Ahsan Abdullah1, Neerja Singh2, Monika S Rathore2, Shobha Tandon2, Balakrishnan Rajkumar3.
Abstract
AIMS: The aim of this study was to evaluate the effectiveness of two different varieties of electronic apex locators and radiovisiography (RVG) for working length determination in primary teeth.Entities:
Keywords: Apex locators; Dentaport ZX; Propex II; Pulpectomy; Radiovisiography; Working length.
Year: 2016 PMID: 27365931 PMCID: PMC4921879 DOI: 10.5005/jp-journals-10005-1346
Source DB: PubMed Journal: Int J Clin Pediatr Dent ISSN: 0974-7052
Fig. 1First calibrated step for working length determination in radiovisiography
Fig. 3Double click of mouse to get the actual working length
Table 1: Range of difference in working length from radiovisiography in two methods being tested
| 0 | 9 | 30 | 21 | 70 | |||||
| 0-0.5 | 12 | 40 | 6 | 20 | |||||
| > 0.5 | 9 | 30 | 3 | 10 | |||||
X2 = 9.800 (df = 2); p = 0.007
Graph 1Graphical representation for range of difference from RVG in two methods being tested for working length
Table 2: Range of difference between Dentaport ZX and Propex II methods
| 0 | 16 | 53.3 | |||
| 0-0.5 | 12 | 40.0 | |||
| > 0.5 | 2 | 6.7 |
Graph 2The two Apex locaters showing Range of difference between Dentaport ZX and Propex II methods
Table 3: Mean difference between radiovisiography length and length measured by electronic measurement techniques
| 1 | RVG | 14.09 | 1.47 | – | – | – | – | ||||||||
| 2 | Dentaport ZX | 14.08 | 1.61 | -0.01 | 0.74 | 0.074 | 0.94 | ||||||||
| 3 | Propex II | 14.05 | 1.51 | -0.04 | 0.45 | 0.526 | 0.60 | ||||||||
SD: Standard deviation; RVG: Radiovisiography
Table 4: Correlation between radiovisiography length and length measured by electronic measurements
| RVG length | 1 | 0.888* | 0.955* | ||||
| Dentaport ZX | – | 1 | 0.949* | ||||
| Propex II | – | – | 1 |
*Is p < 0.001, RVG: Radiovisiography
Graph 3Graphical representation of correlation between the length measured by RVG and Dentaport ZX
Graph 4Graphical representation of correlation between the length measured by RVG and Propex II