| Literature DB >> 27695955 |
Mathieu Mouhat1, James Mercer2, Lina Stangvaltaite1, Ulf Örtengren3,4.
Abstract
OBJECTIVES: To investigate how heat development in the pulp chamber and coronal surface of natural teeth with and without cusps subjected to irradiance using light-emitting diode (LED)-light-curing units (LCUs) is associated with (i) irradiance, (ii) time, (iii) distance, and (iv) radiant exposure.Entities:
Keywords: Curing lights; Dentistry; Light; Temperature; Tooth
Mesh:
Year: 2016 PMID: 27695955 PMCID: PMC5442227 DOI: 10.1007/s00784-016-1962-5
Source DB: PubMed Journal: Clin Oral Investig ISSN: 1432-6981 Impact factor: 3.573
Fig. 1a, b The placement of a tooth in the thin plastic sheet with the root visible on one side and the crown on the other are shown. During an experiment, the thin plastic sheet with the mounted tooth was placed on the water surface, such that the root was submerged in the water. c The position of the LED-LCU being tested. Overview of the experimental setup (d) showing the position of the infrared camera over the opening of the water bath during an experiment
Mean (SD) for the irradiance of the two different light-curing units tested, Bluephase style® battery (n = 10) and Bluephase G2® high mode (n = 10), at three different curing times
| Curing time | Irradiance of Bluephase style® battery in mW/cm2 | Irradiance of Bluephase G2® high mode in mW/cm2 |
|---|---|---|
| 10 s | 1337 (104) ade | 1411 (142) a |
| 20 s | 1477 (240) bd | 1362 (121) b |
| 30 s | 1479 (96) ce | 1382 (102) c |
The letters a-c indicate significant difference (p < 0.05) between Bluephase G2® and Bluephase style@ at the different times tested respectively. The letter d-e indicates significant difference (p < 0.05) in irradiance at different times for Bluephase style®. Wavelength for the two light-curing units was 385–515 nm
Fig. 2Thermograms showing temperature distribution on the surface of a tooth when subjected to a 30-s irradiation with a LED light-curing unit Bluephase style®. a Thermogram of tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2). b Thermogram of tooth with class I cavity (T1)
Temperature variations at the surface and in the pulp chamber at different radiant exposure (product of irradiance and time (s)). Distance from the surface (top of the cusp from tooth with class I cavity (T1)) light-curing unit tip 0 mm. The table shows the results for T1
| Light-curing unit (irradiance—mW/cm2) | Radiant exposure (J/cm2) | Surface temperature (°C) | Pulp chamber temperature (°C) | Time (s) |
|---|---|---|---|---|
| Bluephase style plugged (1184 ± 12) | 35.9 ± 0.4 | 49.3 ± 0.4 | 37.6 ± 0.1 | 30 |
| 23.9 ± 0.3 | 45.6 ± 0.1 | 37.2 ± 0 | 20 | |
| 12 ± 0.2 | 41.2 ± 0.2 | 36.9 ± 0 | 10 | |
| Bluephase G2 high mode (1471 ± 65) | 44.1 ± 1.9 | 53.1 ± 0.3 | 37.9 ± 0.1 | 30 |
| 29.4 ± 1.3 | 48.2 ± 0.4 | 37.4 ± 0.1 | 20 | |
| 14.6 ± 0.8 | 43.7 ± 0.3 | 37.1 ± 0 | 10 | |
| Bluephase G2 low mode (767 ± 46) | 23.0 ± 1.4 | 45.7 ± 0.2 | 37.3 ± 0 | 30 |
| 15.3 ± 0.9 | 42.8 ± 0.6 | 37.1 ± 0.1 | 20 | |
| 7.7 ± 0.5 | 39.6 ± 0.3 | 36.9 ± 0 | 10 | |
| Bluephase style battery (1268 ± 21) | 38.1 ± 0.6 | 49.5 ± 0.3 | 37.6 ± 0.1 | 30 |
