Endoluminal ultrasound applicators for MR-guided thermal ablation of pancreatic tumors: Preliminary design and evaluation in a porcine pancreas model.

Matthew S Adams1, Vasant A Salgaonkar2, Juan Plata-Camargo3, Peter D Jones2, Aurea Pascal-Tenorio4, Hsin-Yu Chen5, Donna M Bouley4, Graham Sommer3, Kim Butts Pauly3, Chris J Diederich1.   

Abstract

PURPOSE: Endoluminal ultrasound may serve as a minimally invasive option for delivering thermal ablation to pancreatic tumors adjacent to the stomach or duodenum. The objective of this study was to explore the basic feasibility of this treatment strategy through the design, characterization, and evaluation of proof-of-concept endoluminal ultrasound applicators capable of placement in the gastrointestinal (GI) lumen for volumetric pancreas ablation under MR guidance.
METHODS: Two variants of the endoluminal applicator, each containing a distinct array of two independently powered transducers (10 × 10 mm 3.2 MHz planar; or 8 × 10 × 20 mm radius of curvature 3.3 MHz curvilinear geometries) at the distal end of a meter long flexible catheter assembly, were designed and fabricated. Transducers and circulatory water flow for acoustic coupling and luminal cooling were contained by a low-profile polyester balloon covering the transducer assembly fixture. Each applicator incorporated miniature spiral MR coils and mechanical features (guiding tips and hinges) to facilitate tracking and insertion through the GI tract under MRI guidance. Acoustic characterization of each device was performed using radiation force balance and hydrophone measurements. Device delivery into the upper GI tract, adjacent to the pancreas, and heating characteristics for treatment of pancreatic tissue were evaluated in MR-guided ex vivo and in vivo porcine experiments. MR guidance was utilized for anatomical target identification, tracking/positioning of the applicator, and MR temperature imaging (MRTI) for PRF-based multislice thermometry, implemented in the real-time RTHawk software environment.
RESULTS: Force balance and hydrophone measurements indicated efficiencies of 48.8% and 47.8% and -3 dB intensity beam-widths of 3.2 and 1.2 mm for the planar and curvilinear transducers, respectively. Ex vivo studies on whole-porcine carcasses revealed capabilities of producing ablative temperature rise (ΔT > 15 °C) contours in pancreatic tissue 4-40 mm long and 4-28 mm wide for the planar transducer applicator (1-13 min sonication duration, ∼4 W/cm(2) applied acoustic intensity). Curvilinear transducers produced more selective heating, with a narrower ΔT > 15 °C contour length and width of up to 1-24 mm and 2-7 mm, respectively (1-7 min sonication duration, ∼4 W/cm(2) applied acoustic intensity). Active tracking of the miniature spiral coils was achieved using a Hadamard encoding tracking sequence, enabling real-time determination of each coil's coordinates and automated prescription of imaging planes for thermometry. In vivo MRTI-guided heating trials in three pigs demonstrated capability of ∼20 °C temperature elevation in pancreatic tissue at 2 cm depths from the applicator, with 5-7 W/cm(2) applied intensity and 6-16 min sonication duration. Dimensions of thermal lesions in the pancreas ranged from 12 to 28 mm, 3 to 10 mm, and 5 to 10 mm in length, width, and depth, respectively, as verified through histological analysis of tissue sections. Multiple-baseline reconstruction and respiratory-gated acquisition were demonstrated to be effective strategies in suppressing motion artifacts for clear evolution of temperature profiles during MRTI in the in vivo studies.
CONCLUSIONS: This study demonstrates the technical feasibility of generating volumetric ablation in pancreatic tissue using endoluminal ultrasound applicators positioned in the stomach lumen. MR guidance facilitates target identification, device tracking/positioning, and treatment monitoring through real-time multislice PRF-based thermometry.

Entities:  

Mesh:

Year:  2016        PMID: 27370138      PMCID: PMC4912561          DOI: 10.1118/1.4953632

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  42 in total

1.  Phase-field dithering for active catheter tracking.

Authors:  Charles L Dumoulin; Richard P Mallozzi; Robert D Darrow; Ehud J Schmidt
Journal:  Magn Reson Med       Date:  2010-05       Impact factor: 4.668

2.  Method for MRI-guided conformal thermal therapy of prostate with planar transurethral ultrasound heating applicators.

