences and average (a-) values between Holter and Polar810i.

Constantβ95% CIp-value
a-RR9.255-0.010-0.018; -0.0020.016
a-SDNN0.086-0.034-0.095; 0.0260.262
a-rMSSD-1.5160.047-0.021; 0.1150.171
a-pNN50-0.252-0.065-0.171; 0.0410.224
a-LF1.898-0.019-0.083; 0.0460.558
a-HF1.142-0.051-0.125; 0.0220.168
a-LF/HF0.281-0.079-0.169; 0.0120.086

Abbreviations: RR, intervals between two peaks, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio, β, linear regression coefficient

Fig 3

Reliability between Holter and Polar S810i for HRV analyses at rest using Bland-Altman plots.

Table 4

Concordance Correlation Coefficients [95% CI] of HRV indices between Holter and PolarS810i at rest, during the 6MW and recovery.

Rest6MWRecovery
R-R, ms0.99 [0.98–0.99]0.99 [0.97–0.99]0.99 [0.98–0.99]
SDNN, ms0.97 [0.96–0.98]0.95 [0.92–0.97]0.97 [0.95–0.98]
rMSSD, ms0.96 [0.94–0.97]0.93 [0.89–0.95]0.94 [0.90–0.96]
pNN50, %0.91 [0.85–0.94]0.86 [0.79–0.91]0.86 [0.78–0.92]
LF nu0.97 [0.95–0.98]0.96 [0.94–0.98]0.94 [0.90–0.96]
HF nu0.97 [0.94–0.98]0.98 [0.96–0.99]0.97 [0.94–0.98]
LF/HF0.95 [0.91–0.97]0.95 [0.92–0.97]0.98 [0.96–0.99]

Abbreviations: HRV, heart rate variability, 6MW, six-min walk test, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio

Abbreviations: RR, intervals between two peaks, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio, β, linear regression coefficient Abbreviations: HRV, heart rate variability, 6MW, six-min walk test, SDNN, standard deviation of all R-R intervals, rMSSD, root mean square successive difference between R-R intervals, and pNN50, percentage of successive differences in the R-R interval of which the absolute values exceeds 50-ms, LF nu, low-frequency normalized unit, HF nu, high-frequency normalized unit, LF/HF, low and high-frequency ratio

