Reproducibility with repeat CT in radiomics study for rectal cancer.

Panpan Hu1,2, Jiazhou Wang1,2, Haoyu Zhong1,2, Zhen Zhou1,2, Lijun Shen1,2, Weigang Hu1,2, Zhen Zhang1,2.   

Abstract

PURPOSE: To evaluate the reproducibility of radiomics features by repeating computed tomographic (CT) scans in rectal cancer. To choose stable radiomics features for rectal cancer.
RESULTS: Volume normalized features are much more reproducible than unnormalized features. The average value of all slices is the most reproducible feature type in rectal cancer. Different filters have little effect for the reproducibility of radiomics features. For the average type features, 496 out of 775 features showed high reproducibility (ICC ≥ 0.8), 225 out of 775 features showed medium reproducibility (0.8 > ICC ≥ 0.5) and 54 out of 775 features showed low reproducibility (ICC < 0.5).
METHODS: 40 rectal cancer patients with stage II were enrolled in this study, each of whom underwent two CT scans within average 8.7 days. 775 radiomics features were defined in this study. For each features, five different values (value from the largest slice, maximum value, minimum value, average value of all slices and value from superposed intermediate matrix) were extracted. Meanwhile a LOG filter with different parameters was applied to these images to find stable filter value. Concordance correlation coefficients (CCC) and inter-class correlation coefficients (ICC) of two CT scans were calculated to assess the reproducibility, based on original features and volume normalized features.
CONCLUSIONS: Features are recommended to be normalized to volume in radiomics analysis. The average type radiomics features are the most stable features in rectal cancer. Further analysis of these features of rectal cancer can be warranted for treatment monitoring and prognosis prediction.to modulate two anticancer compounds in well-defined sets of GBM patients.

Entities:  

Keywords:  radiomics; rectal cancer; reproducibility; textural features

Mesh:

Year:  2016        PMID: 27669756      PMCID: PMC5342090          DOI: 10.18632/oncotarget.12199

Source DB:  PubMed          Journal:  Oncotarget        ISSN: 1949-2553


INTRODUCTION

As human oncology has a strong phenotypic difference from normal tissue, which may be visualized non-invasively by different imaging modalities, such as X-ray computed tomography (CT), positron emission tomography (PET), and magnetic resonance imaging (MRI). Among them, X-ray CT is the most widely applied in oncology, which can assess tissue density in high resolution and exhibit strong contrasts among different tissue types [1]. Nowadays, radiomics become a novel approach because it can utilize medical imaging to quantify the tumor phenotype non-invasively for further study, such as patients' survival, treatment monitoring and outcome prediction [2]. To get a reliable and reproducible result in radiomics study, it is essential to guarantee the repeatability of features extraction process. Recent publications have demonstrated that radiomics features be reproducibly measured from CT images for patients with non-small cell lung cancer [3]. And for nonsmall cell lung cancer (NSCLC) patients, quantitative image features extracted from computed tomography (CT) images can be used to improve tumor diagnosis, staging, and response assessment [4]. However, no studies have yet examined the stability and reproducibility of CT images textural features for rectal cancer. In addition, no researches have been conducted to analyze which kind of extraction process is the best, including max slice, max value, min value, average value or matrix sum. Specifically, max slice means textural features from the slice with the largest GTV area of target images; max value means textural features from all slices of target images and the maximum value was selected; min value means textural features from all slices of target images and the minimum value was selected; average value means textural features from all slices of target images and average all values; and matrix sum means all slices of target images translated into GLCM [5] (gray level co-occurrence matrix) matrices and GLRLM [6] (gray level run-length matrix) matrices and superpose all matrices, then extract textural features from the superposed matrices. Furthermore, in these researches, patients were found who had two sets of CBCT images obtained within 15 minutes, such a short time may not analyze the reproducibility of radiomics. In this study, we analyzed the stability and reproducibility of radiomics features derived from manually segmented rectal tumors in forty patients who underwent two baseline clinical CT scan within average 8.7 days (5 days to 17 days), without any treatment before. Also, we evaluated the five values for each features to find out the most stable value.

