| Literature DB >> 25338067 |
Barbara J Drew1, Patricia Harris1, Jessica K Zègre-Hemsey2, Tina Mammone3, Daniel Schindler1, Rebeca Salas-Boni1, Yong Bai1, Adelita Tinoco1, Quan Ding1, Xiao Hu1.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms.Entities:
Mesh:
Year: 2014 PMID: 25338067 PMCID: PMC4206416 DOI: 10.1371/journal.pone.0110274
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Physiologic monitor device in Intensive Care Unit.
Bedside patient monitor (GE Healthcare, Milwaukee, WI) displays multiple physiologic waveforms and vital sign measurements. The nurse pictured here gave written informed consent to publish this photograph supplied by the San Francisco Chronicle newspaper (with permission) for their story on alarm fatigue at: http://www.sfgate.com/health/article/Hospitals-look-to-reduce-danger-of-alarm-fatigue-4918018.php.
UCSF Alarm Study Units.
| Hospital Unit | # Beds | Patient Population |
| 2 Intensive Care Units | 32 | Critically-ill adults with complex medical disorders and post-operative general surgery, solid organ transplant, acute kidney injury, acute and end-stage liver failure, sepsis, multiple organ dysfunction syndrome, and liver, pancreas or small bowel transplantation. Patients commonly on mechanical ventilation. |
| Cardiac Critical Care | 16 | Critically-ill adults with cardiac disorders, including cardiac medicine, cardiothoracic surgery, transplant (heart, lung, heart & lung), thoracic or vascular surgery. Patients with left ventricular assist devices, pacemakers, & implantable cardioverter devices. |
| 2 Neuroscience Care Units | 29 | Critically-ill adults with neurological impairment (subarachnoid hemorrhage, stroke, brain tumors, traumatic brain injury) who undergo complex surgical and interventional procedures and patients going through the organ donation process. |
|
|
|
Figure 2Hospital infrastructure to automatically store all physiologic monitor waveform and alarm data.
Figure 3Patient monitoring ECG lead configuration.
A 5-electrode lead configuration was used in all study ICUs with Mason-Likar electrode placement of the limb leads on the torso and one chest electrode that is routinely placed in the V1 location.
Alarm Default Settings for Adult ICUs during the Study Period.
| Patient Status Arrhythmia Alarms | ||
| Alarm Sounds: Crisis, 3 beeps continually; Warning, 2 beeps; Advisory, 1 beep; Message, inaudible text | ||
| Alarm | Factory Default | UCSF Default |
| Asystole | Crisis | Crisis |
| Ventricular Fibrillation/Ventricular Tachycardia | Crisis | Crisis |
| Ventricular Tachycardia | Crisis | Crisis |
| Ventricular Tachycardia >2 | Crisis | Advisory |
| Ventricular Bradycardia | Crisis | Warning |
| Accelerated Ventricular Rhythm | Message | Warning |
| Pause | Message | Warning |
| Tachycardia | Message | Advisory |
| Bradycardia | Message | Advisory |
| R on T | Message | Message |
| Couplet | Message | Message |
| Bigeminy | Message | Message |
| Trigeminy | Message | Message |
| Premature ventricular contraction (PVC) | Message | Message |
| Irregular | Message | Message |
| Atrial Fibrillation | Message | Advisory |
|
| ||
|
| ||
|
|
|
|
| Heart Rate | 50/150 Warning | 50/130 Warning |
| PVC/minute | 6 Message | 10 Message |
| Invasive arterial pressure | Advisory | 90/160 Warning |
| Noninvasive blood pressure | Advisory | 90/160 Advisory |
| ST segment | Advisory | Advisory |
| Respiratory Rate | Message | Warning |
| No Breath/Apnea | Warning | Warning |
| Peripheral oxygen saturation (SpO2) | Advisory | 90% Advisory |
|
| ||
|
| ||
|
|
|
|
| Artifact | Message | Message |
| Lead Fail (single lead I or II or III or RL or V) | Message | Message |
| ECG Leads Fail | Warning | Warning |
| Respiratory Leads Fail | Warning | Warning |
| Arrhythmia Suspend | Warning | Warning |
| Invasive Pressure Sensor Fail | Warning | Warning |
| Noninvasive Blood Pressure Deflation Failure | Warning | Warning |
| Noninvasive Blood Pressure Exceed 3 Minutes | Warning | Warning |
| Noninvasive Blood Pressure Excessive Pressure 200 | Warning | Warning |
| Noninvasive Blood Pressure Invalid Command | Warning | Warning |
Alarm Annotation Protocol.