Authors:  Rajiv Chopra; Mathieu Burtnyk; Masoom A Haider; Michael J Bronskill
Journal:  Phys Med Biol       Date:  2005-10-12       Impact factor: 3.609

Review 3.  MR thermometry for monitoring tumor ablation.

Authors:  Baudouin Denis de Senneville; Charles Mougenot; Bruno Quesson; Iulius Dragonu; Nicolas Grenier; Chrit T W Moonen
Journal:  Eur Radiol       Date:  2007-05-22       Impact factor: 5.315

Review 4.  Ultrasound-guided therapeutic focused ultrasound: current status and future directions.

Authors:  Emad S Ebbini; Gail ter Haar
Journal:  Int J Hyperthermia       Date:  2015-01-23       Impact factor: 3.914

Review 5.  Systematic review of innovative ablative therapies for the treatment of locally advanced pancreatic cancer.

Authors:  S J E Rombouts; J A Vogel; H C van Santvoort; K P van Lienden; R van Hillegersberg; O R C Busch; M G H Besselink; I Q Molenaar
Journal:  Br J Surg       Date:  2014-12-18       Impact factor: 6.939

6.  Development and validation of an experimental model for the assessment of radiofrequency ablation of pancreatic parenchyma.

Authors:  Ravindra S Date; Jill Biggins; Ian Paterson; John Denton; Ray F McMahon; Ajith K Siriwardena
Journal:  Pancreas       Date:  2005-04       Impact factor: 3.327

7.  Celiac disease: evaluation with dynamic contrast-enhanced MR imaging.

Authors:  Gabriele Masselli; Antonio Picarelli; Marco Di Tola; Valerio Libanori; Giuseppe Donato; Elisabetta Polettini; Alessandra Piermattei; Piergaspare Palumbo; Angelo Pittalis; Annarita Saponara; Gianfranco Gualdi
Journal:  Radiology       Date:  2010-07-27       Impact factor: 11.105

8.  Early clinical experience using high intensity focused ultrasound for palliation of inoperable pancreatic cancer.

Authors:  Liu Lin Xiong; Joo Ha Hwang; Xiao Bo Huang; Song Sen Yao; Chong Jun He; Xiao Hua Ge; Hui Yu Ge; Xiao Feng Wang
Journal:  JOP       Date:  2009-03-09

9.  Autonomous real-time interventional scan plane control with a 3-D shape-sensing needle.

Authors:  Santhi Elayaperumal; Juan Camilo Plata; Andrew B Holbrook; Yong-Lae Park; Kim Butts Pauly; Bruce L Daniel; Mark R Cutkosky
Journal:  IEEE Trans Med Imaging       Date:  2014-06-23       Impact factor: 10.048

10.  Treatment of esophageal tumors using high intensity intraluminal ultrasound: first clinical results.

Authors:  David Melodelima; Frederic Prat; Jacques Fritsch; Yves Theillere; Dominique Cathignol
Journal:  J Transl Med       Date:  2008-06-05       Impact factor: 5.531

View more
  3 in total

Review 1.  Contactless Thermometry by MRI and MRS: Advanced Methods for Thermotherapy and Biomaterials.

Authors:  Norbert W Lutz; Monique Bernard
Journal:  iScience       Date:  2020-09-14

2.  Integration of deployable fluid lenses and reflectors with endoluminal therapeutic ultrasound applicators: Preliminary investigations of enhanced penetration depth and focal gain.

Authors:  Matthew S Adams; Vasant A Salgaonkar; Serena J Scott; Graham Sommer; Chris J Diederich
Journal:  Med Phys       Date:  2017-08-08       Impact factor: 4.071

3.  Deployable ultrasound applicators for endoluminal delivery of volumetric hyperthermia.

Authors:  Muhammad Zubair; Matthew S Adams; Chris J Diederich
Journal:  Int J Hyperthermia       Date:  2021-08-10       Impact factor: 3.914

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.