RESULTS

Table 1 presents the comparisons of geometry features (volume, area and volume/area) between two scans. It indicates that tumor volume increased 6% with an average 8.7 days' interval.
Table 1

Geometry features comparisons between two scans

Volume (mm3)Area (mm2)V/A (mm)
Scan 145370.8 ± 35271.710921.1 ± 5246.53.9 ± 1.2
Scan 248002.5 ± 28625.210762.3 ± 4712.54.2 ± 0.9
The results of CCC and ICC for different numerical types of unnormalized features and normalized features are listed in the Figure 1 (A, B) and (C, D) respectively:
Figure 1

Results of CCC and ICC distributions for different types of features

(A) Probability density of CCC distribution for unnormalized features, (B) Probability density of ICC distribution for unnormalized features, (C) Probability density of CCC distribution for normalized features, (D) Probability density of ICC distribution for normalized features.

Results of CCC and ICC distributions for different types of features

(A) Probability density of CCC distribution for unnormalized features, (B) Probability density of ICC distribution for unnormalized features, (C) Probability density of CCC distribution for normalized features, (D) Probability density of ICC distribution for normalized features. In Figures 1 (A) and (B), the peak values of probability density for different types of original features are less than 2.5%, but in Figures 1 (C) and (D), the peak values of volume normalized features are more than 6%. For Figures 2 (A) and (B), the mean CCC values and mean ICC values for different types of original features are less than 0.40 and 0.45, but for in Figures 1 (C) and (D), the values are about 0.80 and 0.85, respectively. In Figures 1 (C) and (D), the peak density value and the mean value for average type are larger than other four types' features, so the average type volume normalized features are more reproducible than others.
Figure 2

Results of CCC and ICC for different type of features with different filters

(A) CCC values for unnormalized features, (B) ICC values for unnormalized features, (C) CCC values for normalized features, (D) ICC values for normalized features.

Results of CCC and ICC for different type of features with different filters

(A) CCC values for unnormalized features, (B) ICC values for unnormalized features, (C) CCC values for normalized features, (D) ICC values for normalized features. The mean values of CCC and ICC for different filters of original features are less than 0.45 showed Figures 2 (A) and (B), however the mean values of CCC and ICC for different filters of volume normalized features are 0.75 and 0.80, respectively, showed Figure 2 (C) and (D). Furthermore, no statistically significant difference was observed in CCC and ICC values for different filters, showed in Figure 2 (C) and (D). In Figures 3 (A) and (B), the peak values of probability density for different types of original features are less than 2.5%, but in Figures 3 (C) and (D), the peak values of volume normalized features are 7.5% and 3 %, respectively. In Figure 3 (C), for GLCM features, average value type and matrix sum type are more stable than other types; for GLRLM features, average value type is the most reproduced of all.
Figure 3

Results of ICC values for GLCM and GLRLM features

(A) ICC values for unnormalized GLCM features, (B) ICC values for unnormalized GLRLM features, (C) ICC values for normalized CCC features, (D) ICC values for normalized ICC features.

Results of ICC values for GLCM and GLRLM features

(A) ICC values for unnormalized GLCM features, (B) ICC values for unnormalized GLRLM features, (C) ICC values for normalized CCC features, (D) ICC values for normalized ICC features. Figure 4 illustrates CCC and ICC values with different wavelet characteristics for features, and the mean values for volume normalized features are much larger than unnormalized features. For volume normalized features, CCC and ICC values are all about 0.75 and 0.80, respectively. So there is no significant difference among wavelet features showed in Figure 5 (C) and (D).
Figure 4

Results of CCC and ICC values for different wavelet features

(A) CCC values for unnormalized wavelet features, (B) ICC values for unnormalized wavelet features, (C) CCC values for normalized wavelet features, (D) ICC values for normalized wavelet features.

Figure 5

Histograms of CCC and ICC values of normalized features for the type of average value

Results of CCC and ICC values for different wavelet features

(A) CCC values for unnormalized wavelet features, (B) ICC values for unnormalized wavelet features, (C) CCC values for normalized wavelet features, (D) ICC values for normalized wavelet features. Figure 5 shows histograms of CCC values and ICC values of volume normalized features for the type of average value. Most of features have high reproducibility. For the average type normalized features, 496 out of 775 features (64.0%) showed high reproducibility (ICC ≥ 0.8), 225 out of 775 features (29.0%) showed medium reproducibility (0.8 > ICC ≥ 0.5) and 54 out of 775 features (7.0%) showed low reproducibility (ICC < 0.5).