| Alarm Label & Algorithm Definition | Proof of True versus False Alarm by Investigator |
|
|
|
| 1. Simultaneous drop in invasive arterial or pulmonary artery (PA) pressure to near zero | |
| 2. Documentation from electronic medical record (EMR) of asystolic cardiac arrest at same time | |
|
| |
| 1. No simultaneous decrease in invasive arterial or PA pressure | |
| 2. A visible QRS is evident in at least one ECG lead (examine all 7 available leads) | |
| 3. Good quality SpO2 signal has pulsatile waveform that matches rate of underlying baseline rhythm | |
| 4. ASYSTOLE alarm duration is >60 seconds but there is no EMR documentation that it was recognized clinically (syncope, seizure, loss of consciousness, cardiac arrest) | |
|
|
|
| 1. Simultaneous drop in invasive arterial or PA pressure to near zero | |
| 2. Documentation from EMR of ventricular tachycardia or fibrillation cardiac arrest at same time | |
|
| |
| 1. No simultaneous decrease in invasive arterial or PA pressure | |
| 2. There are QRS complexes with the same R–R intervals as the patient’s baseline rhythm evident in any ECG lead throughout the alarm event | |
| 3. Good quality SpO2 signal has pulsatile waveform that matches rate of underlying baseline rhythm | |
| 4. VFIB/VTAC alarm duration is >60 seconds but there is no EMR documentation that it was recognized clinically (syncope, seizure, loss of consciousness, cardiac arrest) | |
|
|
|
| 1. Wide QRS beats are not preceded by a P wave with a consistent PR interval | |
| 2. Fusion beats are evident at the transition between ventricular rhythm and sinus rhythm | |
|
| |
| 1. Event is sinus rhythm with BBB (P waves prior to each wide beat with consistent PR interval) | |
| 2. Patient is known to have ventricular pacemaker; event QRS matches paced rhythm on standard “diagnostic” 12-lead ECG | |
|
|
|
| 1. Simultaneous drop in invasive arterial or PA pressure | |
| 2. Documentation from EMR of VT at same time; standard 12-lead ECG documentation of VT read by cardiologist | |
| 3. Atrioventricular (AV) dissociation is evident throughout the wide QRS tachycardia in any ECG lead | |
| 4. Event wide QRS morphology is different than patient’s baseline rhythm with BBB | |
|
| |
| 1. No simultaneous change in invasive arterial or PA pressure (if it is “slow” VT with rate 100–150, there will be less decrease in pressure waveform amplitude) | |
| 2. There are QRS complexes with the same R–R intervals as the patient’s baseline rhythm evident in any ECG lead throughout the alarm event | |
| 3. Good quality SpO2 signal has pulsatile waveform that matches rate of underlying baseline rhythm | |
| 4. VTACH alarm duration is >60 seconds but there is no EMR documentation that it was recognized clinically (syncope, seizure, loss of consciousness, cardiac arrest) | |
| 5. Event has the same wide QRS complex morphology in all 7 ECG leads as the patient’s baseline rhythm with right or left BBB; additional confirmation if sinus P waves are evident prior to each QRS or the rhythm has no discernable P waves but is randomly irregular indicating atrial fibrillation | |
| 6. Event is due to intermittent ventricular pacing (visible pacer spikes before each wide QRS or QRS in all 7 leads matches a standard “diagnostic” 12-lead ECG acquired during ventricular pacing) | |
|
|
|
| 1. Simultaneous pause on invasive arterial or PA waveform | |
| 2. Simultaneous pause on good quality SpO2 waveform | |
|
| |
| 1. No simultaneous pause in invasive arterial or PA pressure | |
| 2. No simultaneous pause on good quality SpO2 waveform | |
| 3. There is a visible QRS during the pause (may be low amplitude) in any of the 7 available leads | |
|
|
|
| 1. Rhythm is complete heart block with ventricular escape rhythm | |
| 2. Rhythm is sinus node arrest with ventricular escape rhythm | |
|
| |
| 1. Event is sinus bradycardia with BBB (P waves prior to each beat with consistent PR interval) | |
| 2. Patient is known to have pacemaker; event QRS in all 7 leads matches paced rhythm on “diagnostic” 12-lead ECG |
Figure 4True positive ventricular tachycardia alarm using seven available ECG leads for diagnosis.
