| Literature DB >> 27593408 |
Luca Lombardo1, Antonella Carlucci1, Mario Palone2, Francesco Mollica3, Giuseppe Siciliani1.
Abstract
BACKGROUND: The aim of this study is to investigate the relative stiffness of straight and mushroom lingual archwires of different diameters, cross sections and alloys, plotting their load/deflection graphs and using a modified three-point bending test.Entities:
Keywords: Lingual straight wire; Mushroom archform; Stainless steel; Stiffness; TMA
Mesh:
Substances:
Year: 2016 PMID: 27593408 PMCID: PMC5018471 DOI: 10.1186/s40510-016-0140-2
Source DB: PubMed Journal: Prog Orthod ISSN: 1723-7785 Impact factor: 2.750
Fig. 1Digital set-up of an ideal patient with moderate crowding using a lingual mushroom archwire (UMW and LMW)
Fig. 2Lingual mushroom and lingual straight archwires are generated by the same 3D virtual set up both for maxillary and mandibular arches
Lingual archwires tested in the study cut at their distal straightest portion according to percentage proportion length calculated by virtual set-up
| Archwire size | Commercial name | Lot number |
|---|---|---|
| SS 0.016 × 0.016 | Ormco | 206-0005 |
| SS 0.016 × 0.022 | Ormco | 206-0006 |
| TMA 0.017 × 0.017 | Ormco | 202-0018 |
| SS 0.018 × 0.018 | Ormco | 204-2302 |
| SS 0.018 × 0.025 | Ormco | 206-0008 |
Fig. 3Deflection, with 1 mm blade, of the portion of a real archwire using a modified three binding test (ISO 15.841 guideline)
Fig. 4Example of a Load/Deflection curve of a 0.018 X 0.018 SS lingual straight archwire
D value of the non-parametric Kolgomorov-Smirnov test for each type of lingual archwires and their statistical significance (p < 0.05*)
| Archwire size |
|
|---|---|
| SS 0.016 × 0.016 | 0.20 (NS) |
| SS 0.016 × 0.022 | 0.15 (NS) |
| TMA 0.017 × 0.017 | 0.31 (NS) |
| SS 0.018 × 0.018 | 0.22 (NS) |
| SS 0.018 × 0.025 | 0.36 (NS) |
NS not significant
Fig. 5Mean values of Load/Deflection curves of SS 0.016 × 0.016-inch, SS 0.016 × 0.022-inch, SS 0.018 × 0.018-inch, SS 0.018 × 0.025-inch and β-Ti 0.0175 × 0.0175-inch archwires using a modified three bending test according to ISO 15.841 guideline
Differences in main values of load/deflection curves (g/mm) of each archwire size and their statistical significance using F ANOVA test (F < 0.001*; F < 0.01**; F < 0.05***)
| Differences in main values of load/deflection curves (g/mm) | ||
|---|---|---|
| SS 0.016 × 0.016 | 400* | 69 |
| SS 0.016 × 0.022 | 585** | 246*** |
| TMA 0.017 × 0.017 | 363.4* | 78*** |
| SS 0.018 × 0.018 | 828.9** | 399.3*** |
| SS 0.018 × 0.025 | 605* | 5 |
| Straight wire vs mushroom upper | Straight wire vs mushroom lower | |