Evaluation of kidney motion and target localization in abdominal SBRT patients.

Marcus Sonier1, William Chu, Nafisha Lalani, Darby Erler, Patrick Cheung, Renee Korol.   

Abstract

The purpose of this study was to evaluate bilateral kidney and target translational/rotational intrafraction motion during stereotactic body radiation therapy treatment delivery of primary renal cell carcinoma and oligometastatic adrenal lesions for patients immobilized in the Elekta BodyFIX system. Bilateral kidney motion was assessed at midplane for 30 patients immobilized in a full-body dual-vacuum-cushion system with two patients immobilized via abdominal compression. Intrafraction motion was assessed for 15 patients using kilovoltage cone-beam computed tomography (kV-CBCT) datasets (n = 151) correlated to the planning CT. Patient positioning was corrected for translational and rotational misalign-ments using a robotic couch in six degrees of freedom if setup errors exceeded 1mm and 1°. Absolute bilateral kidney motion between inhale and exhale 4D CT imaging phases for left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions was 1.51 ± 1.00 mm, 8.10 ± 4.33 mm, and 3.08 ± 2.11 mm, respec-tively. Residual setup error determined across CBCT type (pretreatment, intrafrac-tion, and post-treatment) for x (LR), y (SI), and z (AP) translations was 0.63 ± 0.74 mm, 1.08± 1.38 mm, and 0.70 ± 1.00 mm; while for x (pitch), y (roll), and z (yaw) rotations was 0.24 ± 0.39°, 0.19 ± 0.34°, and 0.26 ± 0.43°, respectively. Targets were localized to within 2.1 mm and 0.8° 95% of the time. The frequency of misalignments in the y direction was significant (p < 0.05) when compared to the x and z directions with no significant difference in translations between IMRT and VMAT. This technique is robust using BodyFIX for patient immobilization and reproducible localization of kidney and adrenal targets and daily CBCT image guidance for correction of positional errors to maintain treatment accuracy.
© 2016 The Authors.

Entities:  

Mesh:

Year:  2016        PMID: 27929514      PMCID: PMC5690515          DOI: 10.1120/jacmp.v17i6.6406

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Motion management is an important consideration during radiotherapy treatment, particularly in the case of stereotactic body radiation therapy (SBRT) where a high dose per fraction is utilized. Organ motion due to breathing during treatment delivery can result in an organ at risk (OAR) entering the treatment beam and subsequently exceed the tolerance dose while the target may move outside of the treatment beam, restricting the dose it receives to less than the intended prescription dose. Various CT and MR studies report kidney motion due to shallow respiration to occur in a complex pattern with the largest movement in the superior–inferior direction, an average of 6.7–19.0 mm. , , , , , , In order to limit OAR and target motion, a variety of immobilization techniques are available and explained in detail by the report of AAPM Task Group 76, which contains the recommendation that in the cases where a range of motion is observed, respiratory management techniques should be employed. , In addition to internal motion, patient setup uncertainty and intrafraction movement during radiotherapy treatment limits the accuracy and effectiveness of the delivered treatment plan, particularly in the case of SBRT where small PTV margins are utilized and high doses result in longer treatment times. A direct result of these small margins in conjunction with high doses necessitates correcting rotational misalignments in order to achieve increased target localization accuracy and acceptable setup uncertainties reflective of the PTV margins applied. Thus, the specific aim of this study is to assess maximum kidney motion from 4D CT datasets and intrafraction target motion using kilovoltage cone‐beam computed tomography (kV‐CBCT) for patients immobilized in the BodyFIX system (Elekta AB, Stockholm, Sweden) with patient positioning corrected for both translational and rotational errors using a robotic couch in six degrees of freedom.

II. MATERIALS AND METHODS

This study was approved by the institutional ethics committee. Intrafraction motion was retrospectively evaluated by analyzing 4D CT datasets and kV‐CBCT registration records. Bilateral kidney motion was assessed for 30 patients (a combination of liver and kidney/adrenal SBRT patients) who were immobilized in the BodyFIX system (a molded full‐body vacuum bag into which the patient is vacuum sealed via a plastic sheet placed on the patient's anterior surface) with two patients immobilized in the BlueBAG (Elekta) with abdominal compression. Absolute motion was measured at midplane in the left–right, superior–inferior, and anterior–posterior directions using the inhale and exhale phases of the helical 4D CT acquired at simulation. Intrafraction motion was then determined from the kV‐CBCT setup records of residual error for a subset of 15 patients with renal cell carcinoma or adrenal metastases treated with either VMAT () or IMRT (). The planning CT and CBCT were registered using an automatic gray value match to the PTV and adjacent kidney with manual fine‐tuning adjustments made to ensure that both the target volume and bony anatomy between the image sets were correlated. CBCT registrations were reviewed by a radiation oncologist. For all cases, patient positioning was corrected using a robotic couch in six degrees of freedom if the setup errors exceeded a threshold of 1 mm and 1°. CBCTs were acquired and shifts applied if setup errors exceeded the acceptable threshold following the initial setup registration, pretreatment scan, and intrafraction scan if IMRT was used due to long treatment times. A post‐treatment CBCT scan was acquired to verify the patient's final position. Translational and rotational errors were recorded for each CBCT registration. In total, 151 CBCT records were included in the localization analysis. Residual patient errors were analyzed using analysis of variance (ANOVA) to evaluate directional differences and, provided the ANOVA was significant (), a post hoc analysis was used to evaluate pairwise comparisons. An unpaired t‐test was used to evaluate differences in patient positioning errors between IMRT and VMAT treatment groups. The 95% confidence interval (CI) was calculated for translations and rotations.

