Abstract

OBJECTIVES: To assess device failures as well as early postoperative, late postoperative, and medical complications occurring after cochlear implantation and to discuss their causes and treatments. STUDY
DESIGN: Retrospective study of 500 consecutive cochlear implantations.
SETTING: Tertiary referral center. PATIENTS: All patients receiving cochlear implants at our institution between 1989 and 2006. MAIN OUTCOME MEASURES: All complications and treatments were systematically reviewed with a maximum follow-up of 18 years. The number of reimplantations was calculated according to follow-up duration to determine the cochlear implant survival rate. Specific risks of reimplantation were calculated for groups with differing durations of implantation.
RESULTS: The overall rate of complications was 16.0% (79 of 500), with minor complications accounting for 5.6%; major complications, 3.2%; and reimplantations, 7.2%. Reasons for revision surgery were device failure, infection, trauma, and "soft device failure" (failure despite normal results from integrity testing). Revision surgery was performed in 51 of the 500 cases (10.2%), and other complications were managed medically (28 of 500; 5.6%). The rate of hard and soft device failures was 6% (30 of 500). Seventy-two percent of reimplantations occurred within 5 years. The risk of severe infection (eventually requiring explantation) was 1.4% (7 of 500). There was 1 case of transient facial palsy following surgery (0.2%), and the incidence of postsurgical meningitis was 0.
CONCLUSIONS: Cochlear implantation is a safe technique with a relatively low complication rate; however, certain complications may require specific attention to prevent or correct them.

Entities:  

Mesh:

Year:  2008        PMID: 19075122     DOI: 10.1001/archoto.2008.504

Source DB:  PubMed          Journal:  Arch Otolaryngol Head Neck Surg        ISSN: 0886-4470


  33 in total

1.  Cochlear Implant Reliability: On the Reporting of Rates of Revision Surgery.

Authors:  Graham O'Neill; Neil S Tolley
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2020-04-01

2.  The pattern and degree of capsular fibrous sheaths surrounding cochlear electrode arrays.

Authors:  Reuven Ishai; Barbara S Herrmann; Joseph B Nadol; Alicia M Quesnel
Journal:  Hear Res       Date:  2017-02-17       Impact factor: 3.208

3.  Surgical findings and auditory performance after cochlear implant revision surgery.

Authors:  R Manrique-Huarte; A Huarte; M J Manrique
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-03-27       Impact factor: 2.503

4.  Endoscopic cochlear implant procedure.

Authors:  Daniele Marchioni; Alberto Grammatica; Matteo Alicandri-Ciufelli; Elisabetta Genovese; Livio Presutti
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-04-18       Impact factor: 2.503

5.  Intracochlear inflammatory response to cochlear implant electrodes in humans.

Authors:  Mohammad Seyyedi; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2014-10       Impact factor: 2.311

6.  The feasibility of endoscopic transcanal approach for insertion of various cochlear electrodes: a pilot study.

Authors:  Lela Migirov; Yisgav Shapira; Michael Wolf
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-03-12       Impact factor: 2.503

Review 7.  Surgery of the ear and the lateral skull base: pitfalls and complications.

Authors:  Bernhard Schick; Julia Dlugaiczyk
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2013-12-13

8.  Multicenter evaluation of Neurelec Digisonic® SP cochlear implant reliability.

Authors:  Luminiţa Rădulescu; Sebastian Cozma; Casimir Niemczyk; Nicolas Guevara; Ivan Gahide; John Economides; Jean-Pierre Lavieille; Renaud Meller; Jean-Pierre Bébéar; Emilien Radafy; Philippe Bordure; Djamel Djennaoui; Eric Truy
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-11-18       Impact factor: 2.503

9.  Bhopal Technique of Cochlear Implantation: A Surgical Review of 50 Cases.

Authors:  Satyaparkash Dubey; Jaskaran Singh; Bhanu Bhardwaj
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2020-06-16

10.  Postoperative complications in cochlear implants: a retrospective analysis of 438 consecutive cases.

Authors:  A Ciorba; R Bovo; P Trevisi; M Rosignoli; C Aimoni; A Castiglione; A Martini
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-11-01       Impact factor: 2.503