Literature DB >> 27587953

Diffusion Weighted Imaging for Differentiating Benign from Malignant Orbital Tumors: Diagnostic Performance of the Apparent Diffusion Coefficient Based on Region of Interest Selection Method.

Xiao-Quan Xu1, Hao Hu1, Guo-Yi Su1, Hu Liu2, Hai-Bin Shi1, Fei-Yun Wu1.   

Abstract

OBJECTIVE: To evaluate the differences in the apparent diffusion coefficient (ADC) measurements based on three different region of interest (ROI) selection methods, and compare their diagnostic performance in differentiating benign from malignant orbital tumors.
MATERIALS AND METHODS: Diffusion-weighted imaging data of sixty-four patients with orbital tumors (33 benign and 31 malignant) were retrospectively analyzed. Two readers independently measured the ADC values using three different ROIs selection methods including whole-tumor (WT), single-slice (SS), and reader-defined small sample (RDSS). The differences of ADC values (ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS) between benign and malignant group were compared using unpaired t test. Receiver operating characteristic curve was used to determine and compare their diagnostic ability. The ADC measurement time was compared using ANOVA analysis and the measurement reproducibility was assessed using Bland-Altman method and intra-class correlation coefficient (ICC).
RESULTS: Malignant group showed significantly lower ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS than benign group (all p < 0.05). The areas under the curve showed no significant difference when using ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS as differentiating index, respectively (all p > 0.05). The ROISS and ROIRDSS required comparable measurement time (p > 0.05), while significantly shorter than ROIWT (p < 0.05). The ROISS showed the best reproducibility (mean difference ± limits of agreement between two readers were 0.022 [-0.080-0.123] × 10(-3) mm(2)/s; ICC, 0.997) among three ROI methods.
CONCLUSION: Apparent diffusion coefficient values based on the three different ROI selection methods can help to differentiate benign from malignant orbital tumors. The results of measurement time, reproducibility and diagnostic ability suggest that the ROISS method are potentially useful for clinical practice.

Entities:  

Keywords:  ADC; Apparent diffusion coefficient; DWI; Diffusion weighted imaging; Orbit; Tumor

Mesh:

Year:  2016        PMID: 27587953      PMCID: PMC5007391          DOI: 10.3348/kjr.2016.17.5.650

Source DB:  PubMed          Journal:  Korean J Radiol        ISSN: 1229-6929            Impact factor:   3.500


INTRODUCTION

Accurate differentiation between benign and malignant Neckorbital tumors is very important for the pre-treatment plan (1). Conventional computed tomography (CT) and magnetic resonance imaging (MRI) play an essential role in the display of normal structure and location of the orbital lesions; however, the role of conventional CT and MRI features in differentiating benign from malignant orbital tumors is limited (23). Diffusion-weighted imaging (DWI) allows quantification of the diffusion properties of water molecules (4), and its derived apparent diffusion coefficient (ADC) map is widely accepted as a useful tool to detect and characterize orbital lesions (56789). Generally, due to the enlarged nuclei, hyper-cellularity, and subsequent limited extra-cellular space, malignant orbital tumors demonstrate lower ADC value than the benign tumors (5678910). However, the reproducibility of ADC measurements is still a challenge during clinical practice. Previous study has shown that ADC value can be affected by several factors, such as the magnetic field strength, MRI acquisition parameters, and also the region of interest (ROI) selection methods (10). Previous DWI related studies have mainly utilized three kinds of ROI selection methods to obtain ADC values, including whole tumor ROI, single slice ROI, and reader-defined small sample ROI in selected slices (11). ROI on single or selected slice may lead to inter-reader variability, and does not reflect the tissue heterogeneity (12). Whole-tumor ROI can overcome this problem, however, it is a time consuming process, which limits its clinical application (12). The influence of ROI selection methods on the ADC measurements has been evaluated in the liver (13), ovaries (14), pancreas (1115), and breast (16). However, to our knowledge, no previous study has addressed the influence of the three different ROI selection methods on the ADC measurements, and their diagnostic abilities for differentiating benign from malignant orbital tumors. Therefore, the aim of our study was to assess the diagnostic performance of the ADC values based on different ROI selection methods in differentiating benign from malignant orbital tumors at 3T MR.