III. RESULTS

The evaluation of absolute bilateral kidney motion across the inhale and exhale 4D CT image acquisitions in the left–right, superior–inferior, and anterior–posterior directions represented as a mean and standard deviation showed breathing induced displacements of , and , respectively. Residual setup error, including intrafraction motion, determined by kV‐CBCT for x (left–right), y (superior–inferior), and z (anterior–posterior) translations was , and ; while for x (pitch), y (roll), and z (yaw) rotations was , and , respectively. The frequency of translational misalignments in the y direction was found to be statistically significant () when compared to the x and z directions while the rotational misalignments were statistically significant when considering intrafraction motion () as well as for IMRT vs. VMAT (), as illustrated in Fig. 1. Table 1 displays the frequency of misalignments for both translations and rotations broken down by direction and CBCT type. The 95% CI showed an overall localization accuracy of 2.1 mm (1.5 mm, 2.7 mm, and 2.0 mm in the x, y, and z directions, respectively) and 0.8° over 15 patients using a correction threshold of 1 mm and 1°.
Figure 1

Summary of translational and rotational positional errors based on 151 image registrations. The absolute errors are represented as a mean and standard deviation. The data were separated as: (a) x (LR), y (SI), and z (AP); (b) IMRT and VMAT; and (c) pre‐, intra‐, and post‐treatment CBCT. The * indicates statistical significance between the respective results at .

Table 1

Frequency of misalignments indicated by CBCT matching for all patients

Frequency (%)
Translations Rotations
Pretreatment277
X (LR/pitch)Intrafraction1121
Post‐treatment257
Pretreatment414
Y (SI/roll)Intrafraction320
Post‐treatment416
Pretreatment2613
Z (AP/yaw)Intrafraction217
Post‐treatment297
Summary of translational and rotational positional errors based on 151 image registrations. The absolute errors are represented as a mean and standard deviation. The data were separated as: (a) x (LR), y (SI), and z (AP); (b) IMRT and VMAT; and (c) pre‐, intra‐, and post‐treatment CBCT. The * indicates statistical significance between the respective results at . Frequency of misalignments indicated by CBCT matching for all patients

IV. DISCUSSION

The assessment of bilateral kidney motion for patients immobilized in the dual‐vacuum‐cushion system show the largest deviation in the superior–inferior direction with a subset of patients displaying translations larger than 10 mm, while the left–right and anterior–posterior translations were also substantial, albeit consistently lower. These sizeable motions agree with existing data concerning kidney movement due to breathing motion and corroborate the choice to use an ITV for target localization.(6,7,9–11) The residual setup error determined from the on‐line CBCT () alignment corresponds with the absolute motion of the kidneys where the greatest error in localization accuracy occurred in the y direction (superior–inferior) compared to the x (left–right) and z (anterior–posterior) directions, consistent with studies in the literature. , Despite patient immobilization in the dual‐vacuum‐cushion system this observation illustrates that the impact of breathing motion on target localization remains clinically significant and strategies to further reduce kidney motion will be investigated. Possible techniques involve breath‐hold techniques or respiration‐gated radiation therapy as a method to minimize the volume of tissue irradiated, geometrically sparing OARs. The main strength of this study is the assessment of both translational and rotational setup errors for kidney and adrenal SBRT patients immobilized in the BodyFix system. Patient rotations due to intrafraction motion and between IMRT and VMAT were not clinically significant due to the majority of the measurements falling below the correction threshold of 1°. However, despite rotational setup error remaining clinically insignificant with regard to intrafraction motion, it remains an important factor at initial patient setup where complex breathing motion reduces the accuracy of tattoo alignment to external markers and resulted in 84% and 43% of patients falling above the correction threshold of 1° and 2°, respectively; no patients exhibited a rotational error > 3°. The importance of correcting both translational and rotational setup error is highlighted in the use of steep dose gradients, such as in SBRT, where misalignments can more readily result in a decrease in target coverage and increase in OAR doses. Overall, the 95% confidence interval of 2.1 mm and 0.8° demonstrates that the isotropic PTV margin of 5 mm is safe. Breathing‐induced kidney and target intrafraction motion is greatest in the superior–inferior direction in patients positioned in the dual‐vacuum‐cushion immobilization system. In this system, an ITV contoured from the inhale, exhale, and average phases of the 4D CT dataset with a 5 mm isotropic PTV margin is safe to account for target volume intrafraction motion and patient setup accuracy. Evaluation of translational and rotational patient setup errors during highly conformal hypofractionated treatments is necessary to ensure accurate localization of highly mobile targets. Correction of both translational and rotational misalignments is necessary to achieve the improved dose delivery accuracy required of SBRT. Alternate motion management and treatment delivery strategies may further reduce kidney motion, reduce setup errors, and improve target localization.

