Literature DB >> 17321348

Fracture of a carotid stent: an unexpected complication.

Arif Rahim Valibhoy1, Bibombe Patrice Mwipatayi, Kishore Sieunarine.   

Abstract

We present the first reported case of a carotid artery stent fracture where a Nitinol stent was used for the management of carotid artery stenosis. An 83-year-old man underwent right carotid artery stenting in July 2005. On follow-up ultrasound, stenosis of the right internal carotid artery was noted. In January 2006, he underwent left carotid artery stenting and at this procedure, precontrast assessment of the right neck revealed a complete fracture through the waist of the stent in the right internal carotid artery. This fractured segment correlated with the area of stenosis reported at the previous ultrasound. In July 2006, this gentleman underwent successful removal of the carotid stent with uneventful carotid endarterectomy and synthetic patch repair. Carotid artery stenting is a newer and alternative modality to carotid endarterectomy for the treatment of carotid artery stenting. More data are emerging to support its safety and efficacy. An unexpected complication can happen with any procedure and complications help define the follow-up management of subsequent patients undergoing this procedure.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17321348     DOI: 10.1016/j.jvs.2006.08.086

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  8 in total

1.  The use of flat panel angioCT (DynaCT) for navigation through a deformed and fractured carotid stent.

Authors:  Pasquale Mordasini; Fahmi Al-Senani; Jan Gralla; Dai-Do Do; Caspar Brekenfeld; Gerhard Schroth
Journal:  Neuroradiology       Date:  2009-07-28       Impact factor: 2.804

2.  Effect of head posture on the healthy human carotid bifurcation hemodynamics.

Authors:  Yannis Papaharilaou; Nicolas Aristokleous; Ioannis Seimenis; Mohammad Iman Khozeymeh; Georgios C Georgiou; Brigitta C Brott; Elena Eracleous; Andreas S Anayiotos
Journal:  Med Biol Eng Comput       Date:  2012-11-10       Impact factor: 2.602

3.  Cerebral Air Embolism and Vasospasm after Carotid Revascularization: A Case Report.

Authors:  Tsong-Hai Lee; Ho-Fai Wong
Journal:  Case Rep Neurol       Date:  2020-12-14

4.  Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T.

Authors:  Andreas Simgen; Désirée Ley; Christian Roth; Giorgio Franco Maria Cattaneo; Ruben Mühl-Benninghaus; Andreas Müller; Heiko Körner; Yoo-Jin Kim; Bruno Scheller; Wolfgang Reith; Umut Yilmaz
Journal:  Neuroradiology       Date:  2016-07-25       Impact factor: 2.804

Review 5.  Intravascular foreign bodies: danger of unretrieved fragmented medical devices.

Authors:  Minori Tateishi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

6.  In situ fracture of stents implanted for relief of pulmonary arterial stenosis in patients with congenitally malformed hearts.

Authors:  Doff B McElhinney; Lisa Bergersen; Audrey C Marshall
Journal:  Cardiol Young       Date:  2008-06-18       Impact factor: 1.093

7.  Fracture of a carotid stent and restenosis of common carotid artery.

Authors:  Wojciech Wróbel; Maciej Lesiak; Anna Sinkiewicz
Journal:  Arch Med Sci       Date:  2013-02-10       Impact factor: 3.318

8.  "Whiplash" cervical trauma with fracture and migration of carotid stent fragments.

Authors:  Danilo Almeida Silva; Guilherme Brasileiro de Aguiar; Maurício Jory; Mario Luiz Marques Conti; José Carlos Esteves Veiga
Journal:  Surg Neurol Int       Date:  2020-10-08
  8 in total