Technical evaluation of a laser-based optical surface scanning system for prospective and retrospective breathing adapted computed tomography.

Mattias Jönsson1, Sofie Ceberg, Fredrik Nordström, Charlotte Thornberg, Sven Å J Bäck.   

Abstract

BACKGROUND: For breathing adapted radiotherapy, the same motion monitoring system can be used for imaging and triggering of the accelerator.
PURPOSE: To evaluate a new technique for prospective gated computed tomography (CT) and four-dimensional CT (4DCT) using a laser based surface scanning system (Sentinel(™), C-RAD, Uppsala, Sweden). The system was compared to the AZ-733V respiratory gating system (Anzai Medical, Tokyo, Japan) and the Real-Time Position Management System (RPM(™)) (Varian Medical Systems, Palo Alto, CA, USA).
MATERIAL AND METHODS: Temporal accuracy was evaluated using a moving phantom programmed to move a platform along trajectories following a sin(6)(ωt) function with amplitudes from 6 to 20 mm and periods from 2 to 5 s during 120 s while the motion was recorded. The recorded data was Fourier transformed and the peak area at the fundamental and harmonic frequencies compared to data generated using the same sinusoidal function. For verification of the 4DCT reconstruction process, the phantom was programmed to move along a sinusoidal trajectory. Ten phase series were reconstructed. The distance from the couch to the platform was measured in each image. By fitting the function sin(ωt-ϕ) to the values measured in the images corresponding to each slice, the phase of each image was verified. RESULTS AND
CONCLUSION: In the recorded data, the peak area at the fundamental frequency covered on average 104 ± 4%, 102 ± 4% and 91 ± 27% of the peak area in the generated data for the Sentinel(™), RPM(™) and AZ-733V systems, respectively. All systems managed to resolve both harmonic frequencies. The second experiment showed that all images were sorted into the correct series using breathing data recorded by each system. The systems generated very similar results, however, it is preferable to use the same system both for imaging and treatment.

Mesh:

Year:  2014        PMID: 25383452     DOI: 10.3109/0284186X.2014.948059

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  4 in total

1.  3D Kinect Camera Scheme with Time-Series Deep-Learning Algorithms for Classification and Prediction of Lung Tumor Motility.

Authors:  Utumporn Puangragsa; Jiraporn Setakornnukul; Pittaya Dankulchai; Pattarapong Phasukkit
Journal:  Sensors (Basel)       Date:  2022-04-11       Impact factor: 3.847

2.  Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation.

Authors:  Barbara Zollner; Christian Heinz; Sabrina Pitzler; Farkhad Manapov; Steffi Kantz; Maya Christine Rottler; Maximilian Niyazi; Ute Ganswindt; Claus Belka; Hendrik Ballhausen
Journal:  Radiat Oncol       Date:  2016-12-07       Impact factor: 3.481

Review 3.  AAPM task group report 302: Surface-guided radiotherapy.

Authors:  Hania A Al-Hallaq; Laura Cerviño; Alonso N Gutierrez; Amanda Havnen-Smith; Susan A Higgins; Malin Kügele; Laura Padilla; Todd Pawlicki; Nicholas Remmes; Koren Smith; Xiaoli Tang; Wolfgang A Tomé
Journal:  Med Phys       Date:  2022-03-15       Impact factor: 4.506

Review 4.  Recent advanced in Surface Guided Radiation Therapy.

Authors:  P Freislederer; M Kügele; M Öllers; A Swinnen; T-O Sauer; C Bert; D Giantsoudi; S Corradini; V Batista
Journal:  Radiat Oncol       Date:  2020-07-31       Impact factor: 3.481

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.