Literature DB >> 27782691

Technical Note: scuda: A software platform for cumulative dose assessment.

Seyoun Park1, Todd McNutt1, William Plishker2, Harry Quon1, John Wong1, Raj Shekhar3, Junghoon Lee1.   

Abstract

PURPOSE: Accurate tracking of anatomical changes and computation of actually delivered dose to the patient are critical for successful adaptive radiation therapy (ART). Additionally, efficient data management and fast processing are practically important for the adoption in clinic as ART involves a large amount of image and treatment data. The purpose of this study was to develop an accurate and efficient Software platform for CUmulative Dose Assessment (scuda) that can be seamlessly integrated into the clinical workflow.
METHODS: scuda consists of deformable image registration (DIR), segmentation, dose computation modules, and a graphical user interface. It is connected to our image PACS and radiotherapy informatics databases from which it automatically queries/retrieves patient images, radiotherapy plan, beam data, and daily treatment information, thus providing an efficient and unified workflow. For accurate registration of the planning CT and daily CBCTs, the authors iteratively correct CBCT intensities by matching local intensity histograms during the DIR process. Contours of the target tumor and critical structures are then propagated from the planning CT to daily CBCTs using the computed deformations. The actual delivered daily dose is computed using the registered CT and patient setup information by a superposition/convolution algorithm, and accumulated using the computed deformation fields. Both DIR and dose computation modules are accelerated by a graphics processing unit.
RESULTS: The cumulative dose computation process has been validated on 30 head and neck (HN) cancer cases, showing 3.5 ± 5.0 Gy (mean±STD) absolute mean dose differences between the planned and the actually delivered doses in the parotid glands. On average, DIR, dose computation, and segmentation take 20 s/fraction and 17 min for a 35-fraction treatment including additional computation for dose accumulation.
CONCLUSIONS: The authors developed a unified software platform that provides accurate and efficient monitoring of anatomical changes and computation of actually delivered dose to the patient, thus realizing an efficient cumulative dose computation workflow. Evaluation on HN cases demonstrated the utility of our platform for monitoring the treatment quality and detecting significant dosimetric variations that are keys to successful ART.

Entities:  

Mesh:

Year:  2016        PMID: 27782691      PMCID: PMC5018004          DOI: 10.1118/1.4961985

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

Review 1.  Intensity-modulated radiotherapy: current status and issues of interest.

Authors: 
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-11-15       Impact factor: 7.038

2.  CERR: a computational environment for radiotherapy research.

Authors:  Joseph O Deasy; Angel I Blanco; Vanessa H Clark
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

3.  Image registration using hierarchical B-splines.

Authors:  Zhiyong Xie; Gerald E Farin
Journal:  IEEE Trans Vis Comput Graph       Date:  2004 Jan-Feb       Impact factor: 4.579

4.  Deformable planning CT to cone-beam CT image registration in head-and-neck cancer.

Authors:  Jidong Hou; Mariana Guerrero; Wenjuan Chen; Warren D D'Souza
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  Demons deformable registration of CT and cone-beam CT using an iterative intensity matching approach.

Authors:  Sajendra Nithiananthan; Sebastian Schafer; Ali Uneri; Daniel J Mirota; J Webster Stayman; Wojciech Zbijewski; Kristy K Brock; Michael J Daly; Harley Chan; Jonathan C Irish; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

6.  Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy.

Authors:  Joan Hatton; Boyd McCurdy; Peter B Greer
Journal:  Phys Med Biol       Date:  2009-07-10       Impact factor: 3.609

7.  Validation of an accelerated 'demons' algorithm for deformable image registration in radiation therapy.

Authors:  He Wang; Lei Dong; Jennifer O'Daniel; Radhe Mohan; Adam S Garden; K Kian Ang; Deborah A Kuban; Mark Bonnen; Joe Y Chang; Rex Cheung
Journal:  Phys Med Biol       Date:  2005-06-01       Impact factor: 3.609

Review 8.  Intensity-modulated radiation therapy in head and neck cancers: an update.

Authors:  Nancy Lee; Dev R Puri; Angel I Blanco; K S Clifford Chao
Journal:  Head Neck       Date:  2007-04       Impact factor: 3.147

9.  CT to cone-beam CT deformable registration with simultaneous intensity correction.

Authors:  Xin Zhen; Xuejun Gu; Hao Yan; Linghong Zhou; Xun Jia; Steve B Jiang
Journal:  Phys Med Biol       Date:  2012-10-03       Impact factor: 3.609

10.  Multicriteria optimization in intensity-modulated radiation therapy treatment planning for locally advanced cancer of the pancreatic head.

Authors:  Theodore S Hong; David L Craft; Fredrik Carlsson; Thomas R Bortfeld
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-11-15       Impact factor: 7.038

View more
  1 in total

1.  Evaluation of a software module for adaptive treatment planning and re-irradiation.

Authors:  Anne Richter; Stefan Weick; Thomas Krieger; Florian Exner; Sonja Kellner; Bülent Polat; Michael Flentje
Journal:  Radiat Oncol       Date:  2017-12-28       Impact factor: 3.481

  1 in total