Philomina Akhilesh1, Arti R Kulkarni2, Shramika H Jamhale1, S D Sharma1, Rajesh Kumar1, D Datta1.
Abstract
The purpose of this study was to estimate eye lens dose during brain scans in 16-, 64-, 128- and 256-slice multidetector computed tomography (CT) scanners in helical acquisition mode and to test the feasibility of using radiochromic film as eye lens dosemeter during CT scanning. Eye lens dose measurements were performed using Gafchromic XR-QA2 film on a polystyrene head phantom designed with outer dimensions equivalent to the head size of a reference Indian man. The response accuracy of XR-QA2 film was validated by using thermoluminescence dosemeters. The eye lens dose measured using XR-QA2 film on head phantom for plain brain scanning in helical mode ranged from 43.8 to 45.8 mGy. The XR-QA2 film measured dose values were in agreement with TLD measured dose values within a maximum variation of 8.9%. The good correlation between the two data sets confirms the viability of using XR-QA2 film for eye lens dosimetry.Entities:
Mesh:
Year: 2017 PMID: 27247444 DOI: 10.1093/rpd/ncw132