Literature DB >> 27225516

Optimization of a simultaneous dual-isotope 201Tl/123I-MIBG myocardial SPECT imaging protocol with a CZT camera for trigger zone assessment after myocardial infarction for routine clinical settings: Are delayed acquisition and scatter correction necessary?

Emmanuel D'estanque1, Christophe Hedon2,3, Benoît Lattuca2, Aurélie Bourdon4, Meriem Benkiran4, Aurélie Verd4, François Roubille2,3, Denis Mariano-Goulart4,3.   

Abstract

BACKGROUND: Dual-isotope 201Tl/123I-MIBG SPECT can assess trigger zones (dysfunctions in the autonomic nervous system located in areas of viable myocardium) that are substrate for ventricular arrhythmias after STEMI. This study evaluated the necessity of delayed acquisition and scatter correction for dual-isotope 201Tl/123I-MIBG SPECT studies with a CZT camera to identify trigger zones after revascularization in patients with STEMI in routine clinical settings.
METHODS: Sixty-nine patients were prospectively enrolled after revascularization to undergo 201Tl/123I-MIBG SPECT using a CZT camera (Discovery NM 530c, GE). The first acquisition was a single thallium study (before MIBG administration); the second and the third were early and late dual-isotope studies. We compared the scatter-uncorrected and scatter-corrected (TEW method) thallium studies with the results of magnetic resonance imaging or transthoracic echography (reference standard) to diagnose myocardial necrosis.
RESULTS: Summed rest scores (SRS) were significantly higher in the delayed MIBG studies than the early MIBG studies. SRS and necrosis surface were significantly higher in the delayed thallium studies with scatter correction than without scatter correction, leading to less trigger zone diagnosis for the scatter-corrected studies. Compared with the scatter-uncorrected studies, the late thallium scatter-corrected studies provided the best diagnostic values for myocardial necrosis assessment.
CONCLUSIONS: Delayed acquisitions and scatter-corrected dual-isotope 201Tl/123I-MIBG SPECT acquisitions provide an improved evaluation of trigger zones in routine clinical settings after revascularization for STEMI.

Entities:  

Keywords:  MIBG; Trigger zone; cardiac SPECT; myocardial infarction; scatter correction

Mesh:

Substances:

Year:  2016        PMID: 27225516     DOI: 10.1007/s12350-016-0524-1

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  34 in total

Review 1.  Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association.

Authors:  Manuel D Cerqueira; Neil J Weissman; Vasken Dilsizian; Alice K Jacobs; Sanjiv Kaul; Waren K Laskey; Dudley J Pennell; John A Rumberger; Thomas Ryan; Mario S Verani
Journal:  Int J Cardiovasc Imaging       Date:  2002-02       Impact factor: 2.357

2.  Evaluation of 3D Monte Carlo-based scatter correction for 99mTc cardiac perfusion SPECT.

Authors:  Jianbin Xiao; Tim C de Wit; Steven G Staelens; Freek J Beekman
Journal:  J Nucl Med       Date:  2006-10       Impact factor: 10.057

3.  Evaluation of 3D Monte Carlo-based scatter correction for 201Tl cardiac perfusion SPECT.

Authors:  Jianbin Xiao; Tim C de Wit; Wojciech Zbijewski; Steven G Staelens; Freek J Beekman
Journal:  J Nucl Med       Date:  2007-04       Impact factor: 10.057

Review 4.  Cardiac neuronal imaging: application in the evaluation of cardiac disease.

Authors:  Maureen M Henneman; Frank M Bengel; Ernst E van der Wall; Juhani Knuuti; Jeroen J Bax
Journal:  J Nucl Cardiol       Date:  2008-04-16       Impact factor: 5.952

Review 5.  Review and current status of SPECT scatter correction.

Authors:  Brian F Hutton; Irène Buvat; Freek J Beekman
Journal:  Phys Med Biol       Date:  2011-06-23       Impact factor: 3.609

6.  Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT.

Authors:  T Ichihara; K Ogawa; N Motomura; A Kubo; S Hashimoto
Journal:  J Nucl Med       Date:  1993-12       Impact factor: 10.057

Review 7.  Cardiac sympathetic imaging with mIBG in heart failure.

Authors:  Ignasi Carrió; Martin R Cowie; Junichi Yamazaki; James Udelson; Paolo G Camici
Journal:  JACC Cardiovasc Imaging       Date:  2010-01

8.  Cardiac iodine-123 metaiodobenzylguanidine imaging predicts sudden cardiac death independently of left ventricular ejection fraction in patients with chronic heart failure and left ventricular systolic dysfunction: results from a comparative study with signal-averaged electrocardiogram, heart rate variability, and QT dispersion.

Authors:  Shunsuke Tamaki; Takahisa Yamada; Yuji Okuyama; Takashi Morita; Shoji Sanada; Yasumasa Tsukamoto; Masaharu Masuda; Keiji Okuda; Yusuke Iwasaki; Taku Yasui; Masatsugu Hori; Masatake Fukunami
Journal:  J Am Coll Cardiol       Date:  2009-02-03       Impact factor: 24.094

9.  Simultaneous Tc-99m/I-123 dual-radionuclide myocardial perfusion/innervation imaging using Siemens IQ-SPECT with SMARTZOOM collimator.

Authors:  Yong Du; Manojeet Bhattacharya; Eric C Frey
Journal:  Phys Med Biol       Date:  2014-05-13       Impact factor: 3.609

10.  123 I-mIBG scintigraphy to predict inducibility of ventricular arrhythmias on cardiac electrophysiology testing: a prospective multicenter pilot study.

Authors:  Jeroen J Bax; Otakar Kraft; Alfred E Buxton; Jan Gunnar Fjeld; Petr Parízek; Denis Agostini; Juhani Knuuti; Albert Flotats; James Arrighi; Africa Muxi; Marie-Jeanne Alibelli; Gopa Banerjee; Arnold F Jacobson
Journal:  Circ Cardiovasc Imaging       Date:  2008-07-30       Impact factor: 7.792

View more
  6 in total

1.  First assessment of simultaneous dual isotope (123I/99mTc) cardiac SPECT on two different CZT cameras: A phantom study.

Authors:  Tanguy Blaire; Alban Bailliez; Fayçal Ben Bouallegue; Dimitri Bellevre; Denis Agostini; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2017-03-08       Impact factor: 5.952

2.  Dual-isotope cardiac SPECT: the Twin Peaks Saga in nuclear cardiology.

Authors:  C Marcassa; O Zoccarato
Journal:  J Nucl Cardiol       Date:  2017-04-27       Impact factor: 5.952

Review 3.  Current Clinical Applications and Next Steps for Cardiac Innervation Imaging.

Authors:  Mark I Travin
Journal:  Curr Cardiol Rep       Date:  2017-01       Impact factor: 2.931

4.  Imaging the heart's brain: Simultaneous innervation/perfusion analysis in the era of new CZT cameras.

Authors:  Riccardo Liga; Alessia Gimelli
Journal:  J Nucl Cardiol       Date:  2016-05-18       Impact factor: 5.952

5.  Dual isotope and multidetector camera: The best choices for a specific end-point.

Authors:  Roberta Assante; Wanda Acampa
Journal:  J Nucl Cardiol       Date:  2016-05-19       Impact factor: 5.952

6.  Novel SPECT Technologies and Approaches in Cardiac Imaging.

Authors:  Piotr Slomka; Guang-Uei Hung; Guido Germano; Daniel S Berman
Journal:  Cardiovasc Innov Appl       Date:  2016-12-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.