Evaluation of 3D Monte Carlo-based scatter correction for 99mTc cardiac perfusion SPECT.

Jianbin Xiao1, Tim C de Wit, Steven G Staelens, Freek J Beekman.   

Abstract

UNLABELLED: Cardiac SPECT images are degraded by photons scattered in the thorax. Accurate correction for scatter is complicated by the nonuniform density and varied sizes of thoraxes and by the additional attenuation and scatter caused by female patients' breasts. Monte Carlo simulation is a general and accurate method for detailed modeling of scatter. Hence, for 99mTc we compared statistical reconstruction based on Monte Carlo modeling of scatter with statistical reconstruction that incorporates the more commonly used triple-energy-window scatter correction. Both of these scatter correction methods were used in conjunction with attenuation correction and resolution recovery.
METHODS: Simultaneous attenuation, detector response, and Monte Carlo-based scatter correction were implemented via the dual-matrix ordered-subset expectation maximization algorithm with a Monte Carlo simulator as part of the forward projector (ADS-MC). ADS-MC was compared to (i) ordered-subset expectation maximization with attenuation correction and with detector response modeling (AD); (ii) like (i) but with scatter correction added using the triple-energy-window method (ADS-TEW). A dual-detector SPECT system equipped with 2 153Gd scanning line sources was used for acquiring 99mTc emission data and attenuation maps. Four clinically realistic phantom configurations (a large thorax and a small thorax, each with and without breasts) with a cardiac insert containing 2 cold defects were used to evaluate the proposed reconstruction algorithms. In these phantom configurations, we compared the performance of the algorithms in terms of noise properties, contrast-to-noise ratios, contrast separability of cold defects, and robustness to anatomic variation.
RESULTS: Noise was found to be approximately 14% lower in the ADS-MC images than in the ADS-TEW and AD images. Typically, the contrast obtained with the ADS-MC algorithm was 10%-20% higher than that obtained with the other 2 methods. Furthermore, compared with the other 2 algorithms, the ADS-MC method was less sensitive to anatomic variations.
CONCLUSION: Our results indicate that the imaging performance of 99mTc SPECT can be improved more by Monte Carlo-based scatter correction than by window-based scatter correction.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17015903

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  17 in total

1.  Recommendations for reducing radiation exposure in myocardial perfusion imaging.

Authors:  Manuel D Cerqueira; Kevin C Allman; Edward P Ficaro; Christopher L Hansen; Kenneth J Nichols; Randall C Thompson; William A Van Decker; Marko Yakovlevitch
Journal:  J Nucl Cardiol       Date:  2010-08       Impact factor: 5.952

2.  Single photon-emission computed tomography.

Authors:  Thomas A Holly; Brian G Abbott; Mouaz Al-Mallah; Dennis A Calnon; Mylan C Cohen; Frank P DiFilippo; Edward P Ficaro; Michael R Freeman; Robert C Hendel; Diwakar Jain; Scott M Leonard; Kenneth J Nichols; Donna M Polk; Prem Soman
Journal:  J Nucl Cardiol       Date:  2010-10       Impact factor: 5.952

3.  Rapid Optimization of SPECT Scatter Correction Using Model LROC Observers.

Authors:  Santosh Kulkarni; Parmeshwar Khurd; Lili Zhou; Gene Gindi
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007

4.  Comparison of attenuation, dual-energy-window, and model-based scatter correction of low-count SPECT to 82Rb PET/CT quantified myocardial perfusion scores.

Authors:  R Glenn Wells; Karen Soueidan; Rachel Timmins; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2013-06-05       Impact factor: 5.952

5.  Prospects for advancing nuclear cardiology by means of new detector designs.

Authors:  Kenneth J Nichols; Andrew Van Tosh; Christopher J Palestro
Journal:  J Nucl Cardiol       Date:  2009-07-24       Impact factor: 5.952

6.  (⁹⁹m)Tc-MAA overestimates the absorbed dose to the lungs in radioembolization: a quantitative evaluation in patients treated with ¹⁶⁶Ho-microspheres.

Authors:  Mattijs Elschot; Johannes F W Nijsen; Marnix G E H Lam; Maarten L J Smits; Jip F Prince; Max A Viergever; Maurice A A J van den Bosch; Bernard A Zonnenberg; Hugo W A M de Jong
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-05-13       Impact factor: 9.236

7.  Optimization of a simultaneous dual-isotope 201Tl/123I-MIBG myocardial SPECT imaging protocol with a CZT camera for trigger zone assessment after myocardial infarction for routine clinical settings: Are delayed acquisition and scatter correction necessary?

Authors:  Emmanuel D'estanque; Christophe Hedon; Benoît Lattuca; Aurélie Bourdon; Meriem Benkiran; Aurélie Verd; François Roubille; Denis Mariano-Goulart
Journal:  J Nucl Cardiol       Date:  2016-05-25       Impact factor: 5.952

8.  Optimization of energy window and evaluation of scatter compensation methods in myocardial perfusion SPECT using the ideal observer with and without model mismatch and an anthropomorphic model observer.

Authors:  Michael Ghaly; Jonathan M Links; Eric Frey
Journal:  J Med Imaging (Bellingham)       Date:  2015-01

9.  Low-dose single acquisition rest (99m)Tc/stress (201)Tl myocardial perfusion SPECT protocol: phantom studies and clinical validation.

Authors:  Thomas Dey; Barbra E Backus; R Leo Romijn; Herfried Wieczorek; J Fred Verzijlbergen
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-09-13       Impact factor: 9.236

Review 10.  Small-animal SPECT and SPECT/CT: application in cardiovascular research.

Authors:  Reza Golestani; Chao Wu; René A Tio; Clark J Zeebregts; Artiom D Petrov; Freek J Beekman; Rudi A J O Dierckx; Hendrikus H Boersma; Riemer H J A Slart
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-01-13       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.