Page one of the alarm annotation analysis tool shows a 10-second rhythm strip of all seven available ECG leads at the time that a ventricular tachycardia alarm was triggered. In this and subsequent Figures, ECG Leads are displayed from top to bottom in the following sequence: Lead I, II, III, V (typically V1), aVR, aVL, aVF. As evident at the beginning of the rhythm strip, the patient has an underlying rhythm of atrial fibrillation with a rapid ventricular rate of about 140. There is an isolated ventricular premature beat (4th beat from the end) and its QRS morphology is identical to the initial beat of the alarm event. Knowing that the event is initiated by a ventricular ectopic beat provides strong evidence that this event is a true ventricular tachycardia alarm.
Figure 5True positive ventricular tachycardia alarm using non-ECG waveforms for diagnosis.
Page 2 of the alarm annotation analysis tool depicts the same alarm event as in Figure 4 with all available non-ECG waveforms. Additional proof that this is a true ventricular tachycardia alarm is provided by observing cessation of the arterial blood pressure waveform that falls to near zero during the arrhythmia.
Figure 6False positive ventricular tachycardia alarm using seven available ECG leads for diagnosis.
Page one of the alarm annotation analysis tool in a second patient with a ventricular tachycardia alarm. Proof that this is a false positive alarm is provided by observing Lead III that shows clearly-visible P-QRS-T waveforms indicating normal sinus rhythm. All six remaining ECG leads show artifact that mimics ventricular tachycardia. It is important to point out that Lead III is not one of the two leads routinely displayed on the bedside monitor in our ICUs so unless all available leads are reviewed, a misdiagnosis would be made of rapid polymorphic ventricular tachycardia. This type of rapid, repetitive artifact on the ECG is often created during patient monitoring by motion artifact during activities of daily living. The non-artifact lead (Lead III) uses the left arm and left leg electrodes but not the right arm electrode. So, this is likely to be a right-handed patient doing something like brushing teeth.
Figure 7False positive ventricular tachycardia alarm using non-ECG waveforms for diagnosis.
Page 2 of the alarm annotation analysis tool depicts the same alarm event as in Figure 6 showing all available non-ECG waveforms. Additional proof that this is a false ventricular tachycardia alarm is provided by the following: a.) no change in the arterial pressure waveform during the event, b.) arterial waveform pulsations match the normal sinus rhythm rate, and c.) SpO2 waveform pulsations match the normal sinus rhythm rate. Of interest, the same artifact that contaminates the ECG signal also contaminates the respiratory waveform, as evidenced by an erroneous device-measured respiratory rate of 162 breaths per minute.
Schema for Counting and Reporting Physiologic Monitor Device Alarms.
