Alejandro Tormo1, Francisco Celada1, Silvia Rodriguez2, Rafael Botella3, Antonio Ballesta1, Michael Kasper4, Zoubir Ouhib4, Manuel Santos2, Jose Perez-Calatayud5. 1. Radiotherapy Department, La Fe University and Polytechnic Hospital, Valencia, Spain. 2. Radiotherapy Department, Benidorm Hospital, Alicante, Spain. 3. Dermatology Department, La Fe University and Polytechnic Hospital, Valencia, Spain. 4. Radiation Oncology, Lynn Regional Cancer Center, Boca Raton, Florida, USA. 5. Radiotherapy Department, Benidorm Hospital, Alicante, Spain ; Physics Section, Department of Radiation Oncology, La Fe University Hospital, Valencia, Spain.
Abstract
PURPOSE: Radiotherapy (RT) has played a significant role in treating non melanoma skin cancer (NMSC). High-dose-rate brachytherapy (HDR-BT) approaches have a paramount relevance due to their adaptability, patient protection, and variable dose fractionation schedules. Several innovative applicators have been introduced to the brachytherapy community. The Valencia applicator is a new superficial device that improves the dose distribution compared with the Leipzig applicator. The purpose of this work is to assess the tumor control, cosmesis, and toxicity in patients with NMSC treated with the Valencia applicator and a new regimen of hypofractionation. MATERIAL AND METHODS: From January 2008 to March 2010, 32 patients with 45 NMSC lesions were treated with the Valencia applicator in the Hospital La Fe. The gross tumor volume was visually assessed, but the tumor depth was evaluated using ultrasound imaging. All lesions for the selected cases were limited to 4 mm depth. The prescription dose was 42 Gy in 6 or 7 fractions (biologically effective dose [BED] ≈ 70 Gy), delivered twice a week. RESULTS: Ninety-eight percent of the lesions were locally controlled at 47 months from treatment. Ninety-three percent of patients were out at least 36 months from treatment. The treatment was well tolerated in all cases. The highest skin toxicity was grade 1 RTOG/EORTC, having resolved with topical treatment at 4 weeks in all but one case which required 2 months. There were no grade 2 or higher late adverse events. CONCLUSIONS: In patients with superficial basal cell carcinoma lesions less than 25 mm in maximum diameter, HDRBT treatment with the Valencia applicator using a hypofractionated regimen provides excellent results, for both cosmetic and local control at a minimum of 3 years follow-up. Moreover, the shorter hypofractionated regimen facilitates compliance, which is very relevant for the elderly patients in our series. Valencia applicators offer a simple, safe, quick, and attractive nonsurgical treatment option.
PURPOSE: Radiotherapy (RT) has played a significant role in treating non melanoma skin cancer (NMSC). High-dose-rate brachytherapy (HDR-BT) approaches have a paramount relevance due to their adaptability, patient protection, and variable dose fractionation schedules. Several innovative applicators have been introduced to the brachytherapy community. The Valencia applicator is a new superficial device that improves the dose distribution compared with the Leipzig applicator. The purpose of this work is to assess the tumor control, cosmesis, and toxicity in patients with NMSC treated with the Valencia applicator and a new regimen of hypofractionation. MATERIAL AND METHODS: From January 2008 to March 2010, 32 patients with 45 NMSC lesions were treated with the Valencia applicator in the Hospital La Fe. The gross tumor volume was visually assessed, but the tumor depth was evaluated using ultrasound imaging. All lesions for the selected cases were limited to 4 mm depth. The prescription dose was 42 Gy in 6 or 7 fractions (biologically effective dose [BED] ≈ 70 Gy), delivered twice a week. RESULTS: Ninety-eight percent of the lesions were locally controlled at 47 months from treatment. Ninety-three percent of patients were out at least 36 months from treatment. The treatment was well tolerated in all cases. The highest skin toxicity was grade 1 RTOG/EORTC, having resolved with topical treatment at 4 weeks in all but one case which required 2 months. There were no grade 2 or higher late adverse events. CONCLUSIONS: In patients with superficial basal cell carcinoma lesions less than 25 mm in maximum diameter, HDRBT treatment with the Valencia applicator using a hypofractionated regimen provides excellent results, for both cosmetic and local control at a minimum of 3 years follow-up. Moreover, the shorter hypofractionated regimen facilitates compliance, which is very relevant for the elderly patients in our series. Valencia applicators offer a simple, safe, quick, and attractive nonsurgical treatment option.
