Local filtration based scatter correction for cone-beam CT using primary modulation.

Lei Zhu1.   

Abstract

PURPOSE: Excessive scatter contamination fundamentally limits the image quality of cone-beam CT (CBCT), hindering its quantitative use in clinical applications. The author has previously proposed an effective scatter correction method for CBCT using primary modulation. A Fourier transform-based algorithm (FTPM) was implemented to estimate scatter from modulated projections, with a few limitations including the assumption of uniform modulation frequency and magnitude that becomes less accurate in the presence of beam-hardening and other nonideal effects. This paper aims to overcome the above drawbacks by developing a new algorithm for the primary modulation method with improved accuracy and reliability.
METHODS: Incident x-ray intensities for each detector pixel with and without the interception of the modulator blocker are estimated from a modulated flat-field image. A new signal relationship is then developed to obtain a first scatter estimate from a modulated projection using a spatially varying modulation distribution. The method empirically adjusts the effective modulation magnitude for each projection ray to account for the beam-hardening effects. Estimated scatter signals with high expected errors are discarded in the generation of the final scatter distribution. The author proposes a technique of local filtration to accelerate major portions of the signal processing, and the new algorithm is referred to as local filtration based primary modulation (LFPM).
RESULTS: The study on the Catphan® 600 phantom shows that LFPM effectively removes scatter-induced cupping artifacts on CBCT images and reduces the CT image error from 222 to 15 HU. In addition, the image contrast on eight contrast rods of the phantom is enhanced by a factor of 2 on average. On an anthropomorphic head phantom, LFPM reduces the CT image error from 153 to 18 HU and eliminates the streak artifacts observed on the result of FTPM with substantially improved image uniformity. On the Rando® phantom, LFPM reduces the CT image error from 278 to 4 HU around the object center.
CONCLUSIONS: As compared with the previously developed FTPM algorithm, LFPM enhances the imaging performance by using a more flexible data processing framework that does not require projection data downsampling or uniform modulation frequency and magnitude. It also becomes possible to discard suspicious scatter estimate values prior to the generation of a final scatter distribution and to model the beam-hardening effects on modulation for improved scatter estimation accuracy. The presented research further exploits the potential of the primary modulation method on scatter correction and facilitates its clinical adoption in CBCT imaging.

Mesh:

Year:  2016        PMID: 27806607      PMCID: PMC5097053          DOI: 10.1118/1.4965042

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography).

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

2.  Iterative CT shading correction with no prior information.

Authors:  Pengwei Wu; Xiaonan Sun; Hongjie Hu; Tingyu Mao; Wei Zhao; Ke Sheng; Alice A Cheung; Tianye Niu
Journal:  Phys Med Biol       Date:  2015-10-14       Impact factor: 3.609

3.  Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT.

Authors:  Yiannis Kyriakou; Thomas Riedel; Willi A Kalender
Journal:  Phys Med Biol       Date:  2006-08-30       Impact factor: 3.609

4.  Magnitude and effects of x-ray scatter in a 256-slice CT scanner.

Authors:  Masahiro Endo; Shinichiro Mori; Takanori Tsunoo; Hiroaki Miyazaki
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

5.  Scatter correction for cone-beam CT in radiation therapy.

Authors:  Lei Zhu; Yaoqin Xie; Jing Wang; Lei Xing
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

6.  Scatter correction method for x-ray CT using primary modulation: phantom studies.

Authors:  Hewei Gao; Rebecca Fahrig; N Robert Bennett; Mingshan Sun; Josh Star-Lack; Lei Zhu
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

7.  Robust primary modulation-based scatter estimation for cone-beam CT.

Authors:  Ludwig Ritschl; Rebecca Fahrig; Michael Knaup; Joscha Maier; Marc Kachelrieß
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

8.  Optimization-based scatter estimation using primary modulation for computed tomography.

Authors:  Yi Chen; Ying Song; Jingchen Ma; Jun Zhao
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

9.  Monte Carlo simulation of the scattered radiation distribution in diagnostic radiology.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1988 Sep-Oct       Impact factor: 4.071

10.  Image-domain shading correction for cone-beam CT without prior patient information.

Authors:  Qiyong Fan; Bo Lu; Justin C Park; Tianye Niu; Jonathan G Li; Chihray Liu; Lei Zhu
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

View more
  5 in total

1.  Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.

Authors:  Xi Chen; Luo Ouyang; Hao Yan; Xun Jia; Bin Li; Qingwen Lyu; You Zhang; Jing Wang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

2.  Fast shading correction for cone-beam CT via partitioned tissue classification.

Authors:  Linxi Shi; Adam Wang; Jikun Wei; Lei Zhu
Journal:  Phys Med Biol       Date:  2019-03-13       Impact factor: 3.609

3.  4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction.

Authors:  Cong Zhao; Yuncheng Zhong; Xinhui Duan; You Zhang; Xiaokun Huang; Jing Wang; Mingwu Jin
Journal:  Phys Med Biol       Date:  2018-05-29       Impact factor: 3.609

4.  The role of off-focus radiation in scatter correction for dedicated cone beam breast CT.

Authors:  Linxi Shi; Srinivasan Vedantham; Andrew Karellas; Lei Zhu
Journal:  Med Phys       Date:  2017-12-16       Impact factor: 4.071

5.  Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT.

Authors:  Weihua Mao; Chang Liu; Stephen J Gardner; Farzan Siddiqui; Karen C Snyder; Akila Kumarasiri; Bo Zhao; Joshua Kim; Ning Winston Wen; Benjamin Movsas; Indrin J Chetty
Journal:  Technol Cancer Res Treat       Date:  2019-01-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.