| Arrhythmia Alarms | |||
| Alarm Condition | Short Label | Definition | |
| 1. Accelerated Ventricular Rhythm | Acc Vent | ≥6 ventricular beats with heart rate between 50-100 | |
| 2. Atrial Fibrillation | Afib | Irregular timing of QRS complexes and absenceof preceding P waves | |
| 3. Asystole | Asystole | Heart rate drops to zero; typically no QRS for 5–6 seconds | |
| 4. Pause | Pause | No QRS for a 3-second interval | |
| 5. Ventricular Bradycardia | V Brady | ≥3 consecutive ventricular beats at an average rate ≤50 | |
| 6. Ventricular Fibrillation | Vfib/Vtac | Course flutter waves without QRS complexes | |
| 7. Ventricular Tachycardia | Vtach | ≥6 consecutive ventricular beats at rate ≥100 | |
| 8. Premature VentricularContractions | All PVC | VT>2 | 3-5 consecutive ventricular beats at rate ≥100 |
| PVC | Isolated PVCs | ||
| R on T | PVC falls on the ST or T wave portionof previous beat | ||
| Couplet | Two consecutive PVCs with rate >100 | ||
| Bigeminy | PVC alternates with a non-ventricular beat for ≥3 cycles | ||
| Trigeminy | PVC alternates with 2 non-ventricular beats for ≥3 cycles | ||
| PVC = > X | PVC count is equal to or>than user-defined limit | ||
|
| |||
| 9. Heart Rate | All HR | HR<or>user-defined limit determined from ECG waveform (HR) | |
| HR<or>user-defined limit determined from SpO2 waveform (SpO2 Rate) | |||
| HR<or>user-defined limit determined from arterial pressurewaveform (ART Rate or FEM Rate) | |||
| Bradycardia arrhythmia alarm: 8 R–R intervals fall below user-definedlow HR limit setting | |||
| Tachycardia arrhythmia alarm: 8 R–R intervals occur aboveuser-defined high HR limit setting | |||
| 10. Respiratory Rate | All RR | Respiratory rate<or>user-defined limit | |
| No breaths detected for user-defined period of seconds (Apnea Alarm) | |||
| 11. Oxygen Saturation | All SpO2 | SpO2< or>user-defined limit determined from pulse oximetry sensor | |
| 12. Invasive Arterial Pressure | All ART | Systolic = or<or>user-defined limit | |
| Diastolic = or<or>user-defined limit | |||
| Mean = or<or>user-defined limit | |||
| 13. Noninvasive Blood Pressure | All NIBP | Systolic = or<or>user-defined limit | |
| Diastolic = or<or>user-defined limit | |||
| Mean = or<or>user-defined limit | |||
| 14. Central & Intra-cardiac Pressure with invasive hemodynamic monitoring | All Heart Pressures | Systolic = or<or>user-defined limit for CVP, RAP, PAP, LAP | |
| Diastolic = or<or>user-defined limit for CVP, RAP, PAP, LAP | |||
| Mean = or<or>user-defined limit for CVP, RAP, PAP, LAP | |||
| 15. Intra-cranial Pressure | All ICP | ICP mean = or<or>user-defined limit | |
| 16. ST-segment Amplitude | All ST | Lead I or II or III or aVR or aVL or aVF or V1 or V2 or V3 or V4 or V5 or V6 ST<or>than PR segment amplitude by user-definedsetting (hospital default, ±2 millimeters) | |
|
| |||
| 17. Problem with artifact, sensors,probes, line disconnects, etc. | All Technical | Artifact: noisy signal on ECG | |
| Arrhythmia suspend: no arrhythmia detection due to sustained artifact | |||
| Arrhythmia Off | |||
| ECG Leads Fail or No ECG or individual Lead Fail (I, II, III, RL, V) | |||
| Respiratory Lead Fail | |||
| Sensor Fail for ART or FEM or ICP or CVP or RAP or PAP or LAP | |||
| SpO2 :Probe Off or Probe Fail or Low Signal or IncompatibleCable or Connect Probe | |||
| Noninvasive BP: invalid command or ExcessivePressure 200 or Exceeded 3 min or Deflation Failure orInflation Time Exceeded | |||
| Line Disconnect for PA or ART or FEM or CVP | |||
CVP = central venous pressure; RAP = right atrial pressure; PAP = pulmonary artery pressure; LAP = left atrial pressure; ART = invasive arterial line; FEM = invasive arterial line in femoral site.
Figure 8Frequency of all unique alarms (N = 2,558,760) over a 31-day period.
Figure 9False apnea alarm in a patient breathing adequately on mechanical ventilation.
The respiratory waveform (bottom tracing labelled “Resp”) has a flat line appearance. The detection of respirations from the ECG lead (impedance method) is inaccurate in this patient, displaying an erroneous respiratory rate of 4 per minute.