| 25.4 ± 0.4 | 46.3 ± 0.3 | 37.3 ± 0 | 20 | |
| 12.7 ± 0.2 | 43.1 ± 0.4 | 37.1 ± 0 | 10 |
Temperature variations at the surface and in the pulp chamber at different radiant exposure (product of irradiance and time (s)). The table shows the results for the tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2). Distance from the surface to the light curing unit tip 0 mm
| Light-curing unit (irradiance—mW/cm2) | Radiant exposure (J/cm2) | Surface temperature (°C) | Pulp chamber temperature (°C) | Time (s) |
|---|---|---|---|---|
| Bluephase style plugged (1257 ± 4) | 37.8 ± 0.1 | 44.4 ± 0.8 | 40.8 ± 0.2 | 30 |
| 25.1 ± 0.1 | 42.5 ± 0.4 | 39.2 ± 0.1 | 20 | |
| 12.6 ± 0.1 | 40.1 ± 0.3 | 38.4 ± 0.1 | 10 | |
| Bluephase G2 high mode (1437 ± 14) | 43.1 ± 0.4 | 46.9 ± 1.2 | 43.1 ± 0.9 | 30 |
| 28.7 ± 0.3 | 45.3 ± 1.2 | 41.4 ± 0.3 | 20 | |
| 14.6 ± 0.1 | 42.5 ± 0.3 | 39.1 ± 0.2 | 10 | |
| Bluephase G2 low mode (774 ± 7) | 23.2 ± 0.2 | 41.5 ± 0.3 | 39.1 ± 0.2 | 30 |
| 15.5 ± 0.1 | 40.3 ± 0.3 | 38.5 ± 0.2 | 20 | |
| 7.7 ± 0.1 | 38.1 ± 0.2 | 37.2 ± 0.1 | 10 | |
| Bluephase style battery (1222 ± 9) | 36.7 ± 0.3 | 52.6 ± 0.5 | 42.5 ± 0.4 | 30 |
| 24.4 ± 0.2 | 52.5 ± 0.4 | 41.5 ± 0.3 | 20 | |
| 12.2 ± 0.1 | 48.6 ± 0.7 | 40.0 ± 0.1 | 10 |
Irradiance, radiant exposure (product of irradiance and time (s)) and heat development (surface temperature and pulp chamber temperature) at different distances (on tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2)) with different light-curing units, Bluephase G2® high mode
| Distance (mm) | Irradiance (mW/mm2) | Time (s) | Radiant exposure (J/mm2) | Surface temperature (°C) | Pulp chamber temperature (°C) |
|---|---|---|---|---|---|
| 0 | 1437 ± 14 | 30 | 43.1 ± 0.4 | 46.9 ± 1.2 | 43.1 ± 0.9 |
| 20 | 28.7 ± 0.3 | 45.3 ± 1.2 | 41.4 ± 0.3 | ||
| 10 | 14.6 ± 0.1 | 42.5 ± 0.3 | 39.1 ± 0.2 | ||
| 2 | 1412 ± 16 | 30 | 42.4 ± 0.5 | 52.2 ± 0.6 | 42.7 ± 0.6 |
| 20 | 28.3 ± 0.3 | 49.1 ± 0.2 | 41.9 ± 0.3 | ||
| 10 | 14.1 ± 0.2 | 44.6 ± 0.8 | 39.1 ± 0.2 | ||
| 4 | 1513 ± 16 | 30 | 45.4 ± 0.5 | 48.7 ± 0.5 | 40.7 ± 0.8 |
| 20 | 30.3 ± 0.3 | 46.2 ± 1.1 | 39.3 ± 0.3 | ||
| 10 | 15.1 ± 0.2 | 42.8 ± 0.3 | 37.6 ± 0.1 |
Irradiance, radiant exposure (product of irradiance and time (s)), and heat development (surface temperature and pulp chamber temperature) at different distances (on tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2)) with different light-curing units, Bluephase style® battery
| Distance (mm) | Irradiance (mW/mm2) | Time (s) | Radiant exposure (J/mm2) | Surface temperature (°C) | Pulp chamber temperature (°C) |
|---|---|---|---|---|---|
| 0 mm | 1222 ± 9 | 30 | 36.7 ± 0.3 | 52.6 ± 0.5 | 42.5 ± 0.4 |
| 20 | 24.4 ± 0.2 | 52.5 ± 0.4 | 41.5 ± 0.3 | ||
| 10 | 12.2 ± 0.1 | 48.6 ± 0.7 | 40.0 ± 0.1 | ||
| 2 mm | 1537 ± 13 | 30 | 46.2 ± 0.4 | 58.1 ± 0.9 | 42.0 ± 0.6 |
| 20 | 30.7 ± 0.3 | 51.3 ± 0.6 | 41.3 ± 0.4 | ||
| 10 | 15.4 ± 0.1 | 47.8 ± 0.8 | 40.0 ± 0.3 | ||
| 4 mm | 1366 ± 1 | 30 | 40.1 ± 0.0 | 53.1 ± 0.8 | 39.5 ± 0.3 |
| 20 | 27.3 ± 0.0 | 50.2 ± 0.4 | 39.3 ± 0.4 | ||
| 10 | 13.7 ± 0.0 | 46.3 ± 1.3 | 37.7 ± 0.2 |
