| Literature DB >> 27036170 |
Azizeh Khaled Sowan1, Tiffany Michelle Gomez, Albert Fajardo Tarriela, Charles Calhoun Reed, Bruce Michael Paper.
Abstract
BACKGROUND: Clinical alarm systems safety is a national concern, specifically in intensive care units (ICUs) where alarm rates are known to be the highest. Interventional projects that examined the effect of changing default alarm settings on overall alarm rate and on clinicians' attitudes and practices toward clinical alarms and alarm fatigue are scarce.Entities:
Keywords: alarm fatigue; cardiac monitors; default alarm settings; in-service; intensive care unit; nursing; survey
Year: 2016 PMID: 27036170 PMCID: PMC4797663 DOI: 10.2196/humanfactors.5098
Source DB: PubMed Journal: JMIR Hum Factors ISSN: 2292-9495
Changes in default settings of the cardiac monitors at the transplant/cardiac intensive care unit.
| Type of change | Parameter | Default setting | Changed to... |
|
|
|
|
|
|
| PVCsa/minute | 10 bpm | 6 bpm |
| TachyClampb | 200 bpm | 180 bpm | |
| Limit increase | ExtremeTachyb | 20 bpm > Heart Rate High Limit | 40 bpm > Heart Rate High Limit |
|
|
|
|
|
|
| ABPc | Source: Systolic | Source: Systolic and Mean |
| NBPd | Source: Systolic and Mean | Source: Systolic | |
|
|
|
|
|
|
| System Pulsee | SpO2f | Auto (from ABP, SpO2, etc) |
| Asystole | Standard | Enhanced | |
| Alarm delay | SpO2: Averageg | No | Yes: 10 seconds |
|
|
|
|
|
|
| STh Analysisi | On | Off |
|
| Run PVCs | On | Off |
|
| Pair PVCs | On | Off |
|
| Bigeminy PVCs | On | Off |
|
| Trigeminy PVCs | On | Off |
|
| Multiform PVCs | On | Off |
|
| Pause | On | Off |
|
| Missed Beat | On | Off |
|
| AFIBj | On | Off |
|
| NBP Done Tone | On | Off |
| Decrease alarm volume | Yellow Alarm Volume | 5 | 3 |
| Increase alarm volume | Red Alarm Volume | +0 | +2 |
aPVC: premature ventricular contraction.
bThese alarms are not amenable to change by clinicians. All other alarms can be customized by clinicians based on the patient condition.
cABP: arterial blood pressure.
dNBP: noninvasive blood pressure.
eIf the peripheral capillary oxygen saturation (SpO2) had a poor waveform, the pulse from the pleth would not pick up and would therefore alarm. Changing to Auto allows the monitor to detect a pulse from other sources before alarming.
fSpO2: peripheral capillary oxygen saturation.
gSpO2 will be averaged over 10 seconds to determine a value instead of alarming the second the SpO2 drops below the limit. The nurse can also manually increase this to 20 or 30 seconds.
hST: ST segment in the electrocardiogram.
iThe ST Analysis alarm was disabled but should be turned on for all interventional cardiology cases (eg, require cardiac catheterization) or acute coronary syndrome patients. For these specific patients, the original limit of +/-2.0 mm should be tightened to +/- 1.6 mm as per physicians’ requests.
jAFIB: atrial fibrillation.
Difference in alarm rates between the preproject and postproject periods.
