Effect of distance on irradiance and beam homogeneity from 4 light-emitting diode curing units.

Richard B Price1, Daniel Labrie, J Marc Whalen, Christopher M Felix.   

Abstract

PURPOSE: To quantify the effect of distance on the irradiance and beam homogeneity from 4 curing lights.
METHODS: Four light-emitting diode curing lights were evaluated: Fusion, Bluephase 16i, Demi and FlashLite Magna. The irradiance at the centre of the light beam (ICB) was measured at 1.0 to 9.0 mm from the emitting tip using a 3.9-mm diameter probe connected to a spectrometer. The uniformity of the beam from each curing light was characterized by means of the "top hat factor" at 2.0, 4.0, 6.0 and 8.0 mm from the emitting tip. The useful beam diameter, within which irradiance values were greater than 400 mW/cm2, was calculated. The ICB, top hat factor and useful beam diameter were compared by analysis of variance and Fisher's protected least significant difference test at α = 0.01.
RESULTS: At all distances, the ICB was lowest for the FlashLite Magna and highest for the Fusion. Only the Fusion maintained an ICB above 1000 mW/cm2 at the 8.0 mm distance. For distances between 2.0 and 8.0 mm, the top hat factors were similar for the Fusion and the Demi, lower for the Bluephase 16i and lowest for the FlashLite Magna.
CONCLUSIONS: Beam homogeneity, top hat factors and ICB varied significantly among the curing lights. These results indicate that deep restorations may not be adequately cured if the curing time is based on data obtained when the curing light is positioned close to the radiometer or resin. In addition, a single irradiance value cannot be used to describe the light output from a curing light.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21507291

Source DB:  PubMed          Journal:  J Can Dent Assoc        ISSN: 0709-8936            Impact factor:   1.316


  15 in total

1.  Polymerization pattern characterization within a resin-based composite cured using different curing units at two distances.

Authors:  Afnan O Al-Zain; George J Eckert; Henry Lukic; Spiro Megremis; Jeffrey A Platt
Journal:  Clin Oral Investig       Date:  2019-02-09       Impact factor: 3.573

Review 2.  Polymerization shrinkage assessment of dental resin composites: a literature review.

Authors:  Dalia Kaisarly; Moataz El Gezawi
Journal:  Odontology       Date:  2016-08-19       Impact factor: 2.634

3.  Effects of occlusal cavity configuration on 3D shrinkage vectors in a flowable composite.

Authors:  Dalia Kaisarly; Moataz El Gezawi; Guangyun Lai; Jian Jin; Peter Rösch; Karl-Heinz Kunzelmann
Journal:  Clin Oral Investig       Date:  2017-12-16       Impact factor: 3.573

4.  Influence of irradiation distance on the mechanical performances of resin composites polymerized with high-irradiance light curing units.

Authors:  Soram Oh; Hyun Ju Kim; Hyun-Jung Kim; Sibel A Antonson; Sun-Young Kim
Journal:  Biomater Res       Date:  2022-05-20

Review 5.  Polymer-Based Direct Filling Materials.

Authors:  Carmem S Pfeifer
Journal:  Dent Clin North Am       Date:  2017-10

Review 6.  Factors affecting polymerization of resin-based composites: A literature review.

Authors:  Maan M AlShaafi
Journal:  Saudi Dent J       Date:  2017-03-07

7.  Application of a novel, continuous-feeding ultraviolet light emitting diode (UV-LED) system to disinfect domestic wastewater for discharge or agricultural reuse.

Authors:  Thi Minh Hong Nguyen; Poonyanooch Suwan; Thammarat Koottatep; Sara E Beck
Journal:  Water Res       Date:  2019-01-16       Impact factor: 11.236

8.  Light-curing units used in dentistry: factors associated with heat development-potential risk for patients.

Authors:  Mathieu Mouhat; James Mercer; Lina Stangvaltaite; Ulf Örtengren
Journal:  Clin Oral Investig       Date:  2016-10-01       Impact factor: 3.573

9.  Infection Control Barrier and Curing Time as Factors Affecting the Irradiance of Light-Cure Units.

Authors:  Reem A Ajaj; Hani M Nassar; Fatin A Hasanain
Journal:  J Int Soc Prev Community Dent       Date:  2018-11-29

10.  Shrinkage vectors in flowable bulk-fill and conventional composites: bulk versus incremental application.

Authors:  Dalia Kaisarly; Moataz El Gezawi; Andreas Keßler; Peter Rösch; Karl-Heinz Kunzelmann
Journal:  Clin Oral Investig       Date:  2020-07-11       Impact factor: 3.573

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.