Literature DB >> 18217990

Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.

M R Leonardo1, L A B Silva, P Nelson-Filho, R A B Silva, M S G G Raffaini.   

Abstract

AIM: To evaluate ex vivo the accuracy of two electronic apex locators during root canal length determination in primary incisor and molar teeth with different stages of physiological root resorption.
METHODOLOGY: One calibrated examiner determined the root canal length in 17 primary incisors and 16 primary molars (total of 57 root canals) with different stages of root resorption based on the actual canal length and using two electronic apex locators. Root canal length was measured both visually, with the placement of a K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically using two electronic apex locators (Root ZX II--J. Morita Corp. and Mini Apex Locator--SybronEndo) according to the manufacturers' instructions. Data were analysed statistically using the intraclass correlation (ICC) test.
RESULTS: Comparison of the actual root canal length and the electronic root canal length measurements revealed high correlation (ICC = 0.99), regardless of the tooth type (single-rooted and multi-rooted teeth) or the presence/absence of physiological root resorption.
CONCLUSIONS: Root ZX II and Mini Apex Locator proved useful and accurate for apex foramen location during root canal length measurement in primary incisors and molars.

Mesh:

Year:  2008        PMID: 18217990     DOI: 10.1111/j.1365-2591.2007.01366.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  21 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Evaluation of the multifrequency electronic apex locator Joypex 5 in primary teeth.

Authors:  E J N L Silva; D R Herrera; E J Souza-Júnior; T P Rosa
Journal:  Eur Arch Paediatr Dent       Date:  2013-07-24

3.  In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth.

Authors:  L Z Adriano; J C Barasuol; M Cardoso; M Bolan
Journal:  Eur Arch Paediatr Dent       Date:  2019-01-02

4.  Comparative evaluation of root ZX Mini® apex locator and radiovisiography in determining the working length of primary molars: An In Vivo study.

Authors:  Rucha N Davalbhakta; Niraj S Gokhale; Shivayogi M Hugar; Chandrashekhar M Badakar; A Gowtham; Sanjana P Soneta
Journal:  J Oral Biol Craniofac Res       Date:  2021-02-17

5.  In vivo assessment of the impedance ratio method used in electronic foramen locators.

Authors:  Marcos V H Rambo; Humberto R Gamba; Gustavo B Borba; Joaquim M Maia; Carlos A S Ramos
Journal:  Biomed Eng Online       Date:  2010-09-06       Impact factor: 2.819

6.  Digital radiography for determination of primary tooth length: in vivo and ex vivo studies.

Authors:  Maria D Basso; Fabiano Jeremias; Rita C L Cordeiro; Lourdes Santos-Pinto
Journal:  ScientificWorldJournal       Date:  2015-02-23

7.  Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth.

Authors:  Zahra Bahrololoomi; Ali Asghar Soleymani; Jalil Modaresi; Mahsa Imanian; Malihe Lotfian
Journal:  J Dent (Tehran)       Date:  2015-04

8.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

9.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20

10.  A comparative evaluation of electronic and radiographic determination of root canal length in primary teeth: An in vitro study.

Authors:  Iyer Satishkumar Krishnan; Sheela Sreedharan
Journal:  Contemp Clin Dent       Date:  2012-10
View more