COPYRIGHT

This work is licensed under a Creative Commons Attribution 3.0 Unported License.
  14 in total

1.  Respiration-induced motion of the kidneys in whole abdominal radiotherapy: implications for treatment planning and late toxicity.

Authors:  N R Ahmad; M S Huq; B W Corn
Journal:  Radiother Oncol       Date:  1997-01       Impact factor: 6.280

2.  Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study.

Authors:  Matthias Guckenberger; Juergen Meyer; Dirk Vordermark; Kurt Baier; Juergen Wilbert; Michael Flentje
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-07-01       Impact factor: 7.038

3.  Interfraction and respiratory organ motion during conformal radiotherapy in gastric cancer.

Authors:  Barbara Wysocka; Zahra Kassam; Gina Lockwood; James Brierley; Laura A Dawson; Carol Ann Buckley; David Jaffray; Bernard Cummings; John Kim; Rebecca Wong; Jolie Ringash
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-08-06       Impact factor: 7.038

4.  The use of dual vacuum stabilization device to reduce kidney motion for stereotactic radiotherapy planning.

Authors:  Daniel Pham; Tomas Kron; Colin Styles; May Whitaker; Mathias Bressel; Farshad Foroudi; Michal Schneider; Thomas Devereux; Kim Dang; Shankar Siva
Journal:  Technol Cancer Res Treat       Date:  2014-11-21

5.  Kidney mobility during respiration.

Authors:  L H Schwartz; J Richaud; L Buffat; E Touboul; M Schlienger
Journal:  Radiother Oncol       Date:  1994-07       Impact factor: 6.280

6.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

7.  Respiration-induced movement of the upper abdominal organs: a pitfall for the three-dimensional conformal radiation treatment of pancreatic cancer.

Authors:  Barbara Bussels; Laurence Goethals; Michel Feron; Didier Bielen; Steven Dymarkowski; Paul Suetens; Karin Haustermans
Journal:  Radiother Oncol       Date:  2003-07       Impact factor: 6.280

8.  Repositioning accuracy of a commercially available double-vacuum whole body immobilization system for stereotactic body radiation therapy.

Authors:  Martin Fuss; Bill J Salter; Premavathy Rassiah; D Cheek; Sean X Cavanaugh; Terence S Herman
Journal:  Technol Cancer Res Treat       Date:  2004-02

9.  Individually wide range of renal motion evaluated by four-dimensional computed tomography.

Authors:  Hideomi Yamashita; Mami Yamashita; Masahiko Futaguchi; Ryousuke Takenaka; Shino Shibata; Kentaro Yamamoto; Akihiro Nomoto; Akira Sakumi; Satoshi Kida; Yoshihiro Kaneko; Shigeharu Takenaka; Takashi Shiraki; Keiichi Nakagawa
Journal:  Springerplus       Date:  2014-03-07

10.  An analysis of respiratory induced kidney motion on four-dimensional computed tomography and its implications for stereotactic kidney radiotherapy.

Authors:  Shankar Siva; Daniel Pham; Suki Gill; Mathias Bressel; Kim Dang; Thomas Devereux; Tomas Kron; Farshad Foroudi
Journal:  Radiat Oncol       Date:  2013-10-26       Impact factor: 3.481

View more
  4 in total

1.  The impact of 2D cine MR imaging parameters on automated tumor and organ localization for MR-guided real-time adaptive radiotherapy.

Authors:  Martin J Menten; Martin F Fast; Andreas Wetscherek; Christopher M Rank; Marc Kachelrieß; David J Collins; Simeon Nill; Uwe Oelfke
Journal:  Phys Med Biol       Date:  2018-11-22       Impact factor: 3.609

Review 2.  Patient-specific, touch-based registration during robotic, image-guided partial nephrectomy.

Authors:  Naren Nimmagadda; James M Ferguson; Nicholas L Kavoussi; Bryn Pitt; Eric J Barth; Josephine Granna; Robert J Webster; S Duke Herrell
Journal:  World J Urol       Date:  2021-06-16       Impact factor: 4.226

3.  The Mobius AIRO mobile CT for image-guided proton therapy: Characterization & commissioning.

Authors:  Jasmine A Oliver; Omar A Zeidan; Sanford L Meeks; Amish P Shah; Jason Pukala; Patrick Kelly; Naren R Ramakrishna; Twyla R Willoughby
Journal:  J Appl Clin Med Phys       Date:  2017-04-24       Impact factor: 2.102

4.  Breath-hold versus mid-ventilation in SBRT of adrenal metastases.

Authors:  Melissa de Kuijer; Jaap van Egmond; E Kouwenhoven; Dieke Bruijn-Krist; Heleen Ceha; Mirjam Mast
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2019-12-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.