| Literature DB >> 27455493 |
Grace Tang1, Christopher Moussot, Daniel Morf, Edward Seppi, Howard Amols.
Abstract
Most of the treatment units, both new and old models, are equipped with a megavoltage portal imager but its use for volumetric imaging is limited. This is mainly due to the poor image quality produced by the high-energy treatment beam (> 6MV). A linac at our center is equipped with a prototype 2.5 MV imaging beam. This study evaluates the feasibility of low-dose megavoltage cone-beam imaging with the 2.5MV beam and a thick cesium iodide detector, which is a high-efficiency imager. Basic imaging properties such as spatial resolution and modulation transfer function were assessed for the 2.5 MV prototype imaging system. For image quality and imaging dose, a series of megavoltage cone-beam scans were acquired for the head, thorax, and pelvis of an anthropomorphic phantom and were compared to kilovoltage cone-beam and 6X megavoltage cone-beam images. To demonstrate the advantage of MV imaging, a phantom with metallic inserts was scanned and the image quality was compared to CT and kilovoltage cone-beam scans. With a lower energy beam and higher detector efficiency, the 2.5 MV imaging system generally yields better image quality than does the 6 MV imaging system with the conventional MV imager. In particular, with the anthropomorphic phantom studies, the contrast to noise of bone to tissue is generally improved in the 2.5 MV images compared to 6 MV. With an image quality sufficient for bony alignment, the imaging dose for 2.5 MV cone-beam images is 2.4-3.4 MU compared to 26 MU in 6 MV cone-beam scans for the head, thorax, and pelvis regions of the phantom. Unlike kilovoltage cone-beam, the 2.5 MV imaging system does not suffer from high-Z image artifacts. This can be very useful for treatment planning in cases where high-Z prostheses are present.Entities:
Mesh:
Year: 2016 PMID: 27455493 PMCID: PMC5690043 DOI: 10.1120/jacmp.v17i4.6185
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Schematics of the 2.5 MV imaging beamline in comparison to the treatment beamline in the treatment head (not to scale).
Figure 4CBCT scans of the RANDO head phantom.
Figure 2Contrast‐detail curves of and systems.
Figure 3MTF of 2.5X MVCB and 6X MVCB systems.
Figure 5CBCT scans of the RANDO thorax phantom.
Figure 6CBCT scans of the RANDO pelvis phantom.
Figure 7Fan‐beam CT (kVFB), kVCB, and 2.5X MVCB of a PMMA phantom with metallic inserts.
CNR of different tissues and regions of RANDO for 2.5X MVCB, 6X MVCB, and kVCB. For the low dose 2.5X MVCB scan, 2.4 MU, 2.6 MU, and 3.4 MU were used for the head, thorax, and pelvis, respectively; while for the high dose 2.5X MVCB scan, 8.8 MU, 9.0 MU, and 8.9 MU were used, respectively. For 6X MVCB, 26 MU were used for all three phantom regions
|
| |||||
|---|---|---|---|---|---|
|
|
|
|
|
| |
| RANDO Head | Bone – Tissue | 11.6 | 30.3 | 4.7 | 49.5 |
| RANDO Thorax | Bone – Tissue Bone – Lung | 3.5 7.3 | 3.3 22.1 | 2.6 12.3 | 18.1 69.5 |
| RANDO Pelvis | Bone – Tissue | 2.9 | 7.4 | 5.4 | 39.5 |
Figure 8For similar image quality, the imaging dose of MVCB with CsI imager is approximately 3–5 times less than MVCB with aS1000, as demonstrated in a head scan and a pelvis scan.
Imaging dose of 2.5X MVCB and kVCB with a head and a body CTDI phantom. The nominal MVCB dose is calculated based on the low‐dose scan setting for a head phantom and pelvis phantom, at 2.4 MU and 3.4 MU, respectively
|
|
|
|
|
|
|---|---|---|---|---|
| Head, | Isocenter | 0.56 | 1.34 | 0.39 |
| full fan | 1 cm below surface | 0.73 | 1.75 | 0.60 |
| Body, | Isocenter | 0.41 | 1.39 |
|
| half fan | 1 cm below surface | 0.42 | 1.43 |
|