MATERIALS AND METHODS

Patients

Institutional Review Board approval was obtained for this study, and the requirement for written informed consent was waived because of its retrospective nature. Our MRI database identified 124 consecutive patients who underwent DWI scan as part of pretreatment MRI evaluation for orbital tumors from March 2013 to December 2015. First, according to the definition of orbital indeterminate lesions proposed in previous studies (59), the patients of cavernous malformation (n = 30), lymphangioma (n = 1), venous varix (n = 5), and epidermoid cyst (n = 10) were excluded based on characteristic findings on conventional MRI. Six patients were excluded because the largest diameter of the lesion was < 1 cm, 8 patients were excluded because of the poor image quality of DWI. Finally, 64 patients (36 men and 28 women; mean age, 49.8 ± 15.2 years; range, 18–90 years) were enrolled in our study. Sixty-four patients contained 33 benign and 31 malignant tumors. Demographic and histological information of the two groups was shown in Table 1. The final diagnosis was made based on surgical pathology results in 58 patients, typical imaging features and long-term follow-up in 2 patients with optic nerve sheath meningioma, and follow-up after steroid treatment in 4 patients with inflammatory pseudotumor.
Table 1

Demographic and Histological Information between Two Groups

VariablesBenign Group (n = 33)Malignant Group (n = 31)P
Age (years)44.5 ± 12.955.4 ± 15.70.004
Gender (M/F)19/1417/140.825
DiagnosisInflammatory pseudotumor (10)Lymphoma (22)-
Pleomorphic adenoma (9)Adenoid cystic carcinoma (2)
Schwannoma (6)Metastases (3)
Optic nerve sheath meningioma (4)Basal cell carcinoma (1)
IgG4-related disease (3)Lymphoepithelial carcinoma (1)
Solitary fibrous tumor (1)Ewing sarcoma (1)
Melanoma (1)

Data in parentheses indicates number of corresponding patients in our study. F = female, IgG4 = immunoglobulin G4, M = male

MR Scan

MRI was performed using a 3T MR system (Verio; Siemens, Erlangen, Germany), with a twelve-channel head coil. Conventional MR images were acquired as follows: unenhanced axial T1-weighted imaging (repetition time [TR]/echo time [TE], 600/10 msec); axial, coronal and sagittal T2-weighted imaging (TR/TE, 4700/79 msec) with fat saturation; contrast-enhanced axial, coronal and sagittal T1-weighted imaging (TR/TE, 500/10 msec). For contrast-enhanced axial T1-weighted imaging, a standard dose of 0.1 mmol/kg of gadolinium-diethylene triamine pentaacetic acid (Magnevist; Bayer Schering Pharma AG, Berlin, Germany) was administered at a rate of 4 mL/s, followed by a 20-mL bolus of saline at the same injection rate. Diffusion weighted imaging was obtained in the transverse plane using the single-shot spin-echo echo-planar sequence with three orthogonal diffusion gradients. Imaging parameters for DWI were as follows: TR/TE, 4000/85 ms; b factors, 0 and 800 s/mm2; section thickness, 4 mm with no gap; flip angle, 15°; field of view, 200 × 200 mm; matrix, 384 × 384; number of sections, 10; number of averages, 6. The total acquisition time of DWI was 4 minutes 14 seconds.

Imaging Analysis

Diffusion weighted imaging data were processed using an off-line in-house software (FireVoxel; CAI2R; New York University, New York, NY, USA) (12). The ADC map was constructed and the ADC value was calculated using the following mono-exponential fitting formula: ADC = -ln (Sb/S0) / b, where b represents the diffusion sensitivity coefficients, and Sb and S0 represent the corresponding signal values of the given ROI. When applying the whole-tumor ROI (ROIWT) method, freehand ROIs were drawn along the border of the high signal of the tumor on DWI (b = 800) to cover the entire tumor area on each tumor containing slice, and the mean ADC value was derived and defined as ADC-ROIWT. When applying single-slice ROI (ROISS) method, the slice in which the tumor demonstrated the largest diameter was chosen. Free-hand ROIs were drawn to cover as much tumor tissue as possible in this slice. Similarly, the mean ADC value was derived and defined as ADC-ROISS. When applying reader-defined small sample ROI (ROIRDSS) method, three freehand circular ROIs (area range, 0.467–0.622 cm2; mean, 0.506 cm2) were drawn on the area in which the tumor demonstrated relative higher signal intensity on DW image (b = 800). ADC values obtained from three ROIs were averaged for further analysis and designated ADC-ROIRDSS. For the three ROI methods, the visually cystic and necrotic areas were excluded with reference to T2-weighted and contrast enhanced T1-weighted images; and the time required for ADC measurements based on three different ROI methods was also recorded. Schematic illustration of the three different ROI selection methods was shown in Figure 1.
Fig. 1

Schematic illustration of three ROI selection methods on DW (b = 800 s/mm2) images obtained from 70-year-old patient with orbital lymphoma.