Discussion

To the best of our knowledge, the present study is the first to investigate the reproducibility of short-term HRV analysis using simultaneously Holter and PolarS810i and to examine the reliability of PolarS810i HRV analysis based on Holter in patients with moderate HF during the 6MW. The present study validated Polar in the short-term HRV analysis in this specific population of HF subjects with NYHA functional class II at rest. There was also good agreement between Polar and Holter on the HRV measures during the 6MW, despite a small effect of elevated HR (lower R-R intervals) on the absolute differences between the two devices, Holter and Polar. Minimum 256 R-R intervals for HRV analysis were registered as recommended [1], in the three conditions (at rest, during the 6MW and during recovery) using both Holter and PolarS810i simultaneously. Low values of vagal indices and high values of sympathetic modulation were found in stable patients with moderate HF in all three conditions. These findings are consistent with other studies using portable ECGs and heart rate monitors [8, 24, 25]. The present study showed that short-term HRV indices are stable and reproducible using Holter and PolarS810i at rest because, as expected, the capturing and editing of stationary signals facilitate HRV analysis in stable conditions. Good CCC and Bland-Altman plots also indicated that PolarS810i produces good agreements between short-term HRV results, based on simultaneous recordings, and ECG at the three phases: at rest, during the 6MW and during recovery. The reproducibility of short-term HRV indices using heart rate monitors in two consecutive 6MW was investigated by Corrêa et al. [26] using very short recordings at standing rest (1-min) and at the last 2 min of the 6MW in 38 obese subjects (15 male) aged 65 yrs. Subjects showed a similar performance in the two 6MW (mean difference = 2%), and the HRV indices were reproducible in time and non-linear domains. The authors concluded that these HRV indices may be useful in quantifying the effects of interventions on the autonomic modulation of the heart rate, particularly during exercise. In the present study, patients with HF NYHA II walked 15 meters longer in the second 6MW, which may be attributed to the well-documented high inter-test variability, related to fluctuating motivation, fatigue, and learning effects [27-30]. However, the rMSSD data from 6MW recovery in the first test were not similar to the second test using both devices, Holter and Polar S810i. This result may be associated with the highly skewed distribution of time-domain variables, as suggested by others in studies with ECG recordings [31]. In the short-term HRV index determination using ECG, a review of the literature showed that the reproducibility of these measurements is heterogeneous [32]. For instance, Ponikowski et al. [31] used 5-, 10-, 20- and 40-min recordings at supine rest. Although R-R intervals were reproducible (8% variability), the authors found poor reproducibility (25–139% variability) particularly for pNN50 and SDNN in patients with chronic HF, which is possibly attributed to the skewed distribution of these indices. However, it should be considered that the reproducibility in Ponikowski and co-workers study was tested in two moments with a large interval in between them (7 to 56 days), which might have allowed for changes in the clinical conditions of the patients with chronic HF and in turn changed the HRV indices. In our study, the two 6MW were controlled and had an interval of 60 min in between them, which may have helped avoid possible bias, such as changes in the environment and in the patients’ clinical conditions. The simplicity and low cost of heart rate monitors combined with the good interchangeability with the gold-standard ECG for HRV analyses are advantages that have been largely demonstrated in healthy subjects [10-13] but not female subjects [14]. However, the agreement of short-term HRV indices between heart rate monitors and ECGs, based on ECG is little known in patients with HF. In our study, short-term HRV indices were analysed using two devices simultaneously. The analysis of agreement between the two devices was possible because of the use of a Holter with a roll-over analysis, which allowed us to focus on rapid and transitory changes by dividing time intervals into 5-min periods. We believe that this process of analysis was responsible for the high level of concordance between the values observed with Holter and Polar S810i in patients with moderate HF. In healthy subjects, Kingsley et al. [33] assessed reliability and compared R-R intervals obtained during an incremental cycle ergometer test in healthy subjects using ECG and PolarS810i with a good correlation between devices. They showed good to excellent correlations (0.93–0.99) between the devices. In the present study, we also found excellent reliability of HRV indices at rest, during the 6MW and during recovery in patients with HF. We used the 6MW as a non-pharmacologic intervention to produce sympathetic activation. Fortunately, the signals were stable during the exercise, and by dropping the first minute of the 6MW, we succeeded in having the minimum 256 intervals for HRV analysis in the last five minutes. The present study had some limitations. Participants were predominantly males who were aged, on average, 59 years and taking beta-blockers. However, there are studies that show that HRV results during exercise are independent of age and sex [34]. Previous studies that evaluated the reproducibility between Polar and Holter in the analysis of HRV in healthy subjects showed overall good intra-class correlation but unsatisfactory results in women [14] in frequency-domain variables [15] and rMSSD [11]. The use of beta-blockers is associated with increase in cardiac parasympathetic activity, in normal persons [35], and enhances HRV in patients with coronary artery disease possibly reducing heart rate and/or counteracting the adverse effects of sympathetic activity [36]. However, beta-blocker medications do not affect sudden beat-to-beat changes in R-R intervals [37] and the subjects of our study were simultaneously assessed by Holter and Polar for HRV measures. Additionally, we aimed to implement a real life scenario in which patients with stable HF take beta-blockers and other medications in their prescription. Another limitation is that, the 24-hour registration of the HRV can´t be represented by only 5 minutes analysis. However, several studies have validated the short-term HRV analysis in healthy subjects under controlled conditions ([10-17], Table 5). This study, which investigated the validity of Polar for HRV analysis in the specific population of HF patients, showed high agreement between all variables assessed simultaneously by Holter and Polar, which allowed us to assume the interchangeability of these methods in the evaluation of HRV in specific high-risk subjects. As suggested by others, the agreement that may be considered sufficient for the interchangeable use of two methods should have a 95% CI of > 0.75 [38]. Although there are intrinsic differences between Holter and Polar in signal acquisition, studies have validated Polar for short-term HRV analysis with good accuracy compared with the gold-standard method ECG in healthy subjects [10-17]. It is reasonable to expect that the quality of the signal acquisition during movement would be affected by transitory loss of signal and other factors. As pointed out before, Polar has important limitation for studies with patients with structural cardiac changes as it does not register ECG signals. However, in some way, this limitation seemed to be overcome in the HF population NYHA II. The present study validated Polar for HRV analysis during a sub-maximal functional capacity test can be considered simple and an important issue for future studies to assess the short-term response of beat-to-beat variations and, possibly, the response of the autonomic nervous system to specific interventions (for example, exercise) in HF patients NYHA class II. Although we have demonstrated through the linear regression model a small bias of proportionality in the correlation of the RR interval between the two devices (β = -0.010), this effect was small in magnitude. For instance, each reduction of 100 ms in the R-R interval was associated with an increase in one ms in the absolute difference between Holter and Polar measures. In this sense, besides the routine use of clinical variables and the walking distance registered in the 6MWT, the short-term HRV analysis during the 6MW can be another important tool for the health assessment of highly complex patients.
Table 5

Reliability between short-term ECG and portable heart rate monitors for HRV measurements in healthy subject.