Entities:
Keywords:
HDR; Valencia applicators; skin brachytherapy; skin cancer
Non melanoma skin cancer (NMSC) is the most common malignancy and its incidence has been reported in the last years to be increasing rapidly, even in young populations [1]. It affects 2-3 million people each year in the United States [2], and a similar incidence is expected in Europe. Although, NMSC is usually not a life threatening condition, it is a growing public health concern because of the significant impact on quality of life [2-4]. Surgery, cryotherapy, topical chemotherapy, photodynamic therapy, and radiotherapy (RT) are treatments options for NSMC.Surgery (Mohs or excision) is the most frequently offered treatment. For many decades, RT has played a significant role in treating NMSC [5]. Superficial X-rays, electron beam, megavoltage photons, and low- (LDR) or high-dose-rate (HDR) brachytherapy (BT) have been employed with success. There are also new emerging techniques such as electronic brachytherapy (EBT) [6]. In general, treatment options are chosen based upon institutional resources and the experiences of various specialists. With the incorporation of high-dose-rate brachytherapy (HDR-BT) and remote afterloaders in the 1960's, there has been a renewed interest in BT. High-dose-rate brachytherapy approaches have a paramount relevance due to adaptability, patient protection, and variable dose fractionations schedules, which achieve excellent cures rates and cosmetic results [7-9]. Several innovative applicators have been introduced to the BT community and the use of skin BT has increased significantly. The Valencia applicator is a new superficial device with improved dose distribution when compared to the Leipzig applicator [10, 11]. The Valencia [10, 11] applicators were designed to be used with the HDR afterloader microSelectron v2 (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden), and come in two useful sizes: 2 and 3 cm diameters. These cup-shaped applicators limit the irradiation to the treated area using tungsten-alloy shielding. The design of these applicators was based on the Leipzig applicators [12-15] with the addition of a flattening filter to improve the dose profiles and limit the penumbra. This allows focusing the radiation on the target while normal tissue irradiation is minimized.The purpose of this study is to assess the tumor control, cosmesis, and toxicity events in patients with basal cell carcinoma (BCC) treated with the Valencia applicator using a new regimen of hypofractionation.
Material and methods
Eligibility and study design
From January 2008 to March 2010, 32 patients with 45 NMSC lesions were treated with the Valencia applicators in the Radiotherapy Department of Hospital La Fe, Valencia, Spain.Patients enrolled in this study had a biopsy confirmed BCC located anywhere on the body or face except eyelids, and all cases presented as primary lesions. Initially, the maximum diameter accepted was 24. In patients with multiple lesions, all lesions were treated concurrently. All patients were over 18 years old, but no other age restrictions were applied. Data were collected retrospectively.
