| Literature DB >> 27413753 |
Chung-Feng Hwang1, Hui-Chen Ko2, Yung-Ting Tsou2, Kai-Chieh Chan2, Hsuan-Yeh Fang2, Che-Ming Wu2.
Abstract
Objectives. We evaluated the causes, hearing, and speech performance before and after cochlear implant reimplantation in Mandarin-speaking users. Methods. In total, 589 patients who underwent cochlear implantation in our medical center between 1999 and 2014 were reviewed retrospectively. Data related to demographics, etiologies, implant-related information, complications, and hearing and speech performance were collected. Results. In total, 22 (3.74%) cases were found to have major complications. Infection (n = 12) and hard failure of the device (n = 8) were the most common major complications. Among them, 13 were reimplanted in our hospital. The mean scores of the Categorical Auditory Performance (CAP) and the Speech Intelligibility Rating (SIR) obtained before and after reimplantation were 5.5 versus 5.8 and 3.7 versus 4.3, respectively. The SIR score after reimplantation was significantly better than preoperation. Conclusions. Cochlear implantation is a safe procedure with low rates of postsurgical revisions and device failures. The Mandarin-speaking patients in this study who received reimplantation had restored auditory performance and speech intelligibility after surgery. Device soft failure was rare in our series, calling attention to Mandarin-speaking CI users requiring revision of their implants due to undesirable symptoms or decreasing performance of uncertain cause.Entities:
Mesh:
Year: 2016 PMID: 27413753 PMCID: PMC4927948 DOI: 10.1155/2016/8962180
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Demographic backgrounds of the cases.
| Linkou | Kaohsiung & Xiamen | Total | |
|---|---|---|---|
|
| 401 | 188 | 589 |
|
| 210/191 | 102/86 | 312/277 |
|
| 186/215 | 103/85 | 289/300 |
|
| |||
| ≤3.0 | 160 | 80 | 240 |
| 3.1–5.0 | 88 | 49 | 137 |
| 5.1–18.0 | 100 | 34 | 134 |
| >18.0 | 53 | 25 | 78 |
|
| |||
| CI24M | 61 | 7 | 68 |
| CI24R (CS, CA) | 92 | 81 | 173 |
| CI24RE (Freedom) | 194 | 75 | 269 |
| CI512 | 43 | 14 | 57 |
| CI422 | 11 | 11 | 22 |
|
| |||
| Inverted J | 222 | 33 | 255 |
| Minimally invasive | 179 | 155 | 334 |
|
| |||
| Congenital | 243 | 144 | 387 |
| LVAS | 61 | 13 | 74 |
| Progressive | 21 | 6 | 27 |
| Inner ear anomaly | 28 | 5 | 33 |
| Waardenburg syndrome | 8 | 3 | 11 |
| Meningitis | 7 | 1 | 8 |
| Sudden deafness | 3 | 4 | 7 |
| Multiply handicapped | 5 | 3 | 8 |
| Other | 25 | 9 | 34 |
Causes of major complications following cochlear implantation.
| Linkou | Kaohsiung & Xiamen | Total | |
|---|---|---|---|
|
| |||
| Device failure | 7 | 1 | 8 |
| Hematoma needing revision | 0 | 1 | 1 |
| Permanent facial paralysis | 0 | 0 | 0 |
| Mastoiditis | 2 | 1 | 3 |
| Severe scalp flap infections | 9 | 0 | 9 |
| Meningitis | 0 | 0 | 0 |
| Retraction pocket with cholesteatoma | 0 | 0 | 0 |
| Persistent CSF leakage | 0 | 0 | 0 |
| Electrode misplacement/pull-out | 0 | 0 | 0 |
| Magnet displacement | 1 | 0 | 1 |
| Total number of cases | 19 | 3 | 22 |
| Rate of major complications | 4.74% | 1.60% | 3.74% |
Causes of revision surgery following cochlear implantation.
| Linkou | Kaohsiung & Xiamen | Total | |
|---|---|---|---|
|
| 7 | 1 | 8 |
| Hard failure | 7 | 1 | 8 |
| Soft failure | 0 | 0 | 0 |
|
| 9 | 1 | 10 |
| Infection | 9 | 0 | 9 |
| Hematoma | 0 | 1 | 1 |
|
| 0 | 0 | 0 |
| Total number of cases | 16 | 2 | 18 |
| Rate of CI revision | 3.99% | 1.06% | 3.06% |
Patients undergoing reimplantation and hearing/speech performance before and after surgery.
