Accuracy and precision of four value-added blood glucose meters: the Abbott Optium, the DDI Prodigy, the HDI True Track, and the HypoGuard Assure Pro.

Catherine A Sheffield1, Michael P Kane, Gary Bakst, Robert S Busch, Jill M Abelseth, Robert A Hamilton.   

Abstract

PURPOSE: This study compared the accuracy and precision of four value-added glucose meters.
METHODS: Finger stick glucose measurements in diabetes patients were performed using the Abbott Diabetes Care (Alameda, CA) Optium, Diagnostic Devices, Inc. (Miami, FL) DDI Prodigy, Home Diagnostics, Inc. (Fort Lauderdale, FL) HDI True Track Smart System, and Arkray, USA (Minneapolis, MN) HypoGuard Assure Pro. Finger glucose measurements were compared with laboratory reference results. Accuracy was assessed by a Clarke error grid analysis (EGA), a Parkes EGA, and within 5%, 10%, 15%, and 20% of the laboratory value criteria (chi2 analysis). Meter precision was determined by calculating absolute mean differences in glucose values between duplicate samples (Kruskal-Wallis test).
RESULTS: Finger sticks were obtained from 125 diabetes patients, of which 90.4% were Caucasian, 51.2% were female, 83.2% had type 2 diabetes, and average age of 59 years (SD 14 years). Mean venipuncture blood glucose was 151 mg/dL (SD +/-65 mg/dL; range, 58-474 mg/dL). Clinical accuracy by Clarke EGA was demonstrated in 94% of Optium, 82% of Prodigy, 61% of True Track, and 77% of the Assure Pro samples (P < 0.05 for Optium and True Track compared to all others). By Parkes EGA, the True Track was significantly less accurate than the other meters. Within 5% accuracy was achieved in 34%, 24%, 29%, and 13%, respectively (P < 0.05 for Optium, Prodigy, and Assure Pro compared to True Track). Within 10% accuracy was significantly greater for the Optium, Prodigy, and Assure Pro compared to True Track. Significantly more Optium results demonstrated within 15% and 20% accuracy compared to the other meter systems. The HDI True Track was significantly less precise than the other meter systems.
CONCLUSIONS: The Abbott Optium was significantly more accurate than the other meter systems, whereas the HDI True Track was significantly less accurate and less precise compared to the other meter systems.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19764838     DOI: 10.1089/dia.2008.0143

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  5 in total

1.  In vivo, transcutaneous glucose sensing using surface-enhanced spatially offset Raman spectroscopy: multiple rats, improved hypoglycemic accuracy, low incident power, and continuous monitoring for greater than 17 days.

Authors:  Ke Ma; Jonathan M Yuen; Nilam C Shah; Joseph T Walsh; Matthew R Glucksberg; Richard P Van Duyne
Journal:  Anal Chem       Date:  2011-11-02       Impact factor: 6.986

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  Accuracy evaluation of five blood glucose monitoring systems: the North American comparator trial.

Authors:  Solveig Halldorsdottir; Mary Ellen Warchal-Windham; Jane F Wallace; Scott Pardo; Joan Lee Parkes; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

4.  A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.

Authors:  Eric Zijlstra; Lutz Heinemann; Annelie Fischer; Christoph Kapitza
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

5.  Transcutaneous glucose sensing by surface-enhanced spatially offset Raman spectroscopy in a rat model.

Authors:  Jonathan M Yuen; Nilam C Shah; Joseph T Walsh; Matthew R Glucksberg; Richard P Van Duyne
Journal:  Anal Chem       Date:  2010-10-15       Impact factor: 6.986

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.