| Literature DB >> 27694097 |
Azizeh Khaled Sowan1, Charles Calhoun Reed, Nancy Staggers.
Abstract
BACKGROUND: Large datasets of the audit log of modern physiologic monitoring devices have rarely been used for predictive modeling, capturing unsafe practices, or guiding initiatives on alarm systems safety.Entities:
Keywords: alarm fatigue; audit log; clinical alarms; intensive care unit; large clinical data; nursing; physiologic monitors
Year: 2016 PMID: 27694097 PMCID: PMC5065678 DOI: 10.2196/humanfactors.6427
Source DB: PubMed Journal: JMIR Hum Factors ISSN: 2292-9495
An example of a unit audit log.
| Date | Bed label | MRNa | Alarm and Action | Device nameb |
| 4/20/14 0:00:00 | 9115-S1 | 0000000 | **PAPdc 18 >16 Ended.d | PIIC iX: ixsurv006 |
| 4/20/14 0:00:00 | 9115-S1 | 0000000 | **PAPd 18 >16 Generated.d | PIIC iX: ixsurv006 |
| 4/20/14 0:00:00 | 9115-S1 | 0000000 | Yellow alarm sound played.e | PIIC iX: ixsurv006 |
| 4/20/14 0:00:02 | 9115-S1 | 0000000 | **ABPsf 170 >160 Generated.d | PIIC iX: ixsurv006 |
| 4/20/14 0:01:05 | 9123-S1 | 0000000 | ***Desatg 70 < 78 Generated.h | PIIC iX: ixsurv005 |
| 4/20/14 0:01:05 | 9123-S1 | 0000000 | Red alarm sound played.e | PIIC iX: ixsurv005 |
| 4/20/14 0:01:12 | 9123-S1 | 0000000 | ***Desat 73 < 78 ended.h | PIIC iX: ixsurv005 |
| 4/20/14 0:01:16 | 9115-S1 | 0000000 | **ABPs 168 >160 Ended.d | PIIC iX: ixsurv006 |
| 4/20/14 0:01:20 | 9123-S1 | 0000000 | Silence.i | PIIC iX: ixsurv005 |
| 4/20/14 0:01:40 | 9115-S1 | 0000000 | *Multiform PVCsj Generated.d | PIIC iX: ixsurv006 |
| 4/20/14 0:01:40 | 9115-S1 | 0000000 | Resume All Alarms.i | PIIC iX: ixsurv006 |
| 4/20/14 0:01:40 | 9117-S1 | 0000000 | **RRk 37 >30 Ended.d | PIIC iX: ixsurv006 |
| 4/20/14 0:01:40 | 9111-S1 | 0000000 | Patient transferred to 9035-S1.i | PIIC iX: ixsurv006 |
| 4/20/14 0:01:42 | 9095-S1 | 0000000 | Patient category set to Adult.i | PIIC iX: ixsurv006 |
| 4/20/14 0:01:43 | 9090-S1 | 0000000 | Pacer algorithm set to Pacer Algorithm On.i | PIIC iX: ixsurv006 |
| 4/20/14 0:01:44 | 9123-S1 | 0000000 | ECGl Leads Off Generated.m | PIIC iX: ixsurv006 |
| 4/20/14 0:01:44 | 9123-S1 | 0000000 | INOPn sound played.e | PIIC iX: ixsurv006 |
| 4/20/14 0:01:59 | 9123-S1 | 0000000 | ECG Leads Off Ended.m | PIIC iX: ixsurv006 |
| 4/20/14 0:02:00 | 9093-S1 | 0000000 | Equipment Offline.i | PIIC iX: ixsurv006 |
| 4/20/14 0:02:00 | 9115-S1 | 0000000 | **PAPd 18 >16 Generated.d | PIIC iX: ixsurv006 |
| 4/20/14 0:02:00 | 9093-S1 | 0000000 | Equipment Online.i | PIIC iX: ixsurv006 |
| 4/20/14 0:02:00 | 9085-S1 | 0000000 | ST: Al. Limits ST-V2o High: 1.6 ST-V2 Low: −1.6.p | PIIC iX: ixsurv006 |
| 4/20/14 0:02:00 | 9117-S1 | 0000000 | Arrhythmia Off.q | PIIC iX: ixsurv006 |
| 4/20/14 0:02:02 | 9115-S1 | 0000000 | Pause All Alarms.i | PIIC iX: ixsurv006 |
| 4/20/14 0:02:02 | 9075-S1 | 0000000 | Arrhy: Missed Beat Off.r | PIIC iX: ixsurv006 |
| 4/20/14 0:03:00 | 9085-S1 | 0000000 | SpO2s: Desat Limit 78.p | PIIC iX: ixsurv006 |
aMRN is the medical record number and was presented as zeros for confidentiality purposes.
bDevice name refers to the Information Center host name (eg, PIIC iX: ixsurv006).
cPAPd: pulmonary artery pressure diastolic.
dThese are examples of yellow physiologic alarms.
eThese messages appear if “Alert Sound” was selected as a search criterion from the Alerts category.
fABPs: arterial blood pressure systolic.
gDesat: desaturation.
hAn example of a red physiologic alarm.
iThese are examples of clinicians’ actions. They depend on the “actions” selected from the search boxes.
jPVC: premature ventricular contraction.
kRR: respiratory rate.
lECG: electrocardiographic.
mAn example of INOP alarm.
nINOP: inoperative.
oST-V2: a segment in the electrocardiogram.
pExamples of User Action–Alarm Limit Change.
qAn example of User Action–Measurement Off.
rAn example of User Action–Alarm Off; Arrhy: arrhythmia.
sSpO2: peripheral capillary oxygen saturation.
An example of variations in setting the lower limit of the Desat (desaturation) parameter.
| Date | Bed label | Alarm and Action | Device namea |
| 7/16/14 4:57 | 9101-S1 | *** Desat 89 < 90 Generated.b | PIIC iX: ixsurv007 |
| 7/28/14 21:44 | 9115-S1 | *** Desat 87 < 88 Generated. | PIIC iX: ixsurv006 |
| 7/30/14 1:24 | 9109-S1 | *** Desat 8 < 80 Generated. | PIIC iX: ixsurv006 |
| 8/11/14 11:59 | 9097-S1 | *** Desat 44 < 50 Generated. | PIIC iX: ixsurv007 |
| 8/12/14 10:43 | 9113-S1 | *** Desat 80 < 83 Generated. | PIIC iX: ixsurv006 |
| 9/5/14 21:38 | 9123-S1 | *** Desat 0 < 78 Generated. | PIIC iX: ixsurv005 |
aDevice name refers to the Information Center host name (eg, PIIC iX: ixsurv006).
bThe three starts (***) indicate that Desat (desaturation) is a red or high priority alarm.