Treatment protocol
The protocol used for these patients was developed with the Lynn Cancer Institute of Boca Raton Regional Hospital (Boca Raton, Florida), Hospital La Fe (Valencia, Spain), and Hospital Clinica Benidorm (Benidorm, Alicante, Spain) [16]. The tumor depth was evaluated using ultrasound imaging. Ultrasounds images were performed with the Siemens Acuson S 2000 ultrasound system (Siemens AG, Munich, Germany), using a 19 mHz linear probe. To improve imaging, a gel pad was systematically used. The acquisition was done by longitudinal and transversal planes, evaluating the echogenicity and thickness of both skin and subcutaneous cellular tissue at lesion level, measuring the lesion depth. All lesions for the selected cases were limited to a maximum of 4 mm depth. This limitation was necessary to keep the skin dose at acceptable levels, since the percent depth dose of the 192Ir Valencia applicators has a gradient of about 10% per mm [10].The gross tumor volume (GTV) was generally assessed visually. The appropriate Valencia applicator size (2 or 3 cm) was chosen to allow for a minimum margin of 5 mm from the field edge. In the cases with tighter margins, the GTV was assessed by a dermatologist with a dermatoscope. The treatment dose was prescribed at 3 mm for lesions with depth of 3 mm or less, and at 4 mm for those between 3 mm and 4 mm in depth. High accuracy in tumor depth measurement is critical due to the high dose gradient. Thus, for a typical prescription to 3 mm, the surface dose will reach approximately 130%. The prescription dose was 42 Gy in 6 or 7 fractions (biological effective dose (BED) ≈ 70 Gy), delivered two times a week, with a minimum interval of 48 hours between fractions, and not exceeding 3 weeks. Immobilization was achieved in most cases using an articulated arm device provided by Nucletron. Other methods of immobilization such as simple homemade arm, tape, or thermoplastic mask, when needed, were also occasionally used. Some examples are illustrated on Figure 1. Skin marker delineating the outside applicator circumference was used to ensure reproducible treatment conditions.
Fig. 1
Examples of treatment set-ups
Examples of treatment set-upsAll treatments were delivered under the direct supervision of the radiation oncologist for accurate applicator positioning. Patients were instructed not to apply makeup or ointment prior to treatment.
End points
The end points chosen for this study were the analysis of the efficacy, safety outcomes, and cosmetic results of the treatment of NMSC with BT, using the Valencia applicators with a hypofractionated regimen. Acute and chronic toxicities were evaluated according to the Radiation Therapy Oncology Group (RTOG) and the European Organization for Research and Treatment of Cancer (EORTC) scale [17]. Patients were also consulted about their cosmetic impression using a survey with questions related to their satisfaction.
Results
Patient and treatment characteristics
Forty-five lesions in 32 patients were treated between January 2008 and March 2010. Although the age range was wide (43-97 years), the majority (70%) of these patients were over 70 years old. No recurrent lesions were included. Histology was BCC in all lesions. As usual in BCC [18], most (86%) lesions were located on the head. The median maximum diameter of the lesion was 10 mm (range: 3-25). Only two lesions had a depth greater than 3 mm (4 mm). Table 1 summarizes patient and lesion characteristics. Lesions were treated with a total dose of 42 Gy, using 6 fractions (7 Gy/fraction; twice weekly) in 96% of the cases, and 7 fractions (6 Gy/fraction; twice weekly) in the remaining 4%. The biological effective dose (BED), for both schemes is approximately 70 Gy BED (α/β: 10). The treatment was prescribed at 3 mm depth in 46 lesions and at 4 mm in 2 lesions. Median lesion diameter was 10 mm, ranged 3-25, median depth – 2 mm, ranged 0-4. We used 25 applicators size 2 (55.5%) and 20 applicators size 3 (44.5%).
Table 1
Patient and lesion characteristics
Demographics
Age (median) (years)
78
43-97
Gender
Male
18
54.5%
Female
15
45.5%
Histology
Basal cell carcinoma
45
93%
Bowen's disease
3
7%
Lesions per patient
1
26
79%
2
3
9%
3
3
9%
7
1
3%
Lesion location
Head
39
82%
Nose
17
36%
Face
11
23%
Scalp
10
21%
Ear
1
2%
Extremity
5
10%
Trunk
4
8%
Patient and lesion characteristicsThe median follow-up was 47 months (range: 31-60). In 93% of the patients, follow-up was 36 months or more.
Efficacy results
Ninety-eight percent of the lesions completely regressed. Examples of clinical results are shown in Figure 2. At 3 months, only one lesion persisted and it was centrally located. This patient had an extra-facial BCC. Tumor depth of this lesion was originally assessed in the ultrasound as 2 mm. Following the protocol, the depth used for the dose prescription was 3 mm. After subsequent review of the ultrasound images, it was established that tumor depth was actually greater than 3 mm. No other recurrences were detected after a minimum follow-up of 31 months.
