Jasmine A Oliver1, Omar A Zeidan1, Sanford L Meeks1, Amish P Shah1, Jason Pukala1, Patrick Kelly1, Naren R Ramakrishna1, Twyla R Willoughby1.
Abstract
PURPOSE: The purpose of this study was to characterize the Mobius AIRO Mobile CT System for localization and image-guided proton therapy. This is the first known application of the AIRO for proton therapy.Entities:
Keywords: AIRO; IGPT; mobile CT; proton therapy
Mesh:
Year: 2017 PMID: 28436155 PMCID: PMC5689854 DOI: 10.1002/acm2.12084
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Mevion‐AIRO Room Setup. (a) Treatment room with couch in setup position. (b) Treatment room with couch in imaging position. (c) Birds‐eye‐view of the treatment room. The AIRO is shown in scan mode with 100 cm max scan range and 51.2 cm image field‐of‐view. PA and Left Lateral kV imaging panels are also shown with associated track.
Edge CBCT & AIRO Helical CT clinical pre‐set protocols/techniques
| Edge CBCT techniques | AIRO helical CT protocols | ||||
|---|---|---|---|---|---|
| Techniques | kVp | mAs | Protocols | kVp | mAs |
| Head | 100 | 150 | Head | 120 | 326 |
| Thorax | 125 | 268 | Thorax | 120 | 68 |
| Pelvis | 125 | 1080 | Pelvis | 120 | 149 |
| Pelvis Obese | 140 | 1681 | Abdomen | 120 | 149 |
| Image Gently | 80 | 97 | Shoulder | 120 | 149 |
Image quality tests and calculation methods
| Image quality tests | Definition |
|---|---|
| Modulation Transfer Function (MTF) | The modulation of multiple ROIs in the phantom with various line bar patterns. As spatial frequency increases, ROI's modulation decreases. The modulation is normalized to its highest value and plotted. |
| Modulation |
|
| Scaling Discrepancy |
The maximum possible error in a measurement of distance. |
| Geometric Distortion | Verifies that correct distance measurement occurs in all regions of the CT image. |
| Overall Uniformity |
|
| Minimum Uniformity |
The lowest uniformity of all ROIs. Uniformity is calculated as:
|
| Maximum HU Deviation | The maximum deviation between mean HU value of an ROI subtracted from its defined reference HU absolute value of all calculated HU deviation on an image. |
| Contrast |
|
| High Contrast‐to‐Noise Ratio |
|
| Low Contrast‐to‐Noise Ratio |
|
Definitions derived from DoseLab v.6.6 User Manual.17
AIRO image quality results for varying reconstruction kernels
| Sharp | Standard | Soft | |
|---|---|---|---|
| Scaling discrepancy (mm): | 1.19 | 1.34 | 1.35 |
| Geometric distortion (mm): | 0.14 | 0.14 | 0.15 |
| Spatial resolution (50% MTF, lp/mm): | 0.26 | 0.21 | 0.19 |
| Overall uniformity (%): | 98.85 | 98.94 | 98.99 |
| Minimum uniformity (%): | 99.06 | 99.19 | 99.31 |
| Contrast (%): | 10.68 | 10.56 | 10.41 |
| High CNR: | 30.34 | 21.45 | 17.49 |
| Low CNR: | 5.73 | 7.12 | 8.51 |
| Maximum HU deviation (HU): | 58.98 | 57.50 | 57.68 |
Figure 2AIRO Normalized MTF for sharp, standard, and soft reconstruction kernels—calculated with DoseLAB v6.6.
Figure 3Normalized MTF for Philips CT simulator, AIRO, and EDGE CBCT—calculated with DoseLAB v6.6.
Image quality results—AIRO, Philips CT simulator, Siemens CT simulator and Edge CBCT (DoseLAB)
| Philips | Airo | Siemens | Edge | |
|---|---|---|---|---|
| Scaling discrepancy (mm): | 0.21 | 1.34 | 0.03 | 0.86 |
| Geometric distortion (mm): | 0.11 | 0.14 | 0.11 | 0.13 |
| Spatial resolution (50% MTF, lp/mm): | 0.40 | 0.21 | 0.35 | 0.37 |
| Overall uniformity (%): | 99.21 | 98.94 | 99.37 | 98.93 |
| Minimum uniformity (%): | 99.25 | 99.19 | 99.37 | 99.26 |
| Contrast (%): | 10.30 | 10.56 | 12.03 | 10.79 |
| High CNR: | 21.14 | 21.45 | 28.47 | 23.36 |
| Low CNR: | 2.00 | 7.12 | 2.46 | – |
| Maximum HU deviation (HU): | 48.50 | 57.5 | 89.49 | 21.66 |
EDGE CBCT images were acquired with the Pelvis protocol.
HU Constancy (CTP404) for various materials in Philips, Edge, Siemens and AIRO
| Philips (HU) | Edge (HU) | Airo (HU) | Siemens (HU) | True (HU) | |
|---|---|---|---|---|---|
| Acrylic | 132.44 | 123.08 | 108.15 | 129.67 | 100 |
| Air | −969.86 | −999.41 | −978.49 | −1023.34 | −1000 |
| Polystyrene | −22.69 | −36.62 | −47.83 | −52.26 | −35 |
| LDPE | −79.01 | −94.88 | −102.08 | −119.95 | −100 |
| PMP | −167.91 | −186.91 | −191.44 | −223.49 | −200 |
| Teflon | 941.50 | 968.34 | 932.32 | 1079.30 | 990 |
| CTP404 | 103.64 | 106.06 | 92.99 | 96.00 | 100 |
Measured CTDIvol for AIRO helical CT and Philips CT simulator for head and abdomen protocols
| CTDIvol (mGy) | ||
|---|---|---|
| Airo | Philips | |
| HEAD | 72.05 | 44.91 |
| ABDOMEN | 26.08 | 24.09 |