A formulation of tissue- and water-equivalent materials using the stoichiometric analysis method for CT-number calibration in radiotherapy treatment planning.

Indra Yohannes1, Daniel Kolditz, Oliver Langner, Willi A Kalender.   

Abstract

Tissue- and water-equivalent materials (TEMs) are widely used in quality assurance and calibration procedures, both in radiodiagnostics and radiotherapy. In radiotherapy, particularly, the TEMs are often used for computed tomography (CT) number calibration in treatment planning systems. However, currently available TEMs may not be very accurate in the determination of the calibration curves due to their limitation in mimicking radiation characteristics of the corresponding real tissues in both low- and high-energy ranges. Therefore, we are proposing a new formulation of TEMs using a stoichiometric analysis method to obtain TEMs for the calibration purposes. We combined the stoichiometric calibration and the basic data method to compose base materials to develop TEMs matching standard real tissues from ICRU Report 44 and 46. First, the CT numbers of six materials with known elemental compositions were measured to get constants for the stoichiometric calibration. The results of the stoichiometric calibration were used together with the basic data method to formulate new TEMs. These new TEMs were scanned to validate their CT numbers. The electron density and the stopping power calibration curves were also generated. The absolute differences of the measured CT numbers of the new TEMs were less than 4 HU for the soft tissues and less than 22 HU for the bone compared to the ICRU real tissues. Furthermore, the calculated relative electron density and electron and proton stopping powers of the new TEMs differed by less than 2% from the corresponding ICRU real tissues. The new TEMs which were formulated using the proposed technique increase the simplicity of the calibration process and preserve the accuracy of the stoichiometric calibration simultaneously.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22330195     DOI: 10.1088/0031-9155/57/5/1173

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  13 in total

1.  Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction.

Authors:  Adam S Wang; J Webster Stayman; Yoshito Otake; Gerhard Kleinszig; Sebastian Vogt; Gary L Gallia; A Jay Khanna; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2014-02-07       Impact factor: 3.609

2.  Iterative metal artifact reduction improves dose calculation accuracy : Phantom study with dental implants.

Authors:  Manuel Maerz; Pia Mittermair; Andreas Krauss; Oliver Koelbl; Barbara Dobler
Journal:  Strahlenther Onkol       Date:  2016-03-11       Impact factor: 3.621

3.  Proton-counting radiography for proton therapy: a proof of principle using CMOS APS technology.

Authors:  G Poludniowski; N M Allinson; T Anaxagoras; M Esposito; S Green; S Manolopoulos; J Nieto-Camero; D J Parker; T Price; P M Evans
Journal:  Phys Med Biol       Date:  2014-05-01       Impact factor: 3.609

4.  Liquid tissue surrogates for X-ray and CT phantom studies.

Authors:  Paul F FitzGerald; Robert E Colborn; Peter M Edic; Jack W Lambert; Peter J Bonitatibus; Benjamin M Yeh
Journal:  Med Phys       Date:  2017-10-31       Impact factor: 4.071

5.  Effective atomic numbers of some tissue substitutes by different methods: A comparative study.

Authors:  Vishwanath P Singh; N M Badiger
Journal:  J Med Phys       Date:  2014-01

6.  Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources.

Authors:  Mahdi Ghorbani; Fateme Salahshour; Abbas Haghparast; Toktam Ahmadi Moghaddas; Courtney Knaup
Journal:  J Contemp Brachytherapy       Date:  2014-04-03

7.  Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems.

Authors:  Reza Mahmoudi; Nasrollah Jabbari; Mehdi Aghdasi; Hamid Reza Khalkhali
Journal:  PLoS One       Date:  2016-07-08       Impact factor: 3.240

8.  Noncoplanar verification: a feasibility study using Philips' Pinnacle3 treatment planning system.

Authors:  Indra Yohannes; Heru Prasetio; Christoph Bert
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

9.  Development of clinical application program for radiotherapy induced cancer risk calculation using Monte Carlo engine in volumetric-modulated arc therapy.

Authors:  Dong-Jin Kang; Young-Joo Shin; Seonghoon Jeong; Jae-Yong Jung; Hakjae Lee; Boram Lee
Journal:  Radiat Oncol       Date:  2021-06-12       Impact factor: 3.481

10.  A new tissue segmentation method to calculate 3D dose in small animal radiation therapy.

Authors:  C Noblet; G Delpon; S Supiot; V Potiron; F Paris; S Chiavassa
Journal:  Radiat Oncol       Date:  2018-02-26       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.