AuthorsN healthy subjects Age (years)HRV recordingStatistical analysisECG results Mean (SD) or Median (IQR)PolarS810i results Mean (SD) or Median (IQR)ECG and Polar AgreementAuthors conclusion on agreement
Barbosa et al. [10]30 male subjects Mean age = 21 yrsECG and PolarRS800 5 min—SupinePearson or Spearman correlationSDNN = 58.8 (23.3)SDNN = 57.1 (23.3)SDNN = 0.99For all variables
rMSSD = 48.9 (26.5)rMSSD = 47.4 (24.9)rMSSD = 0.99
LF nu = 51.3 (14.2)LF nu = 50.2 (12.8)LF nu = 0.95
HFnu = 48.7 (14.2)HF nu = 49.8 (12.8)HF nu = 0.87
LF/HF = 1.35 (1.3)LF/HF = 1.16 (0.7)LF/HF = 0.91
Gamelin et al. [11]18 male subjects Mean age = 27 yrsECG and Polar S810i 5 min—SupinePearson or Spearman correlationSDNN = 50.2 (18.8)SDNN = 50.1 (18.8)SDNN = 0.99For all variables
rMSSD = 46.7 (23.0)rMSSD = 46.5 (24.0)rMSSD = 0.99
pNN50 = 26.2 (21.0)pNN50 = 25.9 (21.0)pNN50 = 0.99
LF nu = 44.9 (22.0)LF nu = 45.0 (22.9)LF nu = 0.99
HF nu = 55.0 (23.0)HF nu = 55.0 (22.9)HF nu = 0.99
LF/HF = 1.2 (1.2)LF/HF = 1.3 (1.2)LF/HF = 0.99
Radespiel-Troger et al. [12]36 subjects Mean age = 27 yrs (22 male)ECG and Polar 3 min—SittingPearson or Spearman correlationHR = 79 (11)HR = 79 (12)HR = 0.99For all variables
SDNN = 95.3 (44.0)SDNN = 95.3 (43.0)SDNN = 0.99
rMSSD = 58.0 (33.4)rMSSD = 57.5 (33.0)rMSSD = 0.99
Vanderlei et al. [13]15 male subjects Mean age = 21 yrsECG and PolarS810i 5 min—SupineIntraclass correlationrMSSD = 29.5 (3.3)rMSSD = 29.7 (3.3)rMSSD = 0.99For all variables
pNN50 = 52.8 (4.1)pNN50 = 53.1 (4.2)pNN50 = 0.99
LF nu = 60.5 (3.8)LF nu = 61.7 (3.6)LF nu = 0.97
HF nu = 39.5 (3.8)HF nu = 8.3 (3.6)HF nu = 0.97
LF/HF = 1.9 (0.3)LF/HF = 1.9 (0.2)LF/HF = 0.98
Wállen et al. [14]341 subjects Mean age = 52 yrs (139 male)ECG and PolarRS800 5 min—SupineIntraclass correlationsSDNN = 39.7 (37.7–41.7)SDNN = 40.9 (38.8–43.2)SDNN = 0.84With limitations in female subjects
rMSSD = 25.1 (23.4–26.9)rMSSD = 25.2(23.6–26.9)rMSSD = 0.93
Nunan et al. [15]33 subjects Median age = 34 yrs (19 male) Median age = 48 yrs (14 female)ECG and PolarS810i 5 min—SupineIntraclass correlation Data with log transformationMean R-Rlog = 979.4 (176.9)Mean R-Rlog = 980.6 (178.6)Mean R-R = 0.99Only for time domain variables and LF/HF
SDNNlog = 4.0 (0.5)SDNNlog = 4.1 (0.5)SDNN = 0.87
rMSSDlog = 3.7 (0.7)rMSSDlog = 3.7 (0.6)rMSSD = 0.88
LF nulog = 59.0 (17.8)LF nulog = 62.5 (14.5)LF nu = 0.75
HF nulog = 41.0 (17.7)HF nulog = 37.5 (14.5)HF nu = 0.72
LF/HFlog = 2.1 (2.1)LF/HFlog = 2.2 (1.9)LF/HF = 0.90
Porto et al. [16]25 subjects Mean age = 26 yrs (16 male)ECG and PolarS810i 5 min—SupineLimits of agreement Absolute DifferenceMean R-R = 949.0 (141.0)Mean R-R = 951.0 (151.0)Mean R-R = -6.4 to 2.7For all variables
SDNN = 61.2 (31.2)SDNN = 60.9 (32.7)SDNN = -1.7 to 2.3
rMSSD = 60.4 (35.7)rMSSD = 59.6 (36.5)rMSSD = not shown
pNN50 = 30.1 (32.3)pNN50 = 32.3 (20.9)pNN50 = -5.5 to 1.1
Weippert et al. [17]19 male subjects Mean age = 24 yrsECG and PolarS810i 3 min—SupineIntraclass correlationsLF nulog = 0.6 (0.2)LF nulog = 0.7 (0.2)LF nu = 0.95For all variables
HF nulog = 0.3 (0.1)HF nulog = 0.3 (0.1)HF nu = 0.98