DISCUSSION

Zhao et al. [7] analyzed variability from same-day repeat CT scans, quite a short wait time cannot analyze the reproducibility of radiomics, but this study evaluated the reproducibility of rectal tumor textural features from repeat CT scans which were underwent within average 8.7 days. Nowadays, radiomics studies for predicting prognosis are classified according to different stages of cancer, the interval from 5 days to 17 days showed in Table 2, and rectal cancer is still in the same stage. For a long interval, but textural features still have a good stability, the result has a higher reliability, and it indicates that CT image's textural features are highly stable and reproduced for rectal cancer.
Table 2

Characteristics of forty patients whose images were used in this study

CharacteristicsNumber of patientsPercent of patients (%)
Number of patients40NA
Median age (range)50.5 (23–76)NA
Median GTV volume (range) (cm3)38.2 (2.3–171.3)NA
Gender
Male2460
Female1640
Days between two scans8.7 (5–17)NA
As we know, human's tumor is changing fast, different time the tumor may be different, to evaluate the reproducibility of radiomics study, the wait time of two CT scans should be a little longer. So an extended wait time is needed to analyze the reproducibility of radiomics features. In this study, the values of the GLCM features were unnormalized and normalized to target volumes. The reason we use this normalization method is the connection between target volume and feature values. As slices area increase, more CT pixel will be included for analysis, and there will be more possibility to include some extreme value. But in GLCM matrix calculation, we use same bins (usually 32) to group these pixels. This will change the distribution shape of the pixel values. For example, in GLCM contract calculation, we use eq(1), where P(i,j) is the value of the pixel, n is the number of the pixels. As range of the CT value increase, the value of contract will decrease as more pixels will centralized. Furthermore, we analyzed the correlation between textural features from the type of average value and tumor volumes, ICC values between textural features and tumor volumes illustrate that 252 out of 775 features (32.5%) have a high correlation and 227 out of 775 features (29.3%) have a medium correlation. Besides, Zou et al. [8] also reported that textural features have a strong correlation with volumes. The results of this study indicated that volume normalized features are much more stable than unnormalized ones. 775 textural features were extracted from patients' medical images with rectal cancer, and we failed to find any statistically significant differences among each texture feature. It suggests that the further analysis of these textural features can be warranted for treatment monitoring and outcome prediction for rectal cancer. Furthermore, we innovatively compared the advantage and disadvantage of five types of patient's medical images. Additionally, the results indicated that different filters have little effect to textural features.

MATERIALS AND METHODS

Patients

Forty rectal cancer patients with stage II were included retrospectively in this study. All patients underwent two baseline clinical CT scans within average 8.7 days (5 days to 17 days) at Fudan University Shanghai Cancer Center in China, before any treatment was delivered. Both scans were obtained with the same CT scanner by using the same imaging protocol (350mA tube current, 120 kVp tube voltage, 0.92 0.92 mm pixel size, 5 mm section collimation, 512 512 matrix). These patients' medical images were divided into two groups: scans 1 and scans 2. The patients' characteristics are showed in Table 2.

Contouring

The rectal GTV was distinguished and segmented by an experienced radiation oncologist in Eclipse (11.0, Varian, Palo Alto, CA), the delineation was double-checked and the non-invaded rectal wall and the air inside the rectum were carefully excluded. After contouring, the DICOM images and the DICOM contours were exported to MATLAB (Math works Inc, Natick, USA) for feature extracting and analysis. The contoured regions of the images were cropped from the whole patient CT image. This was realized by creating a binary mask base on contouring (Figure 6).
Figure 6

Contoured region cropping using binary mask

Radiomics features extraction

We defined 775 radiomics image features to quantify tumor textural characteristics (detail of these features were list in Supplementary Table S1). These features were divided into four groups: I) GLCM [5] textural features, II) GLRLM [6] textural features, III) Wavelet GLCM textural features and IV) Wavelet GLRLM textural features. For each patient, all textural features were extracted from five types of patient's medical images, respectively: I) Max Slice, II) Max value, III) Min Value, IV) Average Value and V) Matrix Sum. All textural features were extracted via Matlab R2015a software (Mathworks Inc, Natick, USA). Zou's study shows lots of features were highly correlated with volume of the tumor [8]. So in this study original features and volume normalized features are both analyzed in this [3].