| Number | Cause | Implant/SP | Age at CI (years) | Hearing and speech before reimplantation | Hearing and speech after reimplantation | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| PTA (dB HL) | CAP | SIR | Tone recognition (%) | Easy sentence (%) | Difficult sentence (%) | PB word (%) | Interval (years) | PTA (dB HL) | CAP | SIR | Tone recognition (%) | Easy sentence (%) | Difficult sentence (%) | PB word (%) | ||||
| R1 | Infection (flap) | CI24RE/CP(810) | 6.5 | 32.5 | 7 | 5 | n/a | n/a | n/a | n/a | 12.7 | 25 | 6 | 5 | 80 | 92 | 95 | 36 |
| R2 | Infection (device) | CI24M/Sprint | 1.5 | 27.5 | 6 | 5 | n/a | n/a | n/a | n/a | 4.3 | 30.0 | 6 | 5 | 88 | 84 | 69 | 88 |
| R3 | Infection (flap) | CI24R(CS)/Esprit | 5.0 | 33.8 | 4 | 2 | n/a | n/a | n/a | n/a | 1.9 | 31.3 | 4 | 3 | n/a | n/a | n/a | n/a |
| R4 | Infection (device) | CI24R(CS)/CP(810) | 4.3 | 32.5 | 7 | 5 | 85 | 100 | 95 | 84 | 4.1 | 28.8 | 7 | 5 | 74 | 100 | 95 | 84 |
| R5 | Device failure | CI24R(CS)/CP(810) | 3.1 | 33.8 | 4 | 2 | n/a | n/a | n/a | n/a | 1.4 | 23.8 | 6 | 4 | 53 | 32 | 24 | 36 |
| R6 | Device failure | CI24RE/CP(810) | 2.7 | 30.0 | 6 | 4 | 53 | 100 | 92 | 98 | 3.9 | 31.3 | 6 | 5 | 76 | 98 | 91 | 84 |
| R7 | Device failure | CI24RE/CP(810) | 3.1 | 28.8 | 4 | 2 | n/a | n/a | n/a | n/a | 1.3 | 25 | 5 | 3 | 41 | 40 | 28 | 68 |
| R8 | Device failure | CI24RE/CP(810) | 51.5 | 21.3 | 7 | 5 | 35 | 98 | 94 | 60 | 0.8 | 22.5 | 6 | 5 | 58 | 100 | 98 | 88 |
| R9 | Device failure | CI24RE/CP(810) | 2.8 | 26.3 | 6 | 3 | 45 | 26 | 28 | 76 | 2.5 | 26.3 | 6 | 5 | 43 | 82 | 24 | 92 |
| R10 | Device failure | CI24RE/CP(810) | 26.3 | 31.3 | 6 | 5 | 84 | 98 | 98 | 100 | 1.8 | 32.5 | 6 | 5 | 85 | 100 | 93 | 80 |
| R11 | Device failure | CI24RE/CP(810) | 2.6 | 25.0 | 5 | 3 | 46 | 60 | 54 | 72 | 2.5 | 23.8 | 6 | 4 | 15 | 66 | 59 | 72 |
| R12 | Infection (flap) | CI24RE/CP(810) | 1.8 | 26.3 | 5 | 3 | n/a | n/a | n/a | n/a | 1.6 | 27.5 | 6 | 5 | 79 | 100 | 94 | 92 |
| R13 | Device failure | CI24RE/CP(810) | 1.6 | 29.0 | 5 | 4 | n/a | n/a | n/a | n/a | 1.2 | 29.0 | 6 | 5 | n/a | n/a | n/a | n/a |
SP, speech processor; CI, cochlear implantation; PTA, mean pure tone audiometry thresholds (0.5, 1, 2, and 4 kHz); CAP, categorical auditory performance; SIR, speech intelligibility rating; PB word, phonetically balanced word perception; n/a, not available.
Patients undergoing cochlear implantation and hearing/speech performance after surgery.
| Number | Implant/SP | Age at CI (years) | Hearing and speech after CI implantation | |||||
|---|---|---|---|---|---|---|---|---|
| CAP | SIR | Tone recognition (%) | Easy sentence (%) | Difficult sentence (%) | PB word (%) | |||
| C1 | CI24M/CP(810) | 5.7 | 7 | 5 | 75 | 100 | 94 | 80 |
| C2 | CI24M/CP(910) | 1.8 | 6 | 5 | 55 | 98 | 87 | 80 |
| C3 | CI24R(CS)/Sprint | 4.2 | 6 | 5 | 80 | 48 | 39 | 12 |
| C4 | CI24R(CS)/CP(910) | 3.2 | 6 | 5 | 30 | 40 | 16 | 44 |
| C5 | N512/CP(810) | 2.0 | 6 | 5 | 94 | 100 | 90 | 100 |
| C6 | N512/CP(810) | 3.1 | 6 | 5 | 96 | 100 | 90 | 100 |
| C7 | N512/CP(810) | 48.2 | 6 | 5 | 68 | 100 | 97 | 40 |
| C8 | N512/CP(810) | 2.7 | 6 | 5 | 61 | 100 | 67 | 100 |
| C9 | CI24RE/Freedom | 25.3 | 6 | 5 | 83 | 100 | 97 | 64 |
| C10 | N512/CP(810) | 2.6 | 5 | 4 | 20 | 91 | 67 | 80 |
| C11 | CI24RE/CP(810) | 2.4 | 6 | 5 | 89 | 100 | 95 | 88 |