| Alarm condition | Preproject period | Postproject period | |||
|
| Number of alarms | Total alarm rate/patient day | Number of alarms | Total alarm rate/patient day | |
|
|
|
|
|
| |
|
| Total | 27,930 | 38.05 | 28,049 | 33.67 |
|
| ABPsb (systolic) | 13,776 |
| 14,726 |
|
|
| ABPmc (mean) | 13,548 |
| 12,895 |
|
|
| ABP disconnect | 606 |
| 428 |
|
| Pair PVCsd,e | 8305 | 11.31 | 164 | 0.19 | |
|
|
|
|
|
| |
|
| Total | 7079 | 9.64 | 8290 | 9.95 |
|
| SpO2 | 6741 |
| 7858 |
|
|
| SpO2rg (right) | 338 |
| 323 |
|
|
| SpO2lh (left) | 0 |
| 109 |
|
| Multiform PVCse | 5865 | 7.99 | 19 | 0.02 | |
|
|
|
|
|
| |
|
| Total | 3686 | 5.02 | 3976 | 4.77 |
|
| NBPmj (mean) | 1847 |
| 43 |
|
|
| NBPsk (systolic) | 1837 |
| 3933 |
|
|
| NBPdl (diastolic) | 2 |
| 0 |
|
| PVCs/min | 3233 | 4.40 | 5330 | 6.39 | |
| Run PVCs highe | 2155 | 2.94 | 23 | 0.03 | |
| STe | 1851 | 2.52 | 2609 | 3.13 | |
|
|
|
|
|
| |
|
| Total | 1481 | 2.02 | 32 | 0.04 |
|
| AFIB | 990 |
| 26 |
|
|
| End AFIBn | 491 |
| 6 |
|
| Pausee | 1086 | 1.48 | 8 | 0.01 | |
| Missed Beate | 873 | 1.19 | 89 | 0.11 | |
| Asystole | 323 | 0.44 | 565 | 0.68 | |
|
|
|
|
|
| |
|
| Total | 292 | 0.39 | 153 | 0.18 |
|
| Tachy | 273 |
| 153 |
|
|
| Tachy/pp (tachycardia p wave) | 19 |
| 0 |
|
| Ventq Bigeminye | 234 | 0.32 | 7 | 0 | |
| Vent Trigeminye | 79 | 0.11 | 0 | 0 | |
| Pulse | 28 | 0.04 | 5 | 0.01 | |
| Total | 64,500 | 87.86 | 49,319 | 59.18 | |
aABP: arterial blood pressure.
bABPs: arterial blood pressure systolic.
cABPm: arterial blood pressure mean.
dPVC: premature ventricular contraction.
eThese are the alarms that we disabled.
fSpO2: peripheral capillary oxygen saturation.
gSpO2r: peripheral capillary oxygen saturation right.
hSpO2l: peripheral capillary oxygen saturation left.
iNBP: noninvasive blood pressure.
jNBPm: noninvasive blood pressure mean.
kNBPs: noninvasive blood pressure systolic.
lNBPd: noninvasive blood pressure diastolic.
mAFIB: atrial fibrillation.
nEnd AFIB alarm indicates the end of the AFIB status.
oTachy: tachycardia.
pTachy/p: tachycardia p wave.
qVent: ventricular.
Number and percentage of nurses who agreed or strongly agreed on the statements between the preproject and postproject periods (n=24).
| Item | Statementa | Preproject, | Postproject, | % changeb |
| 1 | Nuisance alarms occur frequently | 24 (100) | 18 (75) | -25.0 |
| 2 | Nuisance alarms disrupt patient care | 23 (96) | 23 (96) | 0 |
| 3 | Nuisance alarms reduce trust in alarms and cause caregivers to inappropriately turn alarms off at times other than setup or procedural events | 21 (88) | 22 (92) | 4.8 |
| 4 | When a number of devices are used with a patient, it can be confusing to determine which device is in an alarm condition | 21 (88) | 19 (79) | -9.5 |
| 5 | Smart alarms (eg, where multiple parameters, rate of change of parameters, and signal quality are automatically assessed in their entirety) would be effective to use for improving clinical response to important patient alarms | 20 (83) | 17 (71) | -15.0 |
| 6 | Central alarm management staff responsible for receiving alarm messages and alerting appropriate staff is helpful | 19 (79) | 18 (75) | -5.3 |
| 7 | Smart alarms (eg, where multiple parameters, rate of change of parameters, and signal quality are automatically assessed in their entirety) would be effective to use for reducing false alarms | 19 (79) | 16 (67) | -15.8 |
| 8c | Unit layout does interfere with alarm recognition and management | 18 (75) | 18 (75) | 0 |
| 9 | Alarm integration and communication systems via pagers, cell phones, and other wireless devices are useful for improving alarms management and response | 15 (63) | 17 (71) | 13.3 |
| 10c | Nearly all alarms are actionable (requiring the nurse to respond and take an action) | 14 (58) | 14 (58) | 0 |
| 11 | Alarm sounds and/or visual displays of the current monitoring systems and devices should clearly differentiate the priority of alarm | 13 (54) | 14 (58) | 7.7 |