Data analysis

To estimate the reproducibility and repeatability of the tumor textural features by using repeat CT data, the concordance correlation coefficient (CCC) and the intra-class correlation coefficient (ICC) were initially used [9, 10]. In statistics, the CCC measures the agreement between two variables, for example to evaluate reproducibility or for inter-rater reliability. McGraw and Wong [11] has the form of the concordance correlation coefficient as follow: Where μx and μy are the means for the two variables, σx and σy are the corresponding variances, ρ is the correlation coefficient between the two variables. R package IRR (inter rater reliability) was used for CCC computation [12]. Statistically, the ICC is a descriptive statistic that can be used when quantitative measurements are made on units that are organized into groups [13]. It ranges between 0 and 1, indicating null and perfect reproducibility. In order to determine the ICC for inter-observer segmentations, variance estimates were obtained from two-way mixed effect model of analysis of variance (ANOVA), Leijenaar [14] defined ICC as follow: Where MSR = mean square for rows, MSW = mean square for residual source of variance, k = number of observers involved and n = number of subjects. R package version 3.1.3 IRR was used for ICC computation [12].
  10 in total

Review 1.  Correlation and simple linear regression.

Authors:  Kelly H Zou; Kemal Tuncali; Stuart G Silverman
Journal:  Radiology       Date:  2003-06       Impact factor: 11.105

Review 2.  Intraclass correlations: uses in assessing rater reliability.

Authors:  P E Shrout; J L Fleiss
Journal:  Psychol Bull       Date:  1979-03       Impact factor: 17.737

3.  Evaluating variability in tumor measurements from same-day repeat CT scans of patients with non-small cell lung cancer.

Authors:  Binsheng Zhao; Leonard P James; Chaya S Moskowitz; Pingzhen Guo; Michelle S Ginsberg; Robert A Lefkowitz; Yilin Qin; Gregory J Riely; Mark G Kris; Lawrence H Schwartz
Journal:  Radiology       Date:  2009-07       Impact factor: 11.105

4.  IBEX: an open infrastructure software platform to facilitate collaborative work in radiomics.

Authors:  Lifei Zhang; David V Fried; Xenia J Fave; Luke A Hunter; Jinzhong Yang; Laurence E Court
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

5.  High quality machine-robust image features: identification in nonsmall cell lung cancer computed tomography images.

Authors:  Luke A Hunter; Shane Krafft; Francesco Stingo; Haesun Choi; Mary K Martel; Stephen F Kry; Laurence E Court
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

6.  Stability of FDG-PET Radiomics features: an integrated analysis of test-retest and inter-observer variability.

Authors:  Ralph T H Leijenaar; Sara Carvalho; Emmanuel Rios Velazquez; Wouter J C van Elmpt; Chintan Parmar; Otto S Hoekstra; Corneline J Hoekstra; Ronald Boellaard; André L A J Dekker; Robert J Gillies; Hugo J W L Aerts; Philippe Lambin
Journal:  Acta Oncol       Date:  2013-09-09       Impact factor: 4.089

7.  Can radiomics features be reproducibly measured from CBCT images for patients with non-small cell lung cancer?

Authors:  Xenia Fave; Dennis Mackin; Jinzhong Yang; Joy Zhang; David Fried; Peter Balter; David Followill; Daniel Gomez; A Kyle Jones; Francesco Stingo; Jonas Fontenot; Laurence Court
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

8.  Robust Radiomics feature quantification using semiautomatic volumetric segmentation.

Authors:  Chintan Parmar; Emmanuel Rios Velazquez; Ralph Leijenaar; Mohammed Jermoumi; Sara Carvalho; Raymond H Mak; Sushmita Mitra; B Uma Shankar; Ron Kikinis; Benjamin Haibe-Kains; Philippe Lambin; Hugo J W L Aerts
Journal:  PLoS One       Date:  2014-07-15       Impact factor: 3.240

9.  Quantitative computed tomographic descriptors associate tumor shape complexity and intratumor heterogeneity with prognosis in lung adenocarcinoma.

Authors:  Olya Grove; Anders E Berglund; Matthew B Schabath; Hugo J W L Aerts; Andre Dekker; Hua Wang; Emmanuel Rios Velazquez; Philippe Lambin; Yuhua Gu; Yoganand Balagurunathan; Edward Eikman; Robert A Gatenby; Steven Eschrich; Robert J Gillies
Journal:  PLoS One       Date:  2015-03-04       Impact factor: 3.240

10.  Decoding tumour phenotype by noninvasive imaging using a quantitative radiomics approach.

Authors:  Hugo J W L Aerts; Emmanuel Rios Velazquez; Ralph T H Leijenaar; Chintan Parmar; Patrick Grossmann; Sara Carvalho; Sara Cavalho; Johan Bussink; René Monshouwer; Benjamin Haibe-Kains; Derek Rietveld; Frank Hoebers; Michelle M Rietbergen; C René Leemans; Andre Dekker; John Quackenbush; Robert J Gillies; Philippe Lambin
Journal:  Nat Commun       Date:  2014-06-03       Impact factor: 14.919

  10 in total
  23 in total

1.  Influence of segmentation margin on machine learning-based high-dimensional quantitative CT texture analysis: a reproducibility study on renal clear cell carcinomas.

