A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.

Eric Zijlstra1, Lutz Heinemann2, Annelie Fischer2, Christoph Kapitza2.   

Abstract

BACKGROUND: The objective was to evaluate the performance (in terms of accuracy, precision, and trueness) of 5 CE-certified and commercially available blood glucose (BG) systems (meters plus test strips) using an innovative clinical-experimental study design with a 3-step glucose clamp approach and frequent capillary BG measurements.
METHODS: Sixteen subjects with type 1 diabetes participated in this open label, single center trial. BG was clamped at 3 levels for 60 minutes each: 60-100-200 mg/dL. Medical staff performed regular finger pricks (up to 10 per BG level) to obtain capillary blood samples for paired BG measurements with the 5 BG systems and a laboratory method as comparison.
RESULTS: Three BG systems displayed significantly lower mean absolute relative deviations (MARD) (ACCU-Chek® Aviva Nano [5.4%], BGStar® [5.1%], iBGStar® [5.3%]) than 2 others (FreeStyle InsuLinx® [7.7%], OneTouch Verio®IQ [10.3%]). The measurement precision of all BG systems was comparable, but relative bias was also lower for the 3 systems with lower MARD (ACCU-Chek [1.3%], BGStar [-0.9%], iBGStar [1.0%]) compared with the 2 others (FreeStyle [-7.2%], OneTouch [8.9%]).
CONCLUSIONS: This 3 range glucose clamp approach enables a systematic performance evaluation of BG systems under controlled and reproducible conditions. The random error of the tested BG systems was comparable, but some showed a lower systematic error than others. These BG systems allow an accurate glucose measurement at low, normal and high BG levels.
© 2016 Diabetes Technology Society.

Entities:  

Keywords:  accuracy; blood glucose self-monitoring; glucose clamp; glucose meters

Mesh:

Substances:

Year:  2016        PMID: 27605592      PMCID: PMC5094346          DOI: 10.1177/1932296816668373

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

1.  Performance evaluation of blood glucose monitoring devices.

Authors:  Ellen T Chen; James H Nichols; Show-Hong Duh; Glen Hortin
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

2.  The variability of results between point-of-care testing glucose meters and the central laboratory analyzer.

Authors:  Adil I Khan; Yolanda Vasquez; Jacquelyn Gray; Frank H Wians; Martin H Kroll
Journal:  Arch Pathol Lab Med       Date:  2006-10       Impact factor: 5.534

Review 3.  Assessing the analytical performance of systems for self-monitoring of blood glucose: concepts of performance evaluation and definition of metrological key terms.

Authors:  Oliver Schnell; Rolf Hinzmann; Bernd Kulzer; Guido Freckmann; Michael Erbach; Volker Lodwig; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

4.  The precision study: examining the inter- and intra-assay variability of replicate measurements of BGStar, iBGStar and 12 other blood glucose monitors.

Authors:  Sanja Ramljak; Petra B Musholt; Christina Schipper; Frank Flacke; Jochen Sieber; Marcus Borchert; Thomas Forst; Andreas Pfützner
Journal:  Expert Opin Med Diagn       Date:  2013-10-03

5.  System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

6.  Clinical assessment of the accuracy of blood glucose measurement devices.

Authors:  Andreas Pfützner; Michael Mitri; Petra B Musholt; Daniela Sachsenheimer; Marcus Borchert; Andrew Yap; Thomas Forst
Journal:  Curr Med Res Opin       Date:  2012-03-21       Impact factor: 2.580

7.  Acute hypoglycemia decreases myocardial blood flow reserve in patients with type 1 diabetes mellitus and in healthy humans.

Authors:  Omar Rana; Christopher D Byrne; David Kerr; David V Coppini; Soha Zouwail; Roxy Senior; Joe Begley; Jeremy J Walker; Kim Greaves
Journal:  Circulation       Date:  2011-09-12       Impact factor: 29.690

8.  Effects of acute insulin-induced hypoglycemia on indices of inflammation: putative mechanism for aggravating vascular disease in diabetes.

Authors:  Rohana J Wright; David E Newby; David Stirling; Christopher A Ludlam; Ian A Macdonald; Brian M Frier
Journal:  Diabetes Care       Date:  2010-07       Impact factor: 17.152

9.  Accuracy and precision of four value-added blood glucose meters: the Abbott Optium, the DDI Prodigy, the HDI True Track, and the HypoGuard Assure Pro.

Authors:  Catherine A Sheffield; Michael P Kane; Gary Bakst; Robert S Busch; Jill M Abelseth; Robert A Hamilton
Journal:  Diabetes Technol Ther       Date:  2009-09       Impact factor: 6.118

10.  A glucose meter accuracy and precision comparison: the FreeStyle Flash Versus the Accu-Chek Advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic.

Authors:  Linda E Thomas; Michael P Kane; Gary Bakst; Robert S Busch; Robert A Hamilton; Jill M Abelseth
Journal:  Diabetes Technol Ther       Date:  2008-04       Impact factor: 6.118

View more
  2 in total

1.  Analysis of "Capillary and Venous Blood Glucose Accuracy in Blood Glucose Meters Versus Reference Standards: The Impact of Study Design on Accuracy Evaluations".

Authors:  Gary H Thorpe
Journal:  J Diabetes Sci Technol       Date:  2018-09-15

2.  Retrospective Analysis of Continuous Glucose Monitoring Data With the Surveillance Error Grid Supports Nonadjunctive Dosing Decisions.

Authors:  John B Welsh; Tomas Walker; David Price
Journal:  J Diabetes Sci Technol       Date:  2017-02-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.