ST-Segment Alarm Durations in a 16-Bed Cardiac ICU.
| Alarm Duration (Seconds) | Number of Alarm Events | Percentage |
| 0<30 | 4,981 | 80% |
| 30<60 | 673 | 11% |
| >60 | 542 | 9% |
| Total: | 6,196 |
Accuracy of 12,671 Arrhythmia Alarms.
| Alarm Type | Number ofAlarms | Number ofPatients | Number of TruePositives | Number of FalsePositives | False PositiveRate |
| 1. Asystole | 792 | 113 | 260 | 531 | 67.0% |
| 2. Ventricular Fibrillation | 158 | 19 | 107 | 51 | 32.3% |
| 3. Ventricular Tachycardia | 3861 | 183 | 502 | 3352 | 86.8% |
| 4. Accelerated Ventricular Rhythm | 4361 | 99 | 224 | 4135 | 94.8% |
| 5. Pause | 2239 | 140 | 272 | 1963 | 87.7% |
| 6. Ventricular Bradycardia | 1260 | 39 | 40 | 1219 | 96.7% |
|
| 12671 | 1405 | 11251 | 88.8% |
*15 alarms were indistinguishable: 1 Asystole, 7 VTach, 2 AccVent, 4 Pause, and 1 Ventricular Brady.
Figure 10False alarm with one non-artifact ECG lead that confirms artifact mimicking ventricular fibrillation.
Six of the seven ECG leads show what looks like a rapid (>400) polymorphic ventricular arrhythmia. However, Lead II clearly shows sinus rhythm at a rate of 94. Without this single non-artifact lead, a misdiagnosis would be made of ventricular fibrillation.
Figure 11False accelerated ventricular rhythm alarm in a patient with left bundle branch block.
Sinus rhythm at a rate in the 60′s is evident by observing P waves preceding each QRS complex with a consistent PR interval. P waves are visible in all seven leads (especially clear-cut in Leads I and II).
Figure 12True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition.
Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].
Figure 13False accelerated ventricular rhythm alarm in a patient with ventricular pacing.
Patient with atrial fibrillation and intermittent ventricular pacing does not have PaceMode activated. As a result, a period of ventricular pacing goes undetected by the algorithm (no pacemaker spikes are “painted” in) and a false alarm is generated of accelerated ventricular rhythm. The investigators determined this to be intermittent pacing (rather than accelerated ventricular rhythm) because the rate matched the pacemaker heart rate setting. Moreover, the QRS morphology across all 7 leads matched the QRS morphology of corresponding leads on a hospital-acquired standard “diagnostic” 12-lead ECG during a known period of pacing.
Figure 14Low amplitude QRS in a patient with an excessive number of alarms.
Standard “diagnostic” 12-lead ECG recorded from the patient who contributed nearly half of the 12,671 arrhythmia alarms for annotation. The ECG shows left bundle branch block with low amplitude QRS complexes in the limb leads but not in the V leads. Since one of the available leads acquired with the physiologic patient monitoring device is a V lead, the arrhythmia algorithm could have avoided the excessive number of false alarms had all available leads been used for QRS detection.
Frequency of Visible QRS Complexes in One or More ECG Leads during False Brady-Arrhythmia Alarms.
| Alarm Type | Visible QRS Complex | |
| Yes | No | |
| False Asystole Alarms, N = 518 | 469 (91%) | 49 (9%) |
| False Pause Alarms, N = 1903 | 1786 (94%) | 117 (6%) |
Figure 15ECG signal quality in 12,671 annotated arrhythmia alarms.
Good signal quality (green) was defined as a clearly visible P-QRS-T waveform across all available leads with little to no noise, baseline wander, or leads off. Fair signal quality (yellow) was defined as moderate noise or baseline wander but having identifiable QRS complexes for basic rhythm/rate detection. Poor signal quality (red) was defined as being unanalyzable because of excessive noise, baseline wander or leads off.
Figure 16Electrode failure causing artifact and a false ventricular tachycardia alarm.
Electrocardiogram in 6 of the 7 available leads shows intermittent loss of signal (signal “squares off” on top and bottom of tracing) due to an electrode problem such as loss of skin contact or dried out electrode gel. One ECG lead (Lead II that uses the right arm and left leg electrodes) does not show electrode failure so the likely electrode that is malfunctioning is the left arm electrode. Failure to apply fresh electrodes in this case will result in numerous false alarms.