Axial T2-weighted image shows lesion molding in right orbit (A). For reader-defined small sample ROI, three freehand circular ROIs with mean area of approximately 0.5 cm2 were drawn on area in which tumor showed relative higher signal intensity on DW image (b = 800) (B). For selected slice ROI, slice in which tumor showed largest diameter was chosen. Free-hand ROIs was drawn to cover as much tumor tissue as possible in this slice (C), then corresponding ADC map could be generated and embed within DW image (D). For whole-tumor ROI, freehand ROIs were drawn along border of tumor to cover entire tumor area on each tumor-containing slice (E, four maps). ADC = apparent diffusion coefficient, DW = diffusion weighted, ROI = regions of interest

Two experienced neuro-radiologists (reader 1 with 14 years of experience; reader 2 with 5 years of experience) independently measured the tumor ADC values according to the three ROI selection methods. The measurement results were averaged between the two readers for further analyses.

Statistical Analysis

Numeric data were reported as mean ± standard deviation. Kolmogorov-Smirnov test was used for analyzing the normality of data distribution. Unpaired t test was used to compare the difference of age and mean ADC values between benign and malignant orbital tumors group. Chi-square test was used to compare the difference of gender between the two groups. Multiple receiver operating characteristic (ROC) curves comparison according to the method of DeLong et al. (17), were used to compare the diagnostic ability of the mean ADC values based on three different ROIs in predicting malignant orbital tumors. The times required for ADC measurement based on the three ROI selection methods were compared with ANOVA analysis. Inter-reader variability of orbital tumor ADC measurements for each ROI selection method was analyzed by using the intra-class correlation coefficient (ICC) (18), and the method of Bland-Altman (19). The statistical significance threshold was set at a two-sided p value < 0.05. All statistical analyses were carried out using two statistical packages (SPSS version 19.0; SPSS Inc., Chicago, IL, USA and MedCalc version 13; MedClac Software, Ostend, Belgium).

RESULTS

There was significant difference in age (p = 0.004), while no difference in gender (p = 0.825) between the benign and malignant groups. Table 2 summarized the comparison of ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS value between benign and malignant group. Malignant group demonstrated lower mean ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS value than benign group (ADC-ROIWT, 0.808 ± 0.280 vs. 1.795 ± 0.470, p < 0.001; ADC-ROISS, 0.773 ± 0.230 vs. 1.754 ± 0.461, p < 0.001; ADC-ROIRDSS, 0.675 ± 0.213 vs. 1.605 ± 0.423, p < 0.001). Comparison of ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS value between the two groups was also shown in box plots (Fig. 2).
Table 2

Comparison of ADC between Two Groups

ADC ValueBenign Group (n = 33)Malignant Group (n = 31)P
ADC-ROIWT1.795 ± 0.4700.808 ± 0.280< 0.001
ADC-ROISS1.754 ± 0.4610.773 ± 0.230< 0.001
ADC-ROIRDSS1.605 ± 0.4230.675 ± 0.213< 0.001

Data are reported as mean ± standard deviation; unit for ADC value is × 10-3 mm2/s. ADC = apparent diffusion coefficient, ROI = region of interest, ROIRDSS = ROI based on reader-defined small sample in selected slices, ROISS = ROI based on single slice, ROIWT = ROI based on whole tumor

Fig. 2

Box-and-whisker plots of ADC-ROIWT (A), ADC-ROISS (B), and ADC-ROIRDSS (C) values between benign and malignant groups.