Conclusion

This study validates this portable, low-cost, easy-to-operate, and low-complexity heart rate monitor for the study of HRV at rest, during the 6MW and during recovery in patients with HF; it can thus be explored as a useful tool in future prospective interventional studies.

Raw data for each patient.

(PDF) Click here for additional data file.
  36 in total

1.  ATS statement: guidelines for the six-minute walk test.

Authors: 
Journal:  Am J Respir Crit Care Med       Date:  2002-07-01       Impact factor: 21.405

2.  Depressed heart rate variability as an independent predictor of death in chronic congestive heart failure secondary to ischemic or idiopathic dilated cardiomyopathy.

Authors:  P Ponikowski; S D Anker; T P Chua; R Szelemej; M Piepoli; S Adamopoulos; K Webb-Peploe; D Harrington; W Banasiak; K Wrabec; A J Coats
Journal:  Am J Cardiol       Date:  1997-06-15       Impact factor: 2.778

3.  Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system.

Authors:  Matthias Weippert; Mohit Kumar; Steffi Kreuzfeld; Dagmar Arndt; Annika Rieger; Regina Stoll
Journal:  Eur J Appl Physiol       Date:  2010-03-12       Impact factor: 3.078

4.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Reproducibility of heart rate variability measures in patients with chronic heart failure.

Authors:  P Ponikowski; M Piepoli; A A Amadi; T P Chua; D Harrington; M Volterrani; R Colombo; G Mazzuero; A Giordano; A J Coats
Journal:  Clin Sci (Lond)       Date:  1996-10       Impact factor: 6.124

7.  Seventh report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure.

Authors:  Aram V Chobanian; George L Bakris; Henry R Black; William C Cushman; Lee A Green; Joseph L Izzo; Daniel W Jones; Barry J Materson; Suzanne Oparil; Jackson T Wright; Edward J Roccella
Journal:  Hypertension       Date:  2003-12-01       Impact factor: 10.190

8.  Agreement of two different methods for measurement of heart rate variability.

Authors:  Martin Radespiel-Tröger; Robert Rauh; Christine Mahlke; Tim Gottschalk; Michael Mück-Weymann
Journal:  Clin Auton Res       Date:  2003-04       Impact factor: 4.435

9.  Relation of short-term heart rate variability to incident heart failure (from the Multi-Ethnic Study of Atherosclerosis).

Authors:  Sidharth A Shah; Thomas Kambur; Cheeling Chan; David M Herrington; Kiang Liu; Sanjiv J Shah
Journal:  Am J Cardiol       Date:  2013-05-16       Impact factor: 2.778

10.  Reproducibility and relation to mean heart rate of heart rate variability in normal subjects and in patients with congestive heart failure secondary to coronary artery disease.

Authors:  D Van Hoogenhuyze; N Weinstein; G J Martin; J S Weiss; J W Schaad; X N Sahyouni; D Fintel; W J Remme; D H Singer
Journal:  Am J Cardiol       Date:  1991-12-15       Impact factor: 2.778

View more
  2 in total

1.  Different acquisition systems for heart rate variability analysis may lead to diverse outcomes.

Authors:  F A de Oliveira Júnior; R A Pereira; A S Silva; J L de Brito Alves; J H Costa-Silva; V A Braga; C M Balarini
Journal:  Braz J Med Biol Res       Date:  2022-02-04       Impact factor: 2.590

2.  Heart rate variability in children: methodological issues and clinical applications.

Authors:  Iván Rodríguez-Núñez; Natalia Rodríguez-Romero; Alejandro Álvarez; Lysien Zambrano; Glaucia Luciano da Veiga; Fernando Romero
Journal:  Arch Cardiol Mex       Date:  2022-04-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.