Combined photoplethysmographic monitoring of respiration rate and pulse: a comparison between different measurement sites in spontaneously breathing subjects.

L Nilsson1, T Goscinski, S Kalman, L-G Lindberg, A Johansson.   

Abstract

BACKGROUND: The non-invasive photoplethysmographic (PPG) signal reflects blood flow and volume in a tissue. The PPG signal shows variation synchronous with heartbeat (PPGc), as used in pulse oximetry, and variations synchronous with breathing (PPGr). PPGr has been used for non-invasive monitoring of respiration with promising results. Our aim was to investigate PPG signals recorded from different skin sites in order to find suitable locations for parallel monitoring of variations synchronous with heartbeat and breathing.
METHODS: PPG sensors were applied to the forearm, finger, forehead, wrist and shoulder on 48 awake healthy volunteers. From these sites, seven PPG signals were simultaneously recorded during normal spontaneous breathing over 10 min. Capnometry served as respiration and electrocardiogram (ECG) as pulse reference signals. PPG signals were compared with respect to power spectral content and squared coherence.
RESULTS: Forearm PPG measurement showed significantly higher power within the respiratory region of the power spectrum [median (quartile range) 42 (26)%], but significantly lower power within the cardiac region [9 (10)%] compared with the other skin sites. PPG finger measurement showed the opposite; in transmission mode, the power within the respiratory region was significantly lower [4 (10)%] and within the cardiac region significantly higher [45 (25)%] than the other sites. PPGc coherence values were generally high [>0.96 (0.08)], and PPGr coherence values lower [0.83 (0.35)-0.94 (0.17)].
CONCLUSION: Combined PPG respiration and pulse monitoring is possible, but there are significant differences between the respiratory and cardiac components of the PPG signal at different sites.

Entities:  

Mesh:

Year:  2007        PMID: 17711563     DOI: 10.1111/j.1399-6576.2007.01375.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  16 in total

1.  Determination of saturation, heart rate, and respiratory rate at forearm using a Nellcor™ forehead SpO2-saturation sensor.

Authors:  Jarkko Harju; Antti Vehkaoja; Ville Lindroos; Pekka Kumpulainen; Sasu Liuhanen; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2016-10-17       Impact factor: 2.502

2.  Photoplethysmography respiratory rate monitoring in patients receiving procedural sedation and analgesia for upper gastrointestinal endoscopy.

Authors:  Hugo R W Touw; Milou H Verheul; Pieter R Tuinman; Jeroen Smit; Deirdre Thöne; Patrick Schober; Christa Boer
Journal:  J Clin Monit Comput       Date:  2016-05-28       Impact factor: 2.502

3.  Photoplethysmography for blood volumes and oxygenation changes during intermittent vascular occlusions.

Authors:  T Y Abay; P A Kyriacou
Journal:  J Clin Monit Comput       Date:  2017-05-25       Impact factor: 2.502

4.  Detection of Atrial Fibrillation Using a Ring-Type Wearable Device (CardioTracker) and Deep Learning Analysis of Photoplethysmography Signals: Prospective Observational Proof-of-Concept Study.

Authors:  Soonil Kwon; Joonki Hong; Eue-Keun Choi; Byunghwan Lee; Changhyun Baik; Euijae Lee; Eui-Rim Jeong; Bon-Kwon Koo; Seil Oh; Yung Yi
Journal:  J Med Internet Res       Date:  2020-05-21       Impact factor: 5.428

5.  Design Rationale and Performance Evaluation of the Wavelet Health Wristband: Benchtop Validation of a Wrist-Worn Physiological Signal Recorder.

Authors:  Onur Dur; Reinier van Mourik; Colleen Rhoades; Man Suen Ng; Ragwa Elsayed; Maulik D Majmudar
Journal:  JMIR Mhealth Uhealth       Date:  2018-10-16       Impact factor: 4.773

6.  Quantifying the Delays Between Multi-Site Photoplethysmography Pulse and Electrocardiogram R-R Interval Changes Under Slow-Paced Breathing.

Authors:  John Allen
Journal:  Front Physiol       Date:  2019-09-25       Impact factor: 4.566

7.  Development of a Brain-Computer Interface Toggle Switch with Low False-Positive Rate Using Respiration-Modulated Photoplethysmography.

Authors:  Chang-Hee Han; Euijin Kim; Chang-Hwan Im
Journal:  Sensors (Basel)       Date:  2020-01-08       Impact factor: 3.576

8.  Cross time-frequency analysis for combining information of several sources: application to estimation of spontaneous respiratory rate from photoplethysmography.

Authors:  M D Peláez-Coca; M Orini; J Lázaro; R Bailón; E Gil
Journal:  Comput Math Methods Med       Date:  2013-12-01       Impact factor: 2.238

9.  To clarify features of photoplethysmography in monitoring balanced anesthesia, compared with Cerebral State Index.

Authors:  Lieliang Zhang; Lei Xu; Juan Zhu; Yujie Gao; Zhonghua Luo; Hongyu Wang; Zhongliang Zhu; Yi Yu; Hongwei Shi; Hongguang Bao
Journal:  Med Sci Monit       Date:  2014-03-25

10.  Investigation of Photoplethysmography Behind the Ear for Pulse Oximetry in Hypoxic Conditions with a Novel Device (SPYDR).

Authors:  Brian Bradke; Bradford Everman
Journal:  Biosensors (Basel)       Date:  2020-04-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.