Fig. 2
Some examples of clinical results, before and after treatment
Some examples of clinical results, before and after treatment
Adverse events and cosmetic results
The treatment was well tolerated in all cases. The highest skin toxicity was grade 1 RTOG/EORTC [17], having resolved with topical treatment at 4 weeks in all but one case, which required 2 months. There were no grade 2 or higher late adverse events. When patients were consulted about their cosmetic impression, all except one (a young woman with hypopigmentation at abdominal location) were very satisfied.
Discussion
Surgery is often the primary treatment for a NMSC lesion due to its reported low rates of recurrence [19, 20]. Typically, RT is an option for lesions located in areas where surgery might be accompanied with functional or cosmetic deficits, for patients with comorbidities or surgical contraindications, and for the elderly. A wide variety of radiotherapy techniques including superficial X-rays, orthovoltage X-rays, megavoltage X-rays, electron beam irradiation, low-dose-rate brachytherapy (LDR-BT), HDR-BT, and EBT are used. The term BT includes superficial and interstitial treatments, which selection depends of the tumour depth, volume to be treated, or both. Applicators in contact with the skin are used in cases of superficial and well circumscribed lesions. Commercially applicators like the Leipzig and Valencia applicators, as well as flaps and moulds are available. Customized moulds made for individual cases with different materials are also utilized. In all cases, the treatment must be comfortable for the patient and easily reproducible to avoid uncertainties.Currently, all of these approaches achieve high rates of tumour control when considering similar tumour and patient characteristics [21]. The published local control rates with HDR-BT vary from 72 to 100%. Large differences between studies were identified and are related to follow-up time, total dose, dose per fraction, time between fractions, prescription depth, and types of applicators [22-31]. There are just a few publications of clinical results of HDR-BT with standardized superficial applicators. In 1999, Köhler-Brock et al. [7] published the results at 10 years of 520 patients treated with Leipzig applicators up to 30-40 Gy at 5-10 Gy/fr. The failure rate was 8%, but Kaposi's sarcoma, melanomas, and skin metastases were included. Gahly et al. [32] and Gauden et al. [27] presented their experience with Leipzig and GM11004590 applicator, achieving complete remission in 95% and 97% of the cases, respectively, including primary and recurrent NMSC lesions (median follow-up 18 and 37 months, respectively).This study is the first one that reports clinical results with the Valencia applicators. Although the size of the study is smaller than other HDR-BT publications, in our opinion it is important for several reasons. Firstly, the study population was very homogenous in tumour histology (BCC) and lesion size (less than 3 cm). Secondly, all patients were treated in a highly consistent manner. Thirdly, the study also demonstrated safety and good tolerance in an older population (> 70 years). And finally, the local control was excellent, with mature follow-up. Performing HDR-BT with the Valencia applicators presents several advantages: dosimetric calculations are quick and safe, patient positioning and reproducibility are easy and comfortable. The hypofractionated regimen, 42 Gy in 6-7 fractions over 3 weeks or less, allows an ambulatory treatment that reduces costs, and facilitates compliance, which is essential in an elderly population as was treated in this series. In addition, the Valencia applicator using this hypofractionated regimen compares favourably with other regimens and techniques for efficacy as well as toxicity. The use of the Valencia applicators must be limited to superficial tumours (less than 4 mm depth) with a maximum diameter of 25 mm in order to guarantee adequate tumour coverage. The lateral homogeneity and flatness of the Valencia applicator is improved with respect to the Leipzig applicator and provides a smaller penumbra. The 80-20% penumbra values are significantly improved from the Leipzig (6.2-9.5 mm) to the Valencia applicators (1.9 mm). As a result, the useful treatment area