Comparison of Annotated Arrhythmia Alarm ICU Databases.
| Database Characteristics | MIMIC II | UCSF |
| Monitor manufacturer | Philips Healthcare | GE Healthcare |
| Number, type of patients | N = 447; had to have both ECG & invasive arterial pressure | N = 461; all consecutive patients |
| Number, type of beds | 48 beds; medical, surgical, cardiac | 77 beds; medical, surgical, cardiac, neurologic |
| Total monitoring hours | 41,301 hours | 48,173 hours |
| Number, type of ECG leads | 1–3 leads; inconsistent, including modified (MCL) leads | 7 leads; consistently I, II, III, aVR, aVL, aVF, V1 |
| Sampling rate | 125 HZ | ECG, 240 HZ; Pressures, 120 HZ; SpO2, 60 HZ |
| Resolution | 8 bit | 12 bit |
| Number of alarms annotated | 5,386 alarms; 5 alarm categories | 12,671 alarms; 6 alarm categories |
| Asystole | 579 | 792 |
| Vfib/Vtach | 313 | 158 |
| Vtach | 1900 | 3861 |
| Brady (<40), 717 | Vbrady (≤50), 1260 | |
| Tachy (>140), 1877 | Accelerated ventricular rhythm, 4361 | |
| Pause, 2239 |
*Annotation subset of the MIMIC II database [7]. Vfib = ventricular fibrillation; Vtach = ventricular tachycardia; Brady = bradycardia; Tachy = tachycardia; VBrady = ventricular bradycardia.
Key Insights into the Problem of Alarm Fatigue and Recommendations.
| Conditions Causing Excessive Alarms | Suggestions for Device Improvements |
| 1. Alarms are not tailored to the individual patient | • Have monitor prompt more appropriate alarm settings; e.g., |
| 2. Persistent atrial fibrillation | • Have monitor trigger alarms only upon new onset or termination of atrial fibrillation |
| • When there are repetitive audible alarms, have monitor prompt | |
| 3. Artifact mimics VT or VFib | • Have arrhythmia algorithm use all available ECG leads to identify a non-artifact lead |
| • Make it easy to view and print out all available ECG leads at the time the alarm was triggered | |
| 4. Low amplitude QRS causes pause, asystole, & bradycardia false alarms | • Have arrhythmia algorithm use all available ECG leads to identify QRS complexes |
| • Detect QRS if ≥1 lead has peak-to-trough amplitude of ≥0.15 mV as allowed by the AAMI standard, especially if rate matches SpO2 or arterial pressure waveforms | |
| 5. Wide QRS due to BBB or pacemaker rhythm triggers ventricular arrhythmia alarms | • Have monitor detect high frequency pacemaker “spikes” without clinician having to tell the monitor the patient has a pacemaker |
| • Have monitor algorithm identify P waves to distinguish sinus rhythm with BBB | |
| 6. VT alarms not “actionable” | • Make VT alarm delays configurable according to criteria for documentation (≥10 seconds) or treatment (30 seconds) established by hospital preferences & practice guidelines |
| • Use VT rate, invasive arterial pressure and SpO2 to identify hemodynamically significant (symptomatic) VT | |
| 7. Electrode failure causes poor signal quality | • Have monitor measure each electrode’s impedance and indicate when one is failing so electrode can be changed |
| 8. ST-segment alarms are not truly indicative of myocardial ischemia | • Make ST alarm delays configurable according to criteria for ischemia (lasting 1 minute) to prevent brief “spikes” in ST amplitude from triggering alarms |
| • Define ischemia only when present in 2 contiguous limb leads in order of aVL, I, minus aVR, II, aVF, III) | |
| 9. Flat-line respiratory wave-form cause false apnea & RR alarms | • Have monitor automatically search for best available ECG lead to measure/display respiration waveform |
| • Investigate ECG-derived respiratory measurement to replace impedance method |
Afib = atrial fibrillation; AAMI = Association for the Advancement of Medical Instrumentation; HR = heart rate; BBB = bundle branch block; VT = ventricular tachycardia; VFib = ventricular fibrillation; RR = respiratory rate.