Literature DB >> 25554556

Differences in alarm events between disposable and reusable electrocardiography lead wires.

Nancy M Albert1, Terri Murray2, James F Bena2, Ellen Slifcak2, Joel D Roach2, Jackie Spence2, Alicia Burkle2.   

Abstract

BACKGROUND: Disposable electrocardiographic lead wires (ECG-LWs) may not be as durable as reusable ones.
OBJECTIVE: To examine differences in alarm events between disposable and reusable ECG-LWs.
METHOD: Two cardiac telemetry units were randomized to reusable ECG-LWs, and 2 units alternated between disposable and reusable ECG-LWs for 4 months. A remote monitoring team, blinded to ECG-LW type, assessed frequency and type of alarm events by using total counts and rates per 100 patient days. Event rates were compared by using generalized linear mixed-effect models for differences and noninferiority between wire types.
RESULTS: In 1611 patients and 9385.5 patient days of ECG monitoring, patient characteristics were similar between groups. Rates of alarms for no telemetry, leads fail, or leads off were lower in disposable ECG-LWs (adjusted relative risk [95% CI], 0.71 [0.53-0.96]; noninferiority P < .001; superiority P = .03) and monitoring (artifact) alarms were significantly noninferior (adjusted relative risk [95% CI]: 0.88, [0.62-1.24], P = .02; superiority P = .44). No between-group differences existed in false or true crisis alarms. Disposable ECG-LWs were noninferior to reusable ECG-LWs for all false-alarm events (N [rate per 100 patient days], disposable 2029 [79.1] vs reusable 6673 [97.9]; adjusted relative risk [95% CI]: 0.81 [0.63-1.06], P = .002; superiority P = .12.)
CONCLUSIONS: Disposable ECG-LWs with patented push-button design had superior performance in reducing alarms created by no telemetry, leads fail, or leads off and significant noninferiority in all false-alarm rates compared with reusable ECG-LWs. Fewer ECG alarms may save nurses time, decrease alarm fatigue, and improve patient safety. ©2015 American Association of Critical-Care Nurses.

Entities:  

Mesh:

Year:  2015        PMID: 25554556     DOI: 10.4037/ajcc2015663

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  6 in total

1.  The Economic Implications of Introducing Single-Patient ECG Systems for Cardiac Surgery in Australia.

Authors:  Rhodri Saunders; Amanda Hansson Hedblom
Journal:  Clinicoecon Outcomes Res       Date:  2021-08-13

Review 2.  Systematic Review of Physiologic Monitor Alarm Characteristics and Pragmatic Interventions to Reduce Alarm Frequency.

Authors:  Christine Weirich Paine; Veena V Goel; Elizabeth Ely; Christopher D Stave; Shannon Stemler; Miriam Zander; Christopher P Bonafide
Journal:  J Hosp Med       Date:  2015-12-14       Impact factor: 2.960

3.  Changes in Default Alarm Settings and Standard In-Service are Insufficient to Improve Alarm Fatigue in an Intensive Care Unit: A Pilot Project.

Authors:  Azizeh Khaled Sowan; Tiffany Michelle Gomez; Albert Fajardo Tarriela; Charles Calhoun Reed; Bruce Michael Paper
Journal:  JMIR Hum Factors       Date:  2016-01-11

4.  Role of Large Clinical Datasets From Physiologic Monitors in Improving the Safety of Clinical Alarm Systems and Methodological Considerations: A Case From Philips Monitors.

Authors:  Azizeh Khaled Sowan; Charles Calhoun Reed; Nancy Staggers
Journal:  JMIR Hum Factors       Date:  2016-09-30

5.  Faster clinical response to the onset of adverse events: A wearable metacognitive attention aid for nurse triage of clinical alarms.

Authors:  Daniel C McFarlane; Alexa K Doig; James A Agutter; Lara M Brewer; Noah D Syroid; Ranjeev Mittu
Journal:  PLoS One       Date:  2018-05-16       Impact factor: 3.240

6.  Addressing vital sign alarm fatigue using personalized alarm thresholds.

Authors:  Sarah Poole; Nigam Shah
Journal:  Pac Symp Biocomput       Date:  2018