*Outliers. ADC = apparent diffusion coefficient, ROI = region of interest, ROIRDSS = ROI based on reader-defined small sample in selected slices, ROISS = ROI based on single slice, ROIWT = ROI based on whole tumor

The average time required for ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS measurement was 82.1 ± 26.3, 30.0 ± 6.5, and 31.5 ± 8.3 seconds. The time required for ADC-ROIWT measurement was significantly longer than ADC-ROISS and ADC-ROIRDSS (both p < 0.001), while no significant difference was found in the time required for ADC-ROISS and ADC-ROIRDSS (p = 0.093). Receiver operating characteristic curves analysis indicated that setting the ADC-ROIWT = 1.251 × 10-3 mm2/s, ADC-ROISS = 1.084 × 10-3 mm2/s, ADC-ROIRDSS = 1.140 × 10-3 mm2/s as threshold value resulted in optimal diagnostic value {(ADC-ROIWT; area under curve [AUC], 0.948; sensitivity, 96.8%; specificity, 84.9%) (ADC-ROISS; AUC, 0.966; sensitivity, 93.6%; specificity, 90.9%) (ADC-ROIRDSS; AUC, 0.969; sensitivity, 100%; specificity, 84.9%)}. Multiple ROC curves comparison indicated that the AUCs showed no significant differences when using ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS as differentiating index, respectively (ADC-ROIWT vs. ADC-ROISS, p = 0.073; ADC-ROISS vs. ADC-ROIRDSS, p = 0.610; ADC-ROIWT vs. ADC-ROIRDSS, p = 0.064). ROC analyses using ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS to predict orbital malignancy were showed in Table 3 and Figure 3.
Table 3

Diagnostic Ability of ADC Based on Different ROIs for Predicting Orbital Malignancy

VariablesCut-Off ValueAUCSensitivity (%)Specificity (%)
ADC value
 ADC-ROIWT1.2510.948 (0.862–0.988)96.8 (83.3–99.9)84.9 (68.1–94.9)
 ADC-ROISS1.0840.966 (0.887–0.995)93.6 (78.6–99.2)90.9 (75.7–98.1)
 ADC-ROIRDSS1.1400.969 (0.892–0.996)100 (88.8–100)84.9 (68.1–94.9)

Data in parentheses are 95% confidence intervals. Unit for ADC value is × 10-3 mm2/s. ADC = apparent diffusion coefficient, AUC = area under curve, ROI = region of interest, ROIRDSS = ROI based on reader-defined small sample in selected slices, ROISS = ROI based on single slice, ROIWT = ROI based on whole tumor

Fig. 3

ROC analyses using ADC-ROI Multiple ROC curves comparison indicated no significant differences on area under curves when using ADC-ROIWT, ADC-ROISS, and ADC-ROIRDSS as differentiating index, respectively (ADC-ROIWT vs. ADC-ROISS, p = 0.073; ADC-ROISS vs. ADC-ROIRDSS, p = 0.610; ADC-ROIWT vs. ADC-ROIRDSS, p = 0.064). ADC = apparent diffusion coefficient, AUC = area under curve, ROC = receiver operating characteristic, ROI = region of interest, ROIRDSS = ROI based on reader-defined small sample in selected slices, ROISS = ROI based on single slice, ROIWT = ROI based on whole tumor

The mean difference (bias) and limits of agreement between two readers were 0.022 (-0.080–0.123) × 10-3 mm2/s for single-slice ROI (ICC, 0.997; 95% confidence interval [CI], 0.994–0.998), 0.024 (-0.245–0.292) × 10-3 mm2/s for ROIWT (ICCs, 0.977; 95% CI, 0.962–0.986), and 0.192 (-0.135–0.518) × 10-3 mm2/s for small solid sample ROI (ICC, 0.959; 95% CI, 0.933–0.974). The mean ADCs in small solid sample ROI were more scattered than the other two ROI methods. Graphic illustrations of these data with Bland-Altman plots were displayed in Figure 4.
Fig. 4

Bland-Altman plots of inter-reader agreement for ADC-ROI Difference of mean ADC value between two readers (y-axis) was plotted against mean ADC value of two readers (x-axis), with mean absolute difference (bias) (solid line) and 95% confidence interval of mean difference (limits of agreement) (dashed lines). ADC = apparent diffusion coefficient, ROI = region of interest, ROIRDSS = ROI based on reader-defined small sample in selected slices, ROISS = ROI based on single slice, ROIWT = ROI based on whole tumor