of the Valencia applicator is larger. In 7 of the treated BCC lesions in this study, at least one of the margins was smaller than 5 mm, but at least 3 mm. In these 7 cases, the GTV was determined by dermatoscope. The other 35 lesions were treated with a minimum of a 5 mm margin. A recent publication based on a surgical series, suggests the following recommendations for CTV margins: 10 mm for BCC less than 2 cm, 13 mm for BCC larger than 2 cm, 11 mm for SCC less than 2 cm, and 14 mm for SCC larger than 2 cm [33]. By contrast, a prospective study with 200 BCCs showed that 2 mm dermascopically detected excision margins achieved histologically confirmed complete excision in 98.5% of cases [34]. According to Brodland et al. [35], Mohs surgery margins of 4 mm for low risk SCC and 6 mm for high risk SCC are adequate to clear the tumor in 95% of cases. Based on the results of this and other BT studies, margins of 5 mm for BCC would appear adequate.In our study, the only recurrence was centrally located. While the lesion depth was initially reported to be less than 3 mm according to the ultrasound study, a subsequent review of that image found that the depth was actually deeper. The treatment dose, in retrospect, was incorrectly prescribed to 3 mm depth. With the new data, the dose prescription would have been 4 mm. While a possible solution to avoid undertreatment is to prescribe at 1 mm beyond the lesion depth, the high dose gradient (10% for each millimeter) has critical implications both for the coverage and for the surface dose. One of the most important limitations of the study is the toxicity data collection. Nevertheless, 98% of the skin toxicity resolved after 4 weeks with topical treatment. Regarding late toxicity, data interpretation is limited by the ambiguity of the scale used (RTOG/EORTC) and the subjective criterion of a unique data collector for each case. Although it may still be early for development of late toxicity, no findings were reported. For cosmesis, no scales were used, but all patients were consulted about their cosmetic impression. The results were favourable with the exception of one young patient with an abdominal lesion reported her dissatisfaction due to hypopigmentation. Using CTC v4.0 scale [36] this event could be classed as Grade 2 due to the associated psychosocial impact. All other patients were highly satisfied.
Conclusions
In patients with superficial BCC lesions less than 25 mm in maximum diameter, HDR-BT treatment with the Valencia applicator using a hypofractionated regimen provides excellent results, for both cosmetic and local control at a minimum of 3 years follow-up. Moreover, the shorter, hypofractionated regimen facilitates compliance, which is very relevant for the elderly patients in our series. Valencia applicators offer a simple, safe, quick, and attractive nonsurgical treatment option.
Authors: Maria Maroñas; Jose Luis Guinot; Leoncio Arribas; Maria Carrascosa; Maria Isabel Tortajada; Ruth Carmona; Marian Estornell; Rodrigo Muelas Journal: Brachytherapy Date: 2010-10-08 Impact factor: 2.362
Authors: Eleni Linos; Rupa Parvataneni; Sarah E Stuart; W John Boscardin; C Seth Landefeld; Mary-Margaret Chren Journal: JAMA Intern Med Date: 2013-06-10 Impact factor: 21.873
Authors: Nicholas G Zaorsky; Charles T Lee; Eddie Zhang; Scott W Keith; Thomas J Galloway Journal: Radiother Oncol Date: 2017-08-23 Impact factor: 6.280
Authors: Anna O Likhacheva; Phillip M Devlin; Shervin M Shirvani; Christopher A Barker; Phillip Beron; Ajay Bhatnagar; Stephen W Doggett; Lawrence Hochman; Charles Hsu; Michael Kasper; Martin Keisch; Subhakar Mutyala; Bradley Prestidge; Silvia Rodriguez Villalba; Vershalee Shukla; Srinath Sundararaman; Mitchell Kamrava Journal: Brachytherapy Date: 2016-11-28 Impact factor: 2.362
Authors: Sofian Benkhaled; Dirk Van Gestel; Carolina Gomes da Silveira Cauduro; Samuel Palumbo; Veronique Del Marmol; Antoine Desmet Journal: Front Med (Lausanne) Date: 2022-06-27
Authors: Olga Pons-Llanas; Rosa Ballester-Sánchez; Francisco Javier Celada-Álvarez; Cristian Candela-Juan; Teresa García-Martínez; Margarita Llavador-Ros; Rafael Botella-Estrada; Christopher A Barker; Antonio Ballesta; Alejandro Tormo-Micó; Silvia Rodríguez; Jose Perez-Calatayud Journal: J Contemp Brachytherapy Date: 2014-12-31