A glucose meter accuracy and precision comparison: the FreeStyle Flash Versus the Accu-Chek Advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic.

Linda E Thomas1, Michael P Kane, Gary Bakst, Robert S Busch, Robert A Hamilton, Jill M Abelseth.   

Abstract

BACKGROUND: This study compared the accuracy and precision of five blood glucose (BG) meters.
METHODS: Diabetes patients undergoing venipuncture for glucose testing were randomized to one of two groups consisting of three meters: FreeStyle Flash (Abbott Diabetes Care, Alameda, CA), Accu-Chek Advantage (Roche Diagnostics Corp., Indianapolis, IN), and Accu-Chek Compact Plus (Roche Diagnostics) or FreeStyle Flash, Ascensia Contour (Bayer Healthcare, Diagnostic Division, Tarrytown, NY), and BD Logic (BD Diabetes Care, Franklin Lake, NJ). Within 5 min following venipuncture, duplicate finger BG measurements from three ipsilateral fingers were taken. Finger glucose measurements were compared with laboratory reference values. Accuracy was assessed by a Clarke error grid analysis (EGA) and within 10% of the laboratory value criteria. Meter precision was determined by calculating the absolute mean differences in glucose values between duplicate samples.
RESULTS: Finger sticks were obtained from 202 patients. Mean venipuncture BG was 148 mg/dL (SD +/- mg/64 dL; range 25-439 mg/dL). Accuracy by Clarke EGA (Zone A results) was demonstrated in 69% of Advantage samples, 75% of Compact Plus, and 96% of the first group of Flash versus 88% of the Contour, 67% of the Logic, and 91% of the second Flash samples (P < 0.05 for both Flash and Contour). Meter accuracy using the 10% criteria was demonstrated in 30%, 38%, 70%, 46%, 48%, and 68% of the samples, respectively (P < 0.05 for both Flash groups compared to each of the other meters). There were no differences in meter precision.
CONCLUSIONS: No statistically significant differences in accuracy were evident using the Clarke EGA criteria (pooled results of Zone A and B), though the more strict 20% accuracy criteria (Zone A results only) found the Flash and Contour to have significantly greater accuracy compared to the Advantage, Compact Plus, and the Logic. Using the 10% accuracy criteria found the Flash to have significantly greater accuracy than each of the other four meters. All five meters demonstrated similar precision.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18260773     DOI: 10.1089/dia.2007.0244

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  11 in total

1.  Fatigue in women with type 2 diabetes.

Authors:  Cynthia Fritschi; Laurie Quinn; Eileen D Hacker; Sue M Penckofer; Edward Wang; Marquis Foreman; Carol E Ferrans
Journal:  Diabetes Educ       Date:  2012-06-19       Impact factor: 2.140

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 3.  Glycemic control in the burn intensive care unit: focus on the role of anemia in glucose measurement.

Authors:  Elizabeth A Mann; Alejandra G Mora; Heather F Pidcoke; Steven E Wolf; Charles E Wade
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 4.  Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems.

Authors:  Gary H Thorpe
Journal:  Diabetes Technol Ther       Date:  2013-02-15       Impact factor: 6.118

5.  A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.

Authors:  Eric Zijlstra; Lutz Heinemann; Annelie Fischer; Christoph Kapitza
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

Review 6.  Glucose meters: a review of technical challenges to obtaining accurate results.

Authors:  Ksenia Tonyushkina; James H Nichols
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

7.  Transcutaneous glucose sensing by surface-enhanced spatially offset Raman spectroscopy in a rat model.

Authors:  Jonathan M Yuen; Nilam C Shah; Joseph T Walsh; Matthew R Glucksberg; Richard P Van Duyne
Journal:  Anal Chem       Date:  2010-10-15       Impact factor: 6.986

8.  A blended- rather than whole-lentil meal with or without α-galactosidase mildly increases healthy adults' appetite but not their glycemic response.

Authors:  Katherene O-B Anguah; Brittany S Wonnell; Wayne W Campbell; George P McCabe; Megan A McCrory
Journal:  J Nutr       Date:  2014-10-08       Impact factor: 4.798

9.  Performance of the CONTOUR® TS Blood Glucose Monitoring System.

Authors:  Joy Frank; Jane F Wallace; Scott Pardo; Joan Lee Parkes
Journal:  J Diabetes Sci Technol       Date:  2011-01-01

10.  Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study.

Authors:  Enrique Campos-Náñez; Kurt Fortwaengler; Marc D Breton
Journal:  J Diabetes Sci Technol       Date:  2017-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.