Literature DB >> 25393097

Effect of glaucoma tube shunt parameters on cornea endothelial cells in patients with Ahmed valve implants.

Euna B Koo1, Jing Hou, Ying Han, Jeremy D Keenan, Robert L Stamper, Bennie H Jeng.   

Abstract

PURPOSE: The aim of this study was to assess the effect of various tube parameters on corneal endothelial cell density (ECD) after insertion of Ahmed valves.
METHODS: Thirty-nine eyes of 33 patients with previous superotemporal (ST) Ahmed valve implantation and 20 eyes of 13 participants with previous uncomplicated phacoemulsification and intraocular lens implantation but no history of glaucoma surgery were evaluated. Various tube parameters were measured with anterior segment optical coherence tomography. ST, central, and inferonasal (IN) ECD and pachymetry were measured. Endothelial cell loss and corneal thickness in the ST cornea was compared with those in the IN cornea.
RESULTS: The mean age of the operated patients was 58 ± 22 years, and the mean time since glaucoma surgery was 2.5 ± 2.6 years. Thirty-two of the 39 study eyes were pseudophakic. The ECD was significantly lower in the ST endothelium than in the IN endothelium in eyes with glaucoma tube surgery (P < 0.001), although this relative reduction in ST ECD was not greater than that seen in pseudophakic control eyes (P = 0.16). In univariate analysis, tube angle relative to the cornea and distance from the tip of the tube to the cornea were significant risk factors for decreased ST endothelial cell loss when assessed relative to the IN ECD (P = 0.01 and P = 0.02, respectively). In multivariate analysis, only the distance of the tube tip to the cornea remained significantly associated with ST endothelial cell loss.
CONCLUSIONS: Although this was a retrospective study with inherent limitations, tubes that are closer to the cornea seem to lead to increased loss of adjacent endothelial cells.

Entities:  

Mesh:

Year:  2015        PMID: 25393097     DOI: 10.1097/ICO.0000000000000301

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  16 in total

1.  Baerveldt surgery outcomes: anterior chamber insertion versus vitreous cavity insertion.

Authors:  Naoki Tojo; Atsushi Hayashi; Tomoko Consolvo-Ueda; Shuichiro Yanagisawa
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-08-31       Impact factor: 3.117

2.  Clinical outcomes of Ahmed glaucoma valve in anterior chamber versus ciliary sulcus.

Authors:  A Bayer; M Önol
Journal:  Eye (Lond)       Date:  2016-12-16       Impact factor: 3.775

3.  Effect of shunt type on rates of tube-cornea touch and corneal decompensation after tube shunt surgery in uveitic glaucoma.

Authors:  Sapna Sinha; Allen Y Ganjei; Melih Ustaoglu; Zeba A Syed; Daniel Lee; Jonathan S Myers; Scott J Fudemberg; Reza Razeghinejad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-01-30       Impact factor: 3.117

4.  Ahmed glaucoma valve implantation with the tube placement in the ciliary sulcus: short-term results.

Authors:  Gul Arikan; Betul Akbulut; Canan Asli Utine; Ziya Ayhan; Mahmut Kaya; Taylan Ozturk; Uzeyir Gunenc
Journal:  Int Ophthalmol       Date:  2021-10-10       Impact factor: 2.031

Review 5.  Corneal Endothelial Cell Loss in Glaucoma and Glaucoma Surgery and the Utility of Management with Descemet Membrane Endothelial Keratoplasty (DMEK).

Authors:  Neeru A Vallabh; Stephnie Kennedy; Riccardo Vinciguerra; Keri McLean; Hannah Levis; Davide Borroni; Vito Romano; Colin E Willoughby
Journal:  J Ophthalmol       Date:  2022-01-30       Impact factor: 1.974

6.  Corneal Edema and Keratoplasty: Risk Factors in Eyes With Previous Glaucoma Drainage Devices.

Authors:  Bradley Beatson; Jiangxia Wang; Michael V Boland; Pradeep Ramulu; Oliver Schein; Michael J Fliotsos; Michael E Sulewski; Divya Srikumaran
Journal:  Am J Ophthalmol       Date:  2022-01-05       Impact factor: 5.488

7.  Modified Descemet's stripping automated endothelial keratoplasty for eyes with glaucoma drainage devices and shallow anterior chambers.

Authors:  Alfonso Vasquez-Perez; Syed Mohammed Naqib; Mayank A Nanavaty
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

8.  The Effect of Tube Location on Corneal Endothelial Cells in Patients with Ahmed Glaucoma Valve.

Authors:  Qi Zhang; Yingna Liu; Sukhumal Thanapaisal; Julius Oatts; Yetao Luo; Gui-Shuang Ying; Junming Wang; Stephen D McLeod; Steven J Gedde; Ying Han
Journal:  Ophthalmology       Date:  2020-06-27       Impact factor: 12.079

Review 9.  Ahmed glaucoma valve implant: surgical technique and complications.

Authors:  Ivano Riva; Gloria Roberti; Francesco Oddone; Anastasios Gp Konstas; Luciano Quaranta
Journal:  Clin Ophthalmol       Date:  2017-02-17

10.  Corneal endothelial cell loss after Baerveldt glaucoma drainage device implantation in the anterior chamber.

Authors:  Annelie N Tan; Carroll A B Webers; Tos T J M Berendschot; John de Brabander; Pauline M de Witte; Rudy M M A Nuijts; Johannes S A G Schouten; Henny J M Beckers
Journal:  Acta Ophthalmol       Date:  2016-08-06       Impact factor: 3.761

View more