DISCUSSION

In this study, we compared the mean ADC values derived from three different ROI selection methods of the benign and malignant orbital tumors. ADC values based on the three ROI selection methods of the malignant orbital tumors were significantly lower than those of the benign tumors. In addition, ROISS showed the best reproducibility among three ROI selection methods; whereas, the ROISS and ROIRDSS showed significantly shorter measurement time than ROIWT. Based on the measurement time, reproducibility and diagnostic ability, ROISS has the potential for use in measurement of the ADC value of orbital tumors in clinical practice. The present results demonstrated that ADC values based on the three different ROI selection methods of the malignant orbital tumors were significantly lower than those of the benign tumors. Enlarged nuclei and hyper-cellularity in malignant tumors would reduce the diffusivity of water protons in the extracellular and intracellular space, and result in decreased ADC values (589). In addition, despite variation in diagnostic performance of the ADC values based on the three ROI selection methods, the AUC values from ROC analyses showed no significant difference among these three methods. Our study results indicated that ADC values based on the three ROI selection methods might be useful metrics for predicting orbital malignancy. Each ROI selection method has its own merits and limitations. During clinical usage, ROIRDSS was the most commonly used ROI selection method for measuring the ADC values. Previous studies indicated that since the ROIRDSS was commonly placed on the area where the tumor showed relative higher signal intensity on DW images, ROIRDSS tended to reflect the most solid, cellular or viable portion of the tumor (1120). However, our study results indicated that inter-reader variability was an major limitation of ROIRDSS. To overcome this limitation, some researchers proposed the ROIWT approach (5712). However, since the ROIWT approach is a time-consuming process, much longer measurement time could be a limitation for its routine application in clinical practice. In contrast, the ROISS had significantly better reproducibility than ROIRDSS, and significantly shorter measurement time than ROIWT. Thus, the ROISS would be the most appropriate approach to measure ADC values of orbital tumors in clinical practice. Our study has some limitations. First, the study population was relatively small. Future studies are required to validate the present results in a larger population. Second, final diagnosis of some cases was based on the typical imaging appearance and long-term follow-up or diagnostic treatment, without histological conformation. Third, 8 of 124 DWI studies could not be interpreted due to inadequate image quality. Future study using periodically rotated overlapping parallel lines with enhanced reconstruction technique or readout-segmented echo-planar imaging technique might be helpful to reduce the artifacts (921). In conclusion, our study showed that the ADC values derived from all three ROI selection methods could help to differentiate benign from malignant orbital tumors. The ROI selection methods had no definite influence on the diagnostic performance of ADC values. The results of measurement time, reproducibility and diagnostic ability suggest that the ROISS method could be used to measure the ADC value of orbital tumors in clinical practice.
  21 in total

1.  DWI-associated entire-tumor histogram analysis for the differentiation of low-grade prostate cancer from intermediate-high-grade prostate cancer.

Authors:  Chen-Jiang Wu; Qing Wang; Hai Li; Xiao-Ning Wang; Xi-Sheng Liu; Hai-Bin Shi; Yu-Dong Zhang
Journal:  Abdom Imaging       Date:  2015-10

2.  Differentiation between benign and malignant orbital tumors at 3-T diffusion MR-imaging.

Authors:  Ahmed Abdel Khalek Abdel Razek; Sahar Elkhamary; Amani Mousa
Journal:  Neuroradiology       Date:  2011-02-01       Impact factor: 2.804

3.  Interobserver variability of selective region-of-interest measurement protocols for quantitative diffusion weighted imaging in soft tissue masses: Comparison with whole tumor volume measurements.

Authors:  Shivani Ahlawat; Paras Khandheria; Filippo Del Grande; John Morelli; Ty K Subhawong; Shadpour Demehri; Laura M Fayad
Journal:  J Magn Reson Imaging       Date:  2015-07-14       Impact factor: 4.813

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Diffusion-weighted imaging of malignant ocular masses: initial results and directions for further study.

Authors:  A R Sepahdari; R Kapur; V K Aakalu; J P Villablanca; M F Mafee
Journal:  AJNR Am J Neuroradiol       Date:  2011-11-24       Impact factor: 3.825

Review 6.  Orbital neoplasms in adults: clinical, radiologic, and pathologic review.

Authors:  Tina D Tailor; Divakar Gupta; Roberta W Dalley; C Dirk Keene; Yoshimi Anzai
Journal:  Radiographics       Date:  2013-10       Impact factor: 5.333

7.  Orbital lymphoproliferative disorders (OLPDs): value of MR imaging for differentiating orbital lymphoma from benign OPLDs.

