Optimization-based scatter estimation using primary modulation for computed tomography.

Yi Chen1, Ying Song2, Jingchen Ma1, Jun Zhao1.   

Abstract

PURPOSE: Scatter reduces the image quality in computed tomography (CT), but scatter correction remains a challenge. A previously proposed primary modulation method simultaneously obtains the primary and scatter in a single scan. However, separating the scatter and primary in primary modulation is challenging because it is an underdetermined problem. In this study, an optimization-based scatter estimation (OSE) algorithm is proposed to estimate and correct scatter.
METHODS: In the concept of primary modulation, the primary is modulated, but the scatter remains smooth by inserting a modulator between the x-ray source and the object. In the proposed algorithm, an objective function is designed for separating the scatter and primary. Prior knowledge is incorporated in the optimization-based framework to improve the accuracy of the estimation: (1) the primary is always positive; (2) the primary is locally smooth and the scatter is smooth; (3) the location of penumbra can be determined; and (4) the scatter-contaminated data provide knowledge about which part is smooth.
RESULTS: The simulation study shows that the edge-preserving weighting in OSE improves the estimation accuracy near the object boundary. Simulation study also demonstrates that OSE outperforms the two existing primary modulation algorithms for most regions of interest in terms of the CT number accuracy and noise. The proposed method was tested on a clinical cone beam CT, demonstrating that OSE corrects the scatter even when the modulator is not accurately registered.
CONCLUSIONS: The proposed OSE algorithm improves the robustness and accuracy in scatter estimation and correction. This method is promising for scatter correction of various kinds of x-ray imaging modalities, such as x-ray radiography, cone beam CT, and the fourth-generation CT.

Entities:  

Mesh:

Year:  2016        PMID: 27487893     DOI: 10.1118/1.4958680

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Local filtration based scatter correction for cone-beam CT using primary modulation.

Authors:  Lei Zhu
Journal:  Med Phys       Date:  2016-11       Impact factor: 4.071

2.  Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations.

Authors:  Amy E Becker; Andrew M Hernandez; Paul R Schwoebel; John M Boone
Journal:  Phys Med Biol       Date:  2020-12-18       Impact factor: 3.609

3.  Image quality improvement in cone-beam CT using the super-resolution technique.

Authors:  Asuka Oyama; Shinobu Kumagai; Norikazu Arai; Takeshi Takata; Yusuke Saikawa; Kenshiro Shiraishi; Takenori Kobayashi; Jun'ichi Kotoku
Journal:  J Radiat Res       Date:  2018-07-01       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.