We report an investigation on the accuracy of dose calculation based on the cone-beam computed tomography (CBCT) images of the nonbowtie filter kV imaging system of the Vero linear accelerator. Different sets of materials and tube voltages were employed to generate the Hounsfield unit lookup tables (HLUTs) for both CBCT and fan-beam CT (FBCT) systems. The HLUTs were then implemented for the dose calculation in a treatment planning system (TPS). Dosimetric evaluation was carried out on an in-house-developed cube phantom that consists of water-equivalent slabs and inhomogeneity inserts. Two independent dosimeters positioned in the cube phantom were used in this study for point-dose and two-dimensional (2D) dose distribution measurements. The differences of HLUTs from various materials and tube voltages in both CT systems resulted in differences in dose calculation accuracy. We found that the higher the tube voltage used to obtain CT images, the better the point-dose calculation and the gamma passing rate of the 2D dose distribution agree to the values determined in the TPS. Moreover, the insert materials that are not tissue-equivalent led to higher dose-calculation inaccuracy. There were negligible differences in dosimetric evaluation between the CBCT- and FBCT-based treatment planning if the HLUTs were generated using the tissue-equivalent materials. In this study, the CBCT images of the Vero system from a complex inhomogeneity phantom can be applied for the TPS dose calculation if the system is calibrated using tissue-equivalent materials scanned at high tube voltage (i.e., 120 kV).Entities:
Mesh:
Year: 2016 PMID: 27455496 PMCID: PMC5690058 DOI: 10.1120/jacmp.v17i4.6194
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1A Vero system equipped with two kV imagers for CBCT.
Calculated chamber doses and their relative differences to the measured dose resulting from different HLUTs at different tube voltages both in FBCT and CBCT
|
|
|
| |||||
|---|---|---|---|---|---|---|---|
|
|
| Δ |
| Δ |
| Δ | |
| FBCT | Catphan |
|
|
|
|
|
|
| Gammex |
| 0.6 |
| 0.5 |
| 0.3 | |
| SAM |
| 0.9 |
| 0.6 |
| 0.5 | |
| SC |
| 1.2 |
| 0.9 |
| 0.8 | |
| CBCT | Catphan |
|
|
|
|
|
|
| Gammex |
|
|
|
|
|
| |
| SAM |
|
|
|
|
|
| |
Figure 2An in‐house phantom used for dosimetry study scanned in (a) FBCT and (b) CBCT with tube voltage of 120 kV.
Figure 3HLUT from different sets of materials resulting from (a) FBCT and (b) CBCT at 120 kV.
Gamma index (3% delta dose/3 mm DTA) for absolute dose values of 2D dose distributions measured by the EDR2 film compared with dose calculation in the TPS using different HLUTs for both CT systems
|
|
|
|
| |
|---|---|---|---|---|
| FBCT | Catphan | 75.31% | 77.99% | 78.76% |
| Gammex | 90.59% | 95.82% | 98.65% | |
| SAM | 86.73% | 92.35% | 97.32% | |
| SC | 85.14% | 90.00% | 95.40% | |
| CBCT | Catphan | 54.27% | 57.29% | 64.56% |
| Gammex | 91.81% | 93.02% | 96.96% | |
| SAM | 92.29% | 94.53% | 96.79% |
Figure 4HLUTs of the Gammex with solid lines taken from CBCT and dashed lines taken from FBCT for all tube voltages.