Authors:  K Haradome; H Haradome; Y Usui; S Ueda; T C Kwee; K Saito; K Tokuuye; J Matsubayashi; T Nagao; H Goto
Journal:  AJNR Am J Neuroradiol       Date:  2014-05-29       Impact factor: 3.825

8.  Utility of histogram analysis of ADC maps for differentiating orbital tumors.

Authors:  Xiao-Quan Xu; Hao Hu; Guo-Yi Su; Hu Liu; Xun-Ning Hong; Hai-Bin Shi; Fei-Yun Wu
Journal:  Diagn Interv Radiol       Date:  2016 Mar-Apr       Impact factor: 2.630

9.  Visibility of sutures of the orbit and periorbital region using multidetector computed tomography.

Authors:  Hubert Gufler; Markus Preiß; Sabrina Koesling
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

10.  Is diffusion-weighted MRI useful for differentiation of small non-necrotic cervical lymph nodes in patients with head and neck malignancies?

Authors:  Hyun Kyung Lim; Jeong Hyun Lee; Hye Jin Baek; Namkug Kim; Hayoung Lee; Jee Won Park; Sang Yoon Kim; Kyung Ja Cho; Jung Hwan Baek
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

View more
  7 in total

Review 1.  Teprotumumab (Tepezza): from the discovery and development of medicines to USFDA approval for active thyroid eye disease (TED) treatment.

Authors:  Faraat Ali; Anushma Chorsiya; Varisha Anjum; Asad Ali
Journal:  Int Ophthalmol       Date:  2021-01-22       Impact factor: 2.031

2.  Histogram analysis of apparent diffusion coefficient maps for differentiating malignant from benign parotid gland tumors.

Authors:  Gao Ma; Liu-Ning Zhu; Guo-Yi Su; Hao Hu; Wen Qian; Shou-Shan Bu; Xiao-Quan Xu; Fei-Yun Wu
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-07-02       Impact factor: 2.503

3.  Single shot echo planar imaging (ssEPI) vs single shot turbo spin echo (ssTSE) DWI of the orbit in patients with ocular melanoma.

Authors:  Ekim Gumeler; Safak Parlak; Gozde Yazici; Erdem Karabulut; Hayyam Kiratli; Kader K Oguz
Journal:  Br J Radiol       Date:  2020-12-22       Impact factor: 3.039

4.  Selection and Reporting of Statistical Methods to Assess Reliability of a Diagnostic Test: Conformity to Recommended Methods in a Peer-Reviewed Journal.

Authors:  Ji Eun Park; Kyunghwa Han; Yu Sub Sung; Mi Sun Chung; Hyun Jung Koo; Hee Mang Yoon; Young Jun Choi; Seung Soo Lee; Kyung Won Kim; Youngbin Shin; Suah An; Hyo-Min Cho; Seong Ho Park
Journal:  Korean J Radiol       Date:  2017-09-21       Impact factor: 3.500

5.  Assessment of lacrimal glands in thyroid eye disease with diffusion-weighted magnetic resonance imaging.

Authors:  Ahmed Abdel Razek; El-Hadidy Mohamed El-Hadidy; Mohamed El-Said Moawad; Nader El-Metwaly; Amr Abd El-Hamid El-Said
Journal:  Pol J Radiol       Date:  2019-03-08

6.  A data-driven statistical model that estimates measurement uncertainty improves interpretation of ADC reproducibility: a multi-site study of liver metastases.

Authors:  Ryan Pathak; Hossein Ragheb; Neil A Thacker; David M Morris; Houshang Amiri; Joost Kuijer; Nandita M deSouza; Arend Heerschap; Alan Jackson
Journal:  Sci Rep       Date:  2017-10-26       Impact factor: 4.379

7.  Evaluation of Effects of TGF-β1 Inhibition on Gastric Cancer in Nude Mice by Diffusion Kurtosis Imaging and In-Line X-ray Phase Contrast Imaging With Sequential Histology.

Authors:  Bowen Shi; Fei Yuan; Fuhua Yan; Huan Zhang; Zilai Pan; Weibo Chen; Guilong Wang; Jingwen Tan; Yang Zhang; Yuqi Ren; Lianjun Du
Journal:  J Magn Reson Imaging       Date:  2018-10-06       Impact factor: 4.813