Effect of various environments and computed tomography scanning parameters on renal volume measurements in vitro: A phantom study.

Wangyan Liu1, Yinsu Zhu1, Lijun Tang1, Xiaomei Zhu1, Yi Xu1, Guanyu Yang2.   

Abstract

Kidney volume is an important parameter in clinical practice, and accurate assessment of kidney volume is vital. The aim of the present study was to evaluate the effect of various environments, tube voltages, tube currents and slice thicknesses on the accuracy of a novel segmentation software in determining renal volume on computed tomography (CT) images. The volumes of potatoes and porcine kidneys were measured on CT images and compared with the actual volumes, which were determined by a water displacement method. CT scans were performed under various situations, including different environments (air or oil); tube voltages/tube currents (80 kVp/200 mAs, 120 kVp/200 mAs, 120 kVp/100 mAs); and reconstructed slice thicknesses (0.75 or 1.5 mm). Percentage errors (PEs) relative to the reference standards were calculated. In addition, attenuation and image noise under different CT scanning parameters were compared. Student's t-test was also used to analyze the effect of various conditions on image quality and volume measurements. The results indicated that the volumes measured in oil were closer to the actual volumes (P<0.05). Furthermore, attenuation and image noise significantly increased when using a tube voltage of 80 kVp, while the mean PEs between 120 and 80 kVp voltages were not significantly different. The mean PEs were greater when using a tube current of 100 mAs compared with a current of 200 mAs (P<0.05). In addition, the volumes measured on 1.5 mm slice thickness were closer to the actual volumes (P<0.05). In conclusion, different environments, tube currents and slice thicknesses may affect the volume measurements. In the present study, the most accurate volume measurements were obtained at 120 kVp/200 mAs and a slice thickness of 1.5 mm.

Entities:  

Keywords:  computed tomography; in vitro; tube current; tube voltage; volume measurements

Year:  2016        PMID: 27446271      PMCID: PMC4950443          DOI: 10.3892/etm.2016.3414

Source DB:  PubMed          Journal:  Exp Ther Med        ISSN: 1792-0981            Impact factor:   2.447


  28 in total

1.  Knowledge-based segmentation of pediatric kidneys in CT for measurement of parenchymal volume.

Authors:  M S Brown; W C Feng; T R Hall; M F McNitt-Gray; B M Churchill
Journal:  J Comput Assist Tomogr       Date:  2001 Jul-Aug       Impact factor: 1.826

2.  Renal volume measurements: accuracy and repeatability of US compared with that of MR imaging.

Authors:  J Bakker; M Olree; R Kaatee; E E de Lange; K G Moons; J J Beutler; F J Beek
Journal:  Radiology       Date:  1999-06       Impact factor: 11.105

3.  Small pulmonary nodules: volume measurement at chest CT--phantom study.

Authors:  Jane P Ko; Henry Rusinek; Erika L Jacobs; James S Babb; Margrit Betke; Georgeann McGuinness; David P Naidich
Journal:  Radiology       Date:  2003-09       Impact factor: 11.105

4.  Low contrast agent and radiation dose protocol for hepatic dynamic CT of thin adults at 256-detector row CT: effect of low tube voltage and hybrid iterative reconstruction algorithm on image quality.

Authors:  Takeshi Nakaura; Shinichi Nakamura; Natsuki Maruyama; Yoshinori Funama; Kazuo Awai; Kazunori Harada; Shouzaburou Uemura; Yasuyuki Yamashita
Journal:  Radiology       Date:  2012-05-24       Impact factor: 11.105

5.  Practical and effective ALARA.

Authors:  Joseph John Bevelacqua
Journal:  Health Phys       Date:  2010-05       Impact factor: 1.316

6.  Low-kilovoltage, high-tube-current MDCT of liver in thin adults: pilot study evaluating radiation dose, image quality, and display settings.

Authors:  Takeshi Nakaura; Kazuo Awai; Seitaro Oda; Yoshinori Funama; Kazunori Harada; Shouzaburou Uemura; Yasuyuki Yamashita
Journal:  AJR Am J Roentgenol       Date:  2011-06       Impact factor: 3.959

7.  A clinical MRI investigation of the relationship between kidney volume measurements and renal function in patients with renovascular disease.

Authors:  S J Gandy; K Armoogum; R S Nicholas; T B McLeay; J G Houston
Journal:  Br J Radiol       Date:  2007-01       Impact factor: 3.039

8.  Correlation between loss of renal function and loss of renal volume after partial nephrectomy for tumor in a solitary kidney.

Authors:  Nidhi Sharma; Jerome O'Hara; Andrew C Novick; Michael Lieber; Erick M Remer; Brian R Herts
Journal:  J Urol       Date:  2008-02-20       Impact factor: 7.450

9.  Determination of split renal function by 3D reconstruction of CT angiograms: a comparison with gamma camera renography.

Authors:  Adam L Summerlin; Mark E Lockhart; Andrew M Strang; Peter N Kolettis; Naomi S Fineberg; J Kevin Smith
Journal:  AJR Am J Roentgenol       Date:  2008-11       Impact factor: 3.959

10.  Renal measurements on CT angiograms: correlation with graft function at living donor renal transplantation.

Authors:  Motoyo Yano; Michael F Lin; Kelsey A Hoffman; Anitha Vijayan; Thomas K Pilgram; Vamsi R Narra
Journal:  Radiology       Date:  2012-07-12       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.