| Literature DB >> 27487206 |
Aeri Yoo1, Ki Yeun Nam2, Hungwon Tchah3, Myoung Joon Kim3.
Abstract
PURPOSE: To compare a modified phacoemulsification tip with the established micro tip, in terms of temperature at the corneal wound, efficiency, and chatter events, using the Centurion® Vision system.Entities:
Mesh:
Year: 2016 PMID: 27487206 PMCID: PMC4972308 DOI: 10.1371/journal.pone.0159049
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1New modified phaco tip and sleeve.
Left: Balanced tip has same inner caliber as Kelman tip, however, its thickness decrease using titanum material. Right: Sleeve has been modified with 4 rib- like space as well.
Fig 2Porcine lens cutting methods.
Left: Own cutting knife was made with acryl plate and laser cutter with 2mm interval. Middle: The lens was trapped between the blades, cutting knife were inserted plate slots. Right: The porcine lens is cut into 2.0 mm3 cubes.
Temperature at the incision site at a70% longitudinal setting.
| Mean Incision Temperature (°C) | ||||
|---|---|---|---|---|
| Baseline | Elevation | Maximum | Corneal burn | |
| Kelman(bare tip) | 22.4 | 13 | 35.4 | no |
| Balanced(bare tip) | 22.1 | 15.5 | 37.6 | no |
| Kelman(+sleeve) | 22.2 | 23.5 | 46.7 | yes |
| Balanced(+new sleeve) | 22.3 | 21.9 | 45.2 | yes |
| Bare tip vs Sleeved tip | Group 1 vs Group 2 | Group 3 vs Group 4 | ||
| P—value | 0.021 | 0.763 | 0.896 | |
Fig 3Temperature change during phacoemulsification.
Temperature increase over time at 70% longitudinal power. Arrows indicate thestart point of the corneal burn.
Temperature at the incision site at a100% torsional setting.
| Mean Incision Temperature (°C) | ||||
|---|---|---|---|---|
| Baseline | Elevation | Maximum | Corneal burn | |
| Kelman(bare tip) | 22.4 | 10.4 | 32.8 | no |
| Balanced(bare tip) | 22.2 | 11 | 33.2 | no |
| Kelman(+sleeve) | 22.1 | 16.6 | 38.7 | no |
| Balanced(+new sleeve) | 21.8 | 14 | 35.8 | no |
| Bare tip vs Sleeved tip | Group 1 vs Group 2 | Group 3 vs Group 4 | ||
| P—value | 0.327 | 0.756 | 0.007 | |
Fig 4Efficiency and chatter events.
A. The balanced tip shows short efficiency times at 100% torsional power. B. Chatter events show no statistical differences between the two phaco tips.