Estimation of Eye Lens Dose During Brain Scans Using Gafchromic Xr-QA2 Film in Various Multidetector CT Scanners.

Philomina Akhilesh1, Arti R Kulkarni2, Shramika H Jamhale1, S D Sharma1, Rajesh Kumar1, D Datta1.   

Abstract

The purpose of this study was to estimate eye lens dose during brain scans in 16-, 64-, 128- and 256-slice multidetector computed tomography (CT) scanners in helical acquisition mode and to test the feasibility of using radiochromic film as eye lens dosemeter during CT scanning. Eye lens dose measurements were performed using Gafchromic XR-QA2 film on a polystyrene head phantom designed with outer dimensions equivalent to the head size of a reference Indian man. The response accuracy of XR-QA2 film was validated by using thermoluminescence dosemeters. The eye lens dose measured using XR-QA2 film on head phantom for plain brain scanning in helical mode ranged from 43.8 to 45.8 mGy. The XR-QA2 film measured dose values were in agreement with TLD measured dose values within a maximum variation of 8.9%. The good correlation between the two data sets confirms the viability of using XR-QA2 film for eye lens dosimetry.
© The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2017        PMID: 27247444     DOI: 10.1093/rpd/ncw132

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Signal-to-noise ratio and dose to the lens of the eye for computed tomography examination of the brain using an automatic tube current modulation system.

Authors:  Supawitoo Sookpeng; Chitsanupong Butdee
Journal:  Emerg Radiol       Date:  2016-12-02
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.