Annelie N Tan1,2, Carroll A B Webers2, Tos T J M Berendschot2, John de Brabander2, Pauline M de Witte3, Rudy M M A Nuijts2, Johannes S A G Schouten2, Henny J M Beckers2.
Abstract
PURPOSE: To investigate central and peripheral corneal endothelial cell density (ECD) in relation to Baerveldt (BV) glaucoma drainage device (GDD) tube corneal (TC) distance.Entities:
Keywords: anterior segment OCT; corneal endothelial cell loss; glaucoma; glaucoma drainage implants
Mesh:
Year: 2016 PMID: 27495264 PMCID: PMC5298043 DOI: 10.1111/aos.13161
Source DB: PubMed Journal: Acta Ophthalmol ISSN: 1755-375X Impact factor: 3.761
Figure 1Endothelial cell density (ECD) measurement, (A) Superior corneal ECD, (B) Central corneal ECD, (C) Inferior corneal ECD, (D) Left superior corneal ECD, (E) Right superior corneal ECD.
Figure 2Anterior segment optical coherence tomography (AS‐OCT) showing the Baerveldt tube and the tube corneal distance in pink. (A) Tube ‘free’ in the AC, (B) Transiridial placement of the tube.
Baseline characteristics
| No of eyes ( | 53 |
| Mean age in years (mean ± SD) | 61 ± 14 |
| Gender (% men) | 46 |
| Eye (% right eye) | 51 |
| Lens status (% pseudophakic) | 84.9 |
| Glaucoma type | |
| Primary open‐angle glaucoma (%) | 24.5 ( |
| Secondary glaucoma (%) | 56.6 ( |
| Primary angle‐closure glaucoma (%) | 18.9 ( |
| Previous trabeculectomy (%) | 67.8 |
| Endothelial cell density preoperatively (cells/mm2) | |
| Tube ‘free’ in the anterior chamber (mean ± SD) | 2091 ± 344 |
| Transiridial placement of the tube (mean ± SD) | 2017 ± 547 |
Central endothelial cell density (ECD) at different time points
| Period | Mean ECD (cells/mm2) ± SD |
|---|---|
| Preoperatively | 2052 ± 572 |
| 3 months postop | 2016 ± 592 |
| 6 months postop | 2012 ± 607 |
| 12 months postop | 1911 ± 640 |
| 24 months postop | 1898 ± 657 |
| 36 months postop | 1771 ± 662 |
β‐coefficients and their 95% confidence intervals showing difference in endothelial cell density (ECD) loss between the quadrant closest to the Baerveldt (BV) and the other quadrants
| Parameter | Estimate | Sig. | 95% Confidence interval | |
|---|---|---|---|---|
| Lower bound | Upper bound | |||
| Intercept | 1939 | <0.001 | 1776 | 2101 |
|
| −10.6 | <0.001 | −12.6 | −8.7 |
|
| 14.3 | 0.408 | −19.6 | 48.2 |
|
| 0 | |||
|
| 3.23 | 0.003 | 1.10 | 5.36 |
|
| 0 | |||
PQC = peripheral quadrant closest to the BV GDD tube.
β‐coefficients and their 95% confidence intervals showing the difference in central endothelial cell density (ECD) (cells/mm2) loss for different tube corneal (TC) distances
| Parameter | Estimate | Sig. | 95% Confidence interval | |
|---|---|---|---|---|
| Lower bound | Upper bound | |||
| Intercept (cells/mm2) | 2229 | <0.001 | 1988 | 2470 |
|
| −15.7 | <0.001 | −20.8 | −10.5 |
|
| −142.5 | 0.008 | −246.8 | −38.2 |
|
| 4.51 | 0.004 | 1.46 | 7.55 |