Results of Primary Total Hip Arthroplasty with 36-mm Femoral Heads on Highly Cross-linked Polyethylene-Minimum Seven-years Follow-up.

Won-Kee Choi1, Myung-Rae Cho1, Joo-Hwan Lee1.   

Abstract

PURPOSE: We evaluate the clinical and radiographic midterm results of primary total hip arthroplasty (THA) using a 36 mm diameter femoral head on highly cross-linked polyethylene (minimum 7-year follow-up).
MATERIALS AND METHODS: We retrospectively reviewed 73 patients (74 hips) that underwent primary THA with a 36 mm diameter femoral head on highly cross-linked polyethylene between July 2004 and February 2007. Clinical follow-ups included specific measurements like modified Harris hip scores (HHS) and Merle d'Aubigne and Postel score. For radiologic evaluations, together with position of acetabular cup at 6 weeks later of post-operation, we separately calculated the penentrations of femoral heads into polyethylene liners during post-operation and one year later check-ups, and during one year later check-ups and final check-ups.
RESULTS: There were no complications except for one case of dislocation. Average modified HHS at final follow-up was 88±7.5 (range, 81-96), and Merle d'Aubigne and Postel scores were more than 15 (range, 15-18). Mean acetabular cup inclination and anteversion were 50.1°(range, 35°-58°) and 23.6°(range, 5°-38°), respectively. Average femoral head penetration during the first postoperative year was 0.071±0.034 mm/year, and steady-state wear rate determined using radiographs taken at one-year postoperatively and at latest follow-up was 0.051±0.022 mm/year. Average femoral head penetration during entire follow-ups was 0.058±0.013 mm/year.
CONCLUSION: Primary THA with a large diameter femoral head on highly cross-linked polyethylene was found to produce the results comparable to previous in vitro laboratory hip simulation studies. And we also find out good scores in terms of patient's functionality.

Entities:  

Keywords:  36-mm femoral head; Highly cross-linked polyethylene; Minimum 7-years follow-up; Total hip arthroplasty

Year:  2014        PMID: 27536585      PMCID: PMC4971397          DOI: 10.5371/hp.2014.26.4.220

Source DB:  PubMed          Journal:  Hip Pelvis        ISSN: 2287-3260


  17 in total

Review 1.  Dislocation after total hip arthroplasty: implant design and orientation.

Authors:  Robert L Barrack
Journal:  J Am Acad Orthop Surg       Date:  2003 Mar-Apr       Impact factor: 3.020

2.  Femoral head sizes larger than 32 mm against highly cross-linked polyethylene.

Authors:  Brian R Burroughs; Harry E Rubash; William H Harris
Journal:  Clin Orthop Relat Res       Date:  2002-12       Impact factor: 4.176

3.  Early failure due to osteolysis associated with contemporary highly cross-linked ultra-high molecular weight polyethylene. A case report.

Authors:  Letitia Bradford; Robert Kurland; Meera Sankaran; Hubert Kim; Lisa A Pruitt; Michael D Ries
Journal:  J Bone Joint Surg Am       Date:  2004-05       Impact factor: 5.284

4.  The measurement of creep in ultrahigh molecular weight polyethylene: a comparison of conventional versus highly cross-linked polyethylene.

Authors:  Daniel M Estok; Charles R Bragdon; Gordon R Plank; Anna Huang; Orhun K Muratoglu; William H Harris
Journal:  J Arthroplasty       Date:  2005-02       Impact factor: 4.757

Review 5.  The pros and cons of using larger femoral heads in total hip arthroplasty.

Authors:  Pranav Rathi; Gavin C Pereira; Mauro Giordani; Paul E Di Cesare
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2013-08

6.  Cementless acetabular components.

Authors:  C A Engh; W L Griffin; C L Marx
Journal:  J Bone Joint Surg Br       Date:  1990-01

7.  High stress conditions do not increase wear of thin highly crosslinked UHMWPE.

Authors:  Natalie H Kelly; Amar D Rajadhyaksha; Timothy M Wright; Suzanne A Maher; Geoffrey H Westrich
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

8.  Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review.

Authors:  R L Barrack; R D Mulroy; W H Harris
Journal:  J Bone Joint Surg Br       Date:  1992-05

9.  Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results.

Authors:  C A Engh; J D Bobyn; A H Glassman
Journal:  J Bone Joint Surg Br       Date:  1987-01

10.  The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.

Authors:  Maiken Stilling; Kristian Larsen; Niels T Andersen; Kjeld Søballe; Søren Kold; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.