Influence of curing time (time), distance, irradiance, and radiant exposure (product of irradiance and time) on temperature of the surface and pulp chamber on tooth with class I cavity (T1) and tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2) according to multiple linear regression analyses
| T1 | T2 | ||||
|---|---|---|---|---|---|
| Surface temperature | Pulp chamber temperature | Surface temperature | Pulp chamber temperature | ||
| Independent variable | B (95 % CI) | ||||
| Model 1 | Time | ||||
| 20 s (vs. 10 s) | 4.2 (3.5–4.9) | 0.3 (0.2–0.4) | 2.9 (1.8–4.0) | 1.6 (1.4–1.9) | |
| 30 s (vs. 10 s) | 7.2 (6.5–7.8) | 0.9 (0.8–1.0) | 5.2 (4.0–6.3) | 2.6 (2.3–2.8) | |
| Irradiance (per 100 units) | 0.8 (0.7–0.9) | 0.001 (0–0.001) | 1.3 (1.1–1.4) | 0.4 (0.3–0.4) | |
| Distance | |||||
| 2 mm (vs. 0 mm) | −1.5 (−2.2–(−0.9)) | −0.05 (−0.1–0.4) | 0.4 (−0.8–1.5) | −0.6 (−0.7–(−0.3)) | |
| 4 mm (vs. 0 mm) | −3.9 (−4.6–(−3.3)) | −0.3 (−0.4–(−0.2)) | −0.9 (−2.0–0.3) | −1.8 (−2.0–(−1.6)) | |
| Explained variance | 0.86 | 0.84 | 0.62 | 0.85 | |
| Model 2 | Radiant exposure (per 10 units) | 2.9 (2.7–3.2) | 0.3 (0.3–0.4) | 3.0 (2.5–3.4) | 1.2 (1.1–1.3) |
| Distance | |||||
| 2 mm (vs. 0 mm) | −1.4 (−2.1–(−0.7)) | −0.03 (−0.1–0.05) | 0.9 (−0.4–2.3) | −0.4 (−0.7–(−0.2)) | |
| 4 mm (vs. 0 mm) | −3.7 (−4.4–(−3.0)) | −0.3 (−0.4–(−0.2)) | −0.5 (−1.8–0.8) | −1.7 (−1.9–(−1.5)) | |
| Explained variance | 0.84 | 0.84 | 0.49 | 0.81 | |
Irradiance, radiant exposure (product of irradiance and time (s)), and heat development (surface temperature and pulp chamber temperature) at different distances (on tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2)) with different light-curing units, Bluephase style® plugged
| Distance (mm) | Irradiance (mW/mm2) | Time (s) | Radiant exposure (J/mm2) | Surface temperature (°C) | Pulp chamber temperature (°C) |
|---|---|---|---|---|---|
| 0 | 1257 ± 4 | 30 | 37.8 ± 0.1 | 44.4 ± 0.8 | 40.8 ± 0.2 |
| 20 | 25.1 ± 0.1 | 42.5 ± 0.4 | 39.2 ± 0.1 | ||
| 10 | 12.6 ± 0.1 | 40.1 ± 0.3 | 38.4 ± 0.1 | ||
| 2 | 1250 ± 9 | 30 | 37.5 ± 0.3 | 44.9 ± 0.4 | 40.5 ± 0.4 |
| 20 | 25.0 ± 0.2 | 42.9 ± 0.3 | 39.6 ± 0.2 | ||
| 10 | 12.5 ± 0.1 | 40.3 ± 0.2 | 37.9 ± 0.2 | ||
| 4 | 1176 ± 6 | 30 | 35.3 ± 0.2 | 42.8 ± 0.3 | 39.4 ± 0.3 |
| 20 | 23.5 ± 0.1 | 41.1 ± 0.4 | 38.3 ± 0.2 | ||
| 10 | 11.8 ± 0.1 | 38.9 ± 0.4 | 37.2 ± 0.1 |
Irradiance, radiant exposure (product of irradiance and time (s)), and heat development (surface temperature and pulp chamber temperature) at different distances (on tooth with a flat dentin surface with a pulpal wall thickness of approximately 0.6 mm (T2)) with different light-curing units, Bluephase G2® low mode
| Distance (mm) | Irradiance (mW/mm2) | Time (s) | Radiant exposure (J/mm2) | Surface temperature (°C) | Pulp chamber temperature (°C) |
|---|---|---|---|---|---|
| 0 | 774 ± 7 | 30 | 23.2 ± 0.2 | 41.5 ± 0.3 | 39.1 ± 0.2 |
| 20 | 15.5 ± 0.1 | 40.3 ± 0.3 | 38.5 ± 0.2 | ||
| 10 | 7.7 ± 0.1 | 38.1 ± 0.2 | 37.2 ± 0.1 | ||
| 2 | 827 ± 6 | 30 | 24.8 ± 0.2 | 41.1 ± 0.7 | 39.2 ± 0.3 |
| 20 | 16.5 ± 0.1 | 40.5 ± 0.2 | 38.2 ± 0.2 | ||
| 10 | 8.3 ± 0.1 | 38.2 ± 0.3 | 37.1 ± 0.2 | ||
| 4 | 864 ± 6 | 30 | 25.9 ± 0.2 | 42.2 ± 0.3 | 38.5 ± 0.3 |
| 20 | 17.3 ± 0.1 | 40.8 ± 0.3 | 37.6 ± 0.2 | ||
| 10 | 8.6 ± 0.1 | 38.7 ± 0.3 | 36.9 ± 0.1 |