C A Engh1, W L Griffin, C L Marx.
Abstract
Four hundred and fifteen patients with cementless acetabular components of either a smooth threaded (130) or porous surfaced (285) variety were compared for clinical symptoms and radiographic signs of component loosening. At a mean 4.8 year follow-up none of the patients with porous acetabular components had signs of component instability. At a mean 3.9 year follow-up 27 (21%) of the patients with a smooth threaded acetabular component showed radiographic signs of instability and 33 (25%) had clinical symptoms. The disappointing short-term results with these threaded cups in our hands have prompted us to abandon their use in favour of the porous surfaced hemispherical cups.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2298795
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X