Authors:  Burak Kocak; Ece Ates; Emine Sebnem Durmaz; Melis Baykara Ulusan; Ozgur Kilickesmez
Journal:  Eur Radiol       Date:  2019-02-12       Impact factor: 5.315

Review 2.  Novel Quantitative Imaging for Predicting Response to Therapy: Techniques and Clinical Applications.

Authors:  Kaustav Bera; Vamsidhar Velcheti; Anant Madabhushi
Journal:  Am Soc Clin Oncol Educ Book       Date:  2018-05-23

Review 3.  Radiomics: an Introductory Guide to What It May Foretell.

Authors:  Stephanie Nougaret; Hichem Tibermacine; Marion Tardieu; Evis Sala
Journal:  Curr Oncol Rep       Date:  2019-06-25       Impact factor: 5.075

4.  Reproducibility of CT-based radiomic features against image resampling and perturbations for tumour and healthy kidney in renal cancer patients.

Authors:  Margherita Mottola; Alessandro Bevilacqua; Stephan Ursprung; Leonardo Rundo; Lorena Escudero Sanchez; Tobias Klatte; Iosif Mendichovszky; Grant D Stewart; Evis Sala
Journal:  Sci Rep       Date:  2021-06-02       Impact factor: 4.379

5.  Reproducibility and non-redundancy of radiomic features extracted from arterial phase CT scans in hepatocellular carcinoma patients: impact of tumor segmentation variability.

Authors:  Qingtao Qiu; Jinghao Duan; Zuyun Duan; Xiangjuan Meng; Changsheng Ma; Jian Zhu; Jie Lu; Tonghai Liu; Yong Yin
Journal:  Quant Imaging Med Surg       Date:  2019-03

6.  Emergence of Radiomics: Novel Methodology Identifying Imaging Biomarkers of Disease in Diagnosis, Response, and Progression.

Authors:  Edward Florez; Ali Fatemi; Pier Paolo Claudio; Candace M Howard
Journal:  SM J Clin Med Imaging       Date:  2018-03-15

7.  Prospective Evaluation of Repeatability and Robustness of Radiomic Descriptors in Healthy Brain Tissue Regions In Vivo Across Systematic Variations in T2-Weighted Magnetic Resonance Imaging Acquisition Parameters.

Authors:  Brendan Eck; Prathyush V Chirra; Avani Muchhala; Sophia Hall; Kaustav Bera; Pallavi Tiwari; Anant Madabhushi; Nicole Seiberlich; Satish E Viswanath
Journal:  J Magn Reson Imaging       Date:  2021-04-16       Impact factor: 5.119

8.  Radiomic Features Are Superior to Conventional Quantitative Computed Tomographic Metrics to Identify Coronary Plaques With Napkin-Ring Sign.

Authors:  Márton Kolossváry; Júlia Karády; Bálint Szilveszter; Pieter Kitslaar; Udo Hoffmann; Béla Merkely; Pál Maurovich-Horvat
Journal:  Circ Cardiovasc Imaging       Date:  2017-12       Impact factor: 7.792

9.  Repeatability of Multiparametric Prostate MRI Radiomics Features.

Authors:  Michael Schwier; Joost van Griethuysen; Mark G Vangel; Steve Pieper; Sharon Peled; Clare Tempany; Hugo J W L Aerts; Ron Kikinis; Fiona M Fennessy; Andriy Fedorov
Journal:  Sci Rep       Date:  2019-07-01       Impact factor: 4.379

10.  Assessment of MRI-Based Radiomics in Preoperative T Staging of Rectal Cancer: Comparison between Minimum and Maximum Delineation Methods.

Authors:  Haidi Lu; Yuan Yuan; Zhen Zhou; Xiaolu Ma; Fu Shen; Yuwei Xia; Jianping Lu
Journal:  Biomed Res Int       Date:  2021-07-10       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.