| 12 | Properly setting alarm parameters and alerts is overly complex in existing devices | 13 (54) | 13 (54) | 0 |
| 13 | Clinical staff is sensitive to alarms and responds quickly | 13 (54) | 15 (63) | 15.4 |
| 14c | When a lethal alarm sounds, it is clearly and quickly recognized and immediate action is taken to address the alarm | 12 (50) | 14 (58) | 16.7 |
| 15 | Environmental background noise has interfered with alarm recognition | 12 (50) | 15 (63) | 25.0 |
| 16 | Alarm sounds and/or visual displays should be distinct based on the parameter or source (eg, device) | 12 (50) | 16 (67) | 33.3 |
| 17d | There is a requirement in my unit to document that the alarms are set and are appropriate for each patient | 11 (46) | 18 (75) | 63.6 |
| 18d | The alarms used on my unit are adequate to alert staff of potential or actual changes in a patient’s condition | 10 (42) | 9 (38) | -10.0 |
| 19 | There have been frequent instances where alarms could not be heard and were missed | 8 (33) | 8 (33) | 0 |
| 20d | The medical devices used on my unit all have distinct outputs (ie, sounds, repetition rates, visual displays) that allow users to identify the source of the alarm | 8 (33) | 15 (63) | 87.5 |
| 21d | Clinical policies and procedures regarding alarm management are effectively used in my unit | 6 (25) | 11 (46) | 83.3 |
| 22 | Newer monitoring systems (eg, < 3 years old) have solved most of the previous problems we experienced with clinical alarms | 1 (4) | 6 (25) | 500 |
aEdited and used with permission from the Healthcare Technology Foundation (HTF) 2011.
bPercent change = ((y2 - y1) / y1) × 100.
cThese are the new statements that we added to our survey. They do not exist in the original HTF survey.
dThese are the statements where the “floor/area of the hospital” or “institution” in the HTF clinical alarms survey were replaced with “unit” in our survey.
Importance of alarm issues related to the cardiac monitors (n=24).
| Item | Statement | Preproject | Postproject |
| ||
|
|
| Item response, mean | Mean rankinga | Item response, mean | Mean ranking |
|
| 1b | Frequent false alarms, which lead to reduced attention or response to alarms when they occur | 2.40 | 1 | 3.40 | 1 | .11 |
| 2b | Difficulty in understanding the priority of an alarm | 3.00 | 2 | 4.32 | 2 | .07 |
| 3b | Noise competition from nonclinical alarms and pages | 3.95 | 3 | 4.55 | 3 | .50 |
| 4c | Lack of available policy on appropriate alarm parameters for individualized patients | 4.40 | 4 | 5.80 | 9 | .08 |
| 5c | The need to frequently reset alarm settings every time they revert back to default when the monitor is disconnected from the patient | 4.42 | 5 | 5.16 | 5 | .24 |
| 6d | Difficulty in hearing alarms when they occur, especially from outside patient room | 4.47 | 6 | 5.37 | 7 | .36 |
| 7d | Difficulty in setting alarms properly because of the complexity of the monitor | 4.84 | 7 | 5.21 | 6 | .70 |
| 8b | Lack of training on alarm systems | 4.90 | 8 | 4.70 | 4 | .83 |
| 9d | Difficulty in setting alarms properly because of lack of knowledge on the appropriate limits for my patient condition | 5.42 | 9 | 5.58 | 8 | .75 |
aItem response means were ranked from 1 (most important) to 9 (least important).
bThese statements were adopted from the Healthcare Technology Foundation (HTF) survey.
cThese statements were added to the survey to reflect the cardiac monitors.
dThese statements were modified from the HTF survey. Original statements were as follows: item 6 “Difficulty in hearing alarms when they occur”; items 7 and 9 “Difficulty in setting alarms properly.”
Figure 1Percentage of nurses who modify the bedside alarm parameters in the pre- and postproject periods (n=24).
Figure 2Percentage of nurses who replace patients' electrodes in the pre- and postproject periods (n=24).
Figure 3Percentage of nurses who received and needed monitor training in the pre- and postproject periods (n=24).