Literature DB >> 24088217

The precision study: examining the inter- and intra-assay variability of replicate measurements of BGStar, iBGStar and 12 other blood glucose monitors.

Sanja Ramljak1, Petra B Musholt, Christina Schipper, Frank Flacke, Jochen Sieber, Marcus Borchert, Thomas Forst, Andreas Pfützner.   

Abstract

OBJECTIVE: Self-monitoring of blood glucose is a key element in diabetes management. Accurate and precise performance of blood glucose monitors (BGMs) ensures that valid values are obtained to guide treatment decisions by patients and physicians. BGStar and iBGStar are hand-held BGMs that use dynamic electrochemistry to correct for potential interferences and thereby minimize system errors. RESEARCH DESIGN AND METHODS: A single-center, in vitro diagnostic device performance evaluation with heparinized oxygenated venous blood samples (intra-assay precision) and control solutions (interassay precision) was performed in a laboratory setting, comparing BGStar and iBGStar with 12 competitors. MAIN OUTCOME MEASURES: The primary outcome was the coefficient of variation percent (CV%) of the BGMs investigated.
RESULTS: In inter-assay precision analyses, all but GlucoMen LX had a CV <5%, and in intra-assay precision analyses, 10 of the 14 devices tested had CV <5%. BGStar and iBGStar had a CV <5% in both the inter- and intra-assay precision analyses. The smallest variation was found in the near-normoglycemic glucose range (5.3 - 8.0 mmol/l) for both BGStar and iBGStar in the inter-assay precision analysis.
CONCLUSIONS: BGStar and iBGStar were proven to have very good inter-assay and high intra-assay precision, demonstrating low scattering of replicate measurements with both clinical samples and control solutions.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24088217     DOI: 10.1517/17530059.2013.839984

Source DB:  PubMed          Journal:  Expert Opin Med Diagn        ISSN: 1753-0059


  4 in total

1.  A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.

Authors:  Eric Zijlstra; Lutz Heinemann; Annelie Fischer; Christoph Kapitza
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

2.  Colony-stimulating factor-1: a potential biomarker for lupus nephritis.

Authors:  Julia Menke; Kerstin Amann; Lorenzo Cavagna; Maria Blettner; Arndt Weinmann; Andreas Schwarting; Vicki R Kelley
Journal:  J Am Soc Nephrol       Date:  2014-07-10       Impact factor: 10.121

3.  Role of Mobile Technology to Improve Diabetes Care in Adults with Type 1 Diabetes: The Remote-T1D Study iBGStar® in Type 1 Diabetes Management.

Authors:  Satish K Garg; Viral N Shah; Halis K Akturk; Christie Beatson; Janet K Snell-Bergeon
Journal:  Diabetes Ther       Date:  2017-05-29       Impact factor: 2.945

4.  Brazilian multicenter study for the evaluation of patients' satisfaction of blood glucose self-monitoring with BGStar(®) blood glucose meter in insulinized patients with diabetes mellitus type 1 and 2.

Authors:  Miguel Nasser Hissa
Journal:  Diabetol Metab Syndr       Date:  2016-09-13